Literature DB >> 21029124

Impedance changes on defibrillation coils after atrial fibrillation ablation: lead damage or electromechanical interference?

Subba Reddy Vanga1, Sudharani Bommana, Mark Kroll, Dhanunjaya Lakkireddy.   

Abstract

Fluctuation of impedance on defibrillator leads is highly suggestive of lead failure. A drop in impedance is associated with insulation defects while high impedance is suggestive of lead fracture. In this case report, we described a patient where electromagnetic interference from radiofrequency ablation near the site of superior vena cava coil caused the sensed impedance to fluctuate without uneventful outcomes and we discussed the possible mechanism. Both electromagnetic interference and defibrillator lead insulation defect can result in low threshold and inappropriate shocks. Sometimes electromagnetic interference may result in physical damage of the implanted system and may result in subsequent fluctuations in impedance. Awareness of electromagnetic interference that can cause fluctuating impedance without structural damage can help in making appropriate diagnosis and may avoid unnecessary procedures. ©2010, The Authors. Journal compilation ©2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2010        PMID: 21029124     DOI: 10.1111/j.1540-8159.2010.02927.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Radiofrequency and Cryo-Ablation Effect on Transvenous Pacing and Defibrillatory Lead Integrity: An In Vitro Study.

Authors:  Yousef H Darrat; Anuj Agarwal; Gustavo X Morales; Joseph Thompson; Ahmed Abdel-Latif; Kelly Waespe; Luigi DI Biase; Andrea Natale; Abhijit Patwardan; Claude-Samy Elayi
Journal:  J Cardiovasc Electrophysiol       Date:  2016-06-14

Review 2.  The Effects of Catheter Ablation on Permanent Pacemakers and Implantable Cardiac Defibrillators.

Authors:  Yousef H Darrat; Gustavo X Morales; Claude S Elayi
Journal:  J Innov Card Rhythm Manag       Date:  2017-03-15
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.