Sarah A Kane1, James R Blake1, Frank J McArdle1, Philip Langley2, Andrew J Sims1.
Abstract
BACKGROUND: Atrial fibrillation (AF) affects around 2% of the population and early detection is beneficial, allowing patients to begin potentially life-saving anticoagulant therapies. Blood pressure (BP) monitors may offer an opportunity to screen for AF. AIM: To identify and appraise studies which report the diagnostic accuracy of automated BP monitors used for opportunistic AF detection.Entities:
Keywords: Opportunistic Screening; Systematic Review
Year: 2016 PMID: 27099760 PMCID: PMC4836305 DOI: 10.1136/openhrt-2015-000362
Source DB: PubMed Journal: Open Heart ISSN: 2053-3624
Figure 1PRISMA diagram showing the results from each stage of the search process. PRISMA, Preferred Reporting Items for Systematic Review and Meta-Analyses.
Summary of study characteristics for each of the included studies
| Population | |||||||
|---|---|---|---|---|---|---|---|
| Paper | Device studied | Reference standard | Number of measurements per participant | Number of participants | Male (%) | Mean age (years) | AF prevalence (%) |
| Wiesel | Omron 712C | 12-lead ECG | 2 | 450 | 59 | 69 | 12.0 |
| Stergiou | Microlife BPA100 Plus | 12-lead ECG+1-lead ECG | 3 | 73 | 66 | 71 | 37.0 |
| Wiesel | Microlife BPM BP3MQ1–2D | 12-lead ECG | 3 | 405 | 51 | 73 | 23.0 |
| Marazzi | Microlife BP A200 Plus | 12-lead ECG | 1 | 503 | 54 | 67 | 20.1 |
| Marazzi | Omron M6 | 12-lead ECG | 1 | 503 | 54 | 67 | 20.1 |
| Wiesel | Microlife BPM BP3MQ1–2D | ECG event monitor | 1–4 per day | 139 | 37 | 67 | 12.9 |
| Kearley | WatchBP | 12-lead ECG | 1 | 999 | 49 | 80 | 7.9 |
| Wiesel | Microlife BP A200 Plus | 12-lead ECG | 3 | 183 | 59 | 74 | 16.4 |
| Wiesel | Omron M6 | 12-lead ECG | 1 | 183 | 59 | 74 | 16.4 |
AF, atrial fibrillation.
Summary of the results from the QUADAS-2 tool for each of the included papers
| Risk of bias | Applicability concerns | ||||||
|---|---|---|---|---|---|---|---|
| Study | Patient selection | Index test | Reference standard | Flow and timing | Patient selection | Index test | Reference standard |
| Wiesel | |||||||
| Stergiou | |||||||
| Wiesel | |||||||
| Marazzi | |||||||
| Wiesel | |||||||
| Kearley | |||||||
| Wiesel | |||||||
, low risk; , high risk; , unclear risk.
Figure 2Sensitivity and specificity of AF detection for each of the included studies. CIs have been calculated using a binomial approximation as described by Clopper and Pearson.9