Radiofrequency and Cryo-Ablation Effect on Transvenous Pacing and Defibrillatory Lead Integrity: An In Vitro Study.

Yousef H Darrat1, Anuj Agarwal2, Gustavo X Morales1, Joseph Thompson1, Ahmed Abdel-Latif1, Kelly Waespe1, Luigi DI Biase3, Andrea Natale3, Abhijit Patwardan2, Claude-Samy Elayi1.   

Abstract

INTORDUCTION: Medical societies and cardiac implantable electronic device (CIED) manufacturers recommend avoiding close or direct contact between the body of transvenous leads and ablation catheters used to treat cardiac arrhythmias. These recommendations are made despite the lack of clinical studies. However, the target myocardium for successful ablation can be contiguous to CIED leads. METHODS AND
RESULTS: We examine in vitro the effects of direct application of radiofrequency (RF) and cryo-ablation energy on the integrity and functionality of CIED leads (excluding the pacing electrodes and defibrillation coils). A saline bath was created to mimic the body milieu. CIED leads, including all commercially available lead insulation materials, were connected to a CIED pulse generator and placed in direct contact with an ablation catheter in the tissue bath. RF and cryo-ablation energy were delivered under various conditions, including maximal ablation power, temperature, and impedance via the RF generator. CIED lead functionality, reflective of conductor integrity, was evaluated through lead impedance monitoring during ablation. CIED leads were then visually inspected, and examined with optic and electron microscopy as per protocol. A total of 42 leads were studied. All leads showed the absence of insulation damage at the site of ablation visually and with microscopy. Lead functionality was also preserved in all leads.
CONCLUSION: Catheter ablation in contact with CIED leads using radiofrequency or cryo-ablation in vitro did not affect lead body integrity and function despite aggressive ablation settings. It may be reasonable to perform ablation in contact with the body of CIED leads when clinically necessary.
© 2016 Wiley Periodicals, Inc.

Entities:  

Keywords:  cardiac implantable electronic device leads; catheter ablation; cryo-ablation; implantable cardioverter defibrillator; pacemaker

Mesh:

Year:  2016        PMID: 27138905      PMCID: PMC5572796          DOI: 10.1111/jce.13007

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  12 in total

1.  Administration of isoproterenol and adenosine to guide supplemental ablation after pulmonary vein antrum isolation.

Authors:  Claude S Elayi; Luigi Di Biase; Rong Bai; J David Burkhardt; Prasant Mohanty; Pasquale Santangeli; Javier Sanchez; Richard Hongo; G Joseph Gallinghouse; Rodney Horton; Shane Bailey; Salwa Beheiry; Andrea Natale
Journal:  J Cardiovasc Electrophysiol       Date:  2013-09-10

2.  The effect of radiofrequency catheter ablation on permanent pacemakers: an experimental study.

Authors:  M C Chin; M Rosenqvist; M A Lee; J C Griffin; J J Langberg
Journal:  Pacing Clin Electrophysiol       Date:  1990-01       Impact factor: 1.976

3.  Safety and efficacy of radiofrequency energy catheter ablation of atrial fibrillation in patients with pacemakers and implantable cardiac defibrillators.

Authors:  Dhanunjaya Lakkireddy; Dimpi Patel; Kay Ryschon; Rohit Bhateja; Mihir Bhakru; Sergio Thal; Atul Verma; Oussama Wazni; Fethi Kilicaslan; Ashok Kondur; Subramanya Prasad; Jennifer Cummings; William Belden; David Burkhardt; Walid Saliba; Robert Schweikert; Mandeep Bhargava; Mina Chung; Bruce Wilkoff; Patrick Tchou; Andrea Natale
Journal:  Heart Rhythm       Date:  2005-12       Impact factor: 6.343

4.  Practice advisory for the perioperative management of patients with cardiac implantable electronic devices: pacemakers and implantable cardioverter-defibrillators: an updated report by the american society of anesthesiologists task force on perioperative management of patients with cardiac implantable electronic devices.

Authors: 
Journal:  Anesthesiology       Date:  2011-02       Impact factor: 7.892

5.  Acute effects of radiofrequency ablation of atrial arrhythmias on implanted permanent pacing systems.

Authors:  K A Ellenbogen; M A Wood; B S Stambler
Journal:  Pacing Clin Electrophysiol       Date:  1996-09       Impact factor: 1.976

6.  Relation between impedance and endocardial contact during radiofrequency catheter ablation.

Authors:  S A Strickberger; V R Vorperian; K C Man; B D Williamson; S J Kalbfleisch; C Hasse; F Morady; J J Langberg
Journal:  Am Heart J       Date:  1994-08       Impact factor: 4.749

7.  Effect of radiofrequency current on previously implanted pacemaker and defibrillator ventricular lead systems.

Authors:  M C Burke; D E Kopp; M Alberts; A Patel; A C Lin; J G Kall; M Arruda; P Mazeika; D J Wilber
Journal:  J Electrocardiol       Date:  2001       Impact factor: 1.438

8.  Radiofrequency ablation of atrial flutter and atrial tachycardias in patients with permanent indwelling catheters.

Authors:  K H Newby; L Zimerman; J M Wharton; V Kent; M Kearney; M J Brandon; A Natale
Journal:  Pacing Clin Electrophysiol       Date:  1996-11       Impact factor: 1.976

9.  Adenosine testing in atrial flutter ablation: unmasking of dormant conduction across the cavotricuspid isthmus and risk of recurrence.

Authors:  Gustavo X Morales; Laurent Macle; Paul Khairy; Richard Charnigo; Evan Davidson; Sergio Thal; C K Ching; Nicolas Lellouche; Matthew Whitbeck; Brian Delisle; Jenks Thompson; Luigi Di Biase; Andrea Natale; Stanley Nattel; Claude S Elayi
Journal:  J Cardiovasc Electrophysiol       Date:  2013-05-23

10.  Impedance changes on defibrillation coils after atrial fibrillation ablation: lead damage or electromechanical interference?

Authors:  Subba Reddy Vanga; Sudharani Bommana; Mark Kroll; Dhanunjaya Lakkireddy
Journal:  Pacing Clin Electrophysiol       Date:  2010-10-28       Impact factor: 1.976

View more
  2 in total

Review 1.  The Effects of Catheter Ablation on Permanent Pacemakers and Implantable Cardiac Defibrillators.

Authors:  Yousef H Darrat; Gustavo X Morales; Claude S Elayi
Journal:  J Innov Card Rhythm Manag       Date:  2017-03-15

2.  Epicardial Unipolar Lead Loss of Ventricular Capture during Radiofrequency Ablation of Atrial Fibrillation.

Authors:  Daniel J Chu; Wilson W Lam
Journal:  Case Rep Cardiol       Date:  2020-01-20
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.