Edmond W L Tang1, Benjamin H K Yip1, Chun-Pong Yu2, Samuel Y S Wong1, Eric K P Lee1.
Abstract
Background: Atrial fibrillation (AF) is a prevalent and preventable cause of stroke and mortality. Aim: This systematic review and meta-analysis aimed to investigate the sensitivity and specificity of office and out-of-office automated blood pressure (BP) devices to detect AF.Entities:
Keywords: atrial fibrillation; blood pressure; electrocardiogram (ECG); hypertension; meta-analysis; sensitivity and specificity
Year: 2022 PMID: 36035905 PMCID: PMC9411860 DOI: 10.3389/fcvm.2022.956542
Source DB: PubMed Journal: Front Cardiovasc Med ISSN: 2297-055X
Figure 1PRISMA flowchart.
Characteristics of included studies.
|
|
|
|
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|---|---|---|---|
| Office BP | Hong Kong | No | Yes | Patients with HT or DM, excluded: pacemaker, and/or an implanted defibrillator | Microlife WatchBP Home A | 12 lead ECG by cardiologist | 67.2 | 0.01 | 5,969 | 0.81 | 0.97 |
| AlAwwa ( | Jordan | No | No | Patients with ESRD undergoing maintenance hemodialysis | Omron M6 | 12-lead ECG by cardiologist | 57 | 0.08 | 227 | 0.83 | 0.94 |
| Kearley ( | UK | No | Yes | Aged 75 years or over, living at home; excluded: pacemaker, and/or an implanted defibrillator | Microlife WatchBP | 12 lead ECG by cardiologists | 79.7 | 0.08 | 999 | 0.94 | 0.9 |
| Wiesel ( | USA | No | No | Unselected outpatients in urban cardiology practise who had an ECG performed, excluded: pacemaker | Omron 712C automatic sphygmomanometer | 12-lead ECG | 69 | 0.12 | 446 | 1 | 0.91 |
| Wiesel ( | USA | Yes | No | Patients aged >= 50 years; exclude: pacemakers or defibrillators | Microlife BP A200 Omron M6 Comfort (HEM-7223-E) | 12-lead ECG | 74 | 0.16 | 183 | 1 (Microlife) | 0.92 (Microlife) 0.97 (Ormon) |
| Gandolfo ( | Italy | No | No | Recent onset stroke or TIA patients, excluded: pacemakers or defibrillators | Microlife AFib BP3MQ1-2D | 12-lead ECG by cardiologist | 77.7 | 0.18 | 207 | 0.89 | 0.99 |
| Lown ( | UK | No | Yes | Aged>65 both with and without a diagnosis of AF; excluded: pacemaker | Microlife Watch BP | 12-lead ECG by cardiologists | 73.9 | 0.19 | 418 | 0.96 | 0.94 |
| Marazzi | Italy | No | No | Consecutive patients referred to hypertension clinic. | Microlife BP A200 Plus OMRON M6 | 12-lead ECG | 67 | 0.2 | 503 | 0.92 (Microlife) 1 (Omron) | 0.95 (Microlife) 0.94 (Omron) |
| Balanis ( | Germany | Yes | No | older than 20 years old, exclusion: pacemakers | Omron BP785N (HEM-7321-Z) | 12-lead ECG | 70.2 | 0.2 | 99 | 1 | 0.85 |
| Wiesel ( | USA | Yes | No | Unselected general cardiology outpatients, Excluded: pacemaker, and/or an implanted defibrillator | BP3MQ1-2D; Microlife USA | 12-lead ECG | 73 | 0.23 | 405 | 0.97 | 0.88 |
| Watanabe ( | Japan | Yes | No | Patients with AF and cheque-up examinees | UA-1020; A&D | ECG | 57.02 | 0.27 | 280 | 1 | 0.97 |
| Stergiou ( | Greece | No | No | Patients with AF, other arrthymia or sinus rhythm. Excluded: pacemaker, and/or an implanted defibrillator | Microlife BPA100 Plus | 12-lead ECG | 70.5 | 0.37 | 73 | 1 | 0.89 |
| Kao ( | Taiwan | No | No | patients with AF and without, Excluded: pacemaker, and/or an implanted defibrillator | The Heart Spectrum Blood Pressure Monitor (P2; OSTAR Meditech Corp.) | 12-lead ECG | 67 | 0.53 | 62 | 0.9-1 (testing multiple methods) | 0.94-1 (testing multiple methods) |
| HBPM | USA | Yes | No | Patients with HT, DM, CHF, stroke; excluded: pacemakers or implantable defibrillators | Microlife BPM BP3MQ1-2D | Event recorder when conducting HBPM | 67 | 0.11 | 139 | 1 | 0.9 |
| ABPM | UK | Yes | No | An implanted pacemaker for sick sinus syndrome and previously documented AF or AHRE, excluded: diagnosed with permanent AF, or (iv) had a VVI pacemaker | Microlife WatchBPO3 AFIB | Implantable pacemaker | 71.6 | 0.08 | 48 | 0.76 | 0.81 |
| Kollias ( | Greece | Yes | No | Patients referred to a Hypertension Clinic with symptoms suggesting arrhythmias or with stroke or AF history, exclusion: pacemaker implantation | Microlife WatchBP O3 Afib | 24-h Holter monitoring | 70.55 | 0.21 | 100 | 0.93 | 0.87 |
Not listed as preferred devices or validated devices on StrideBP.org.
AF, atrial fibrillation; BP, blood pressure; DM, diabetes mellitus; ECG, electrocardiogram; ESRD, end-stage renal disease; HT, hypertension; TIA, transient ischemic attack; UK, United Kingdom.
Figure 2Quality assessment.
Figure 3Sensitivity and specificity of office blood pressure (BP) devices to detect atrial fibrillation (AF).
Figure 4Ambulatory blood pressure measurement (ABPM) reading-to-reading (A) sensitivity and (B) specificity.