[Intraocular pressure difference in Goldmann applanation tonometry versus a transpalpebral tonometer TGDc-01"PRA" in glaucoma patients].

T Schlote1, H Landenberger.   

Abstract

PURPOSE: The object of this study was to compare intraocular pressure measurements obtained with the TGDc-01"PRA", a new, transpalpebral indentation tonometer, with those from Goldmann applanation tonometry in normal and glaucomatous eyes. METHODS AND PATIENTS: Forty healthy eyes and 185 eyes suffering from glaucoma were included in the study. For Goldmann tonometry three measurements and for the TGDc-01 ten measurements were performed in a random order. All participants were placed in an upright position for all measurements.
RESULTS: In both groups a systematic increase of intraocular pressure was found within the TGDc-01-measurements. Therefore, the first 3 measurements of each device were used for further statistical analysis. No learning curve could be demonstrated for the TGDc-01-measurement with normal eyes. Within the group of normal eyes the mean IOD obtained with the TGDc-01 was 1.84 mmHg lower than the mean IOD obtained with Goldmann tonometry (two-sided Student's t-test; P = 0.003). In the group of glaucomatous eyes, the mean intraocular pressure obtained with the Goldmann tonometry was 19.7 +/- 10.1 mmHg, with the TGDC-01 18.1 +/- 7.1 mmHg (coefficient of correlation r = 0.64, P < 0.001). The mean standard deviation of intraocular pressure measurements with Goldmann tonometry was 1.2 +/- 0.9 mmHg, with the TGDc-01 3.1 +/- 2.1 mmHg. TGDc-01-measurements overestimated intraocular pressure compared to Goldmann tonometry up to values of 16 mmHg and underestimated intraocular pressure at values over 16 mmHg. The difference increased by 5.5 mmHg per 10 mmHg Goldmann tonometry. The probability of success, defined as TGDc-01-recordings within +/- 3 mmHg of the Goldmann tonometry recordings, was less than 53 % between 5 and 20 mmHg and less than 30 % between 20 and 30 mmHg. Intraocular pressure (Goldmann tonometry) over 30 mmHg was always accompanied by TGDc-01-measurements lower than 3 mmHg.
CONCLUSION: In eyes with elevated intraocular pressure, the TGDc-01"PRA" significantly underestimated the intraocular pressure measurement when compared to the gold standard, Goldmann tonometry. At present, measurement of the intraocular pressure with the TGDc-01 should not be used for clinical management of patients with glaucoma.

Entities:  

Mesh:

Year:  2005        PMID: 15719316     DOI: 10.1055/s-2005-857881

Source DB:  PubMed          Journal:  Klin Monbl Augenheilkd        ISSN: 0023-2165            Impact factor:   0.700


  7 in total

Review 1.  [Measuring intraocular pressure by different methods].

Authors:  J Lamparter; E M Hoffmann
Journal:  Ophthalmologe       Date:  2009-08       Impact factor: 1.059

2.  Comparison of Diaton transpalpebral tonometer with applanation tonometry in keratoconus.

Authors:  Robert Pl Wisse; Natalie Peeters; Saskia M Imhof; Allegonda van der Lelij
Journal:  Int J Ophthalmol       Date:  2016-03-18       Impact factor: 1.779

3.  Central corneal thickness and Diaton transpalpebral tonometry.

Authors:  Mustafa Ilker Toker; Ayse Vural; Haydar Erdogan; Aysen Topalkara; Mustafa Kemal Arici
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-02-20       Impact factor: 3.117

4.  Comparison of Intraocular Pressure before and after Laser In Situ Keratomileusis Refractive Surgery Measured with Perkins Tonometry, Noncontact Tonometry, and Transpalpebral Tonometry.

Authors:  Isabel Cacho; Juan Sanchez-Naves; Laura Batres; Jesús Pintor; Gonzalo Carracedo
Journal:  J Ophthalmol       Date:  2015-06-08       Impact factor: 1.909

5.  Dual Optical Signal-based Intraocular Pressure-sensing Principle Using Pressure-sensitive Mechanoluminescent ZnS:Cu/PDMS Soft Composite.

Authors:  Yooil Kim; Sunanda Roy; Gwang-Yong Jung; Jung-Sik Oh; Gi-Woo Kim
Journal:  Sci Rep       Date:  2019-10-23       Impact factor: 4.379

6.  New intraocular pressure measurement method using reflected pneumatic pressure from cornea deformed by air puff of ring-type nozzle.

Authors:  Hyung Jin Kim; Yeong Ho Seo; Byeong Hee Kim
Journal:  PLoS One       Date:  2017-12-07       Impact factor: 3.240

7.  Utilizing the principles of Gagne's nine events of instruction in the teaching of Goldmann Applanation Tonometry.

Authors:  Yee Ling Wong
Journal:  Adv Med Educ Pract       Date:  2018-01-11
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.