Literature DB >> 8946458

Radiofrequency ablation of atrial flutter and atrial tachycardias in patients with permanent indwelling catheters.

K H Newby1, L Zimerman, J M Wharton, V Kent, M Kearney, M J Brandon, A Natale.   

Abstract

UNLABELLED: The presence of chronic indwelling leads in the area targeted for RF ablation may pose a technical challenge and reduce the chance of success of the ablation. In addition, application of lesions in close proximity to pacemaker leads or other permanent catheters could affect their function. Fourteen patients referred for RF ablation of atrial flutter/fibrillation and atrial tachycardia, who had a permanent dual chamber pacemaker (10 patients), ICD (1 patient), or both (3 patients) were studied to assess the safety, efficacy, and effects of the ablative procedure on device function. Lead impedance, R and P wave amplitude, and pacing threshold of the defibrillator and pacemaker were measured before and after ablation. The procedure was successful in all patients. In one patient who underwent both atrial flutter and atrial fibrillation ablation, the atrial pacing threshold increased from 1.0 preablation to 2.0 V postablation. No P wave was detectable after ablation. In another patient, the P wave amplitude went from 4.0 to 2.0 mV postablation. In both patients the device converted to the power reset mode. No changes were observed in the remaining patients. Postablation defibrillator testing showed no malfunction. Follow-up reinterrogation of the devices revealed no alterations. IN
CONCLUSION: (1) RF ablation of atrial flutter and/or tachycardia is feasible even in patients with multiple chronic atrial and ventricular indwelling catheters; and (2) RF applications in close proximity of defibrillator and pacing catheters does not appear to alter their function unless lesions are produced in the area surrounding the distal pacing electrode.

Entities:  

Mesh:

Year:  1996        PMID: 8946458     DOI: 10.1111/j.1540-8159.1996.tb03188.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Radiofrequency and Cryo-Ablation Effect on Transvenous Pacing and Defibrillatory Lead Integrity: An In Vitro Study.

Authors:  Yousef H Darrat; Anuj Agarwal; Gustavo X Morales; Joseph Thompson; Ahmed Abdel-Latif; Kelly Waespe; Luigi DI Biase; Andrea Natale; Abhijit Patwardan; Claude-Samy Elayi
Journal:  J Cardiovasc Electrophysiol       Date:  2016-06-14

Review 2.  The Effects of Catheter Ablation on Permanent Pacemakers and Implantable Cardiac Defibrillators.

Authors:  Yousef H Darrat; Gustavo X Morales; Claude S Elayi
Journal:  J Innov Card Rhythm Manag       Date:  2017-03-15
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.