| Literature DB >> 27050659 |
Saravanabavaan Suntharalingam1, Franz Ferdinand Stecker1, Nika Guberina1, Adrian Ringelstein1, Thomas Schlosser1, Jens Matthias Theysohn1, Michael Forsting1, Kai Nassenstein1.
Abstract
PURPOSE: To investigate the dose variation between follow-up CT examinations, when a patient is examined several times on the same scanner with the identical scan protocol which comprised automated exposure control.Entities:
Mesh:
Year: 2016 PMID: 27050659 PMCID: PMC4822887 DOI: 10.1371/journal.pone.0152961
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1Schematic summary of the direct and indirect determinates of the radiation exposure in CT.
Radiation dose (CTDIvol, DLP), manually (scan length, table height) or automatically (tube current, mean mAs) adjusted scan parameters as well as the maximal effective abdominal diameter as surrogate for patient’s weight of the 3 CT follow-up examinations each composed of a non-contrast-enhanced scan of the liver and a contrast-enhanced scan of the chest and abdomen in 60 patients.
| Contrast-enhanced CT of the chest | Non-enhanced CT of the liver | Contrast-enhanced CT of the abdomen | |
|---|---|---|---|
| Number of examinations (n) | 180 | 180 | 180 |
| CTDIvol [mGy] | |||
| Median: | 5.9 | 8.1 | 7.7 |
| 95% CI for the median: | 5.4 to 6.4 | 7.6 to 8.9 | 7.1 to 8.2 |
| IQR: | 2.4 | 4.3 | 3.1 |
| DLP [mGy*cm] | |||
| Median: | 226.0 | 183.0 | 363.5 |
| 95% CI for the median: | 207.2 to 241.0 | 173.4 to 197.6 | 329.7 to 394.0 |
| IQR: | 130.0 | 108.5 | 180.5 |
| Scan length [cm] | |||
| Mean: | 39.1 | 22.8 | 48.2 |
| 95% CI for the mean: | 38.3 to 39.9 | 22.3 to 23.2 | 47.6 to 48.7 |
| SD: | 5.5 | 3.0 | 3.8 |
| Table height [cm] | |||
| Mean: | 16.2 | 16.2 | 16.2 |
| 95% CI for the mean: | 15.9 to 16.5 | 15.9 to 16.5 | 15.9 to 16.5 |
| SD: | 1.9 | 1.9 | 1.9 |
| Tube voltage | |||
| 80 kV (n) | 1 | 0 | 0 |
| 100 kV (n) | 168 | 38 | 179 |
| 120 kV (n) | 11 | 111 | 1 |
| 140 kV (n) | 0 | 31 | 0 |
| Mean mAs | |||
| Median: | 80.5 | 81.5 | 114.3 |
| 95% CI for the median: | 75.8 to 86.7 | 78.6 to 83.5 | 105.8 to 119.9 |
| IQR: | 35.7 | 23.3 | 45.9 |
| Max. effective diameter [cm] | |||
| Mean: | 32.2 | 32.2 | 32.2 |
| 95% CI for the mean: | 31.6 to 32.7 | 31.6 to 32.7 | 31.6 to 32.7 |
| SD: | 3.6 | 3.6 | 3.6 |
Abbreviations: CTDIvol = volume computed tomography dose index, DLP = dose-length-product, CI = confidence interval, IQR = interquartile range, SD = standard deviation.
Differences between the follow-up examinations in radiation exposure, scan parameters and maximal effective abdominal diameter.
| Contrast-enhanced CT of the chest | Non-enhanced CT of the liver | Contrast-enhanced CT of the abdomen | |
|---|---|---|---|
| Δ CTDIvol [mGy] | |||
| Median: | 0.4 | 0.65 | 0.59 |
| 95% CI for the median: | 0.37 to 0.50 | 0.51 to 0.73 | 0.51 to 0.71 |
| IQR: | 0.63 | 0.78 | 0.88 |
| range: | 0 to 5.5 | 0 to 5.8 | 0 to 5.0 |
| Δ DLP [mGy*cm] | |||
| Median: | 21.0 | 19.0 | 37.0 |
| 95% CI for the median: | 17.4 to 23.6 | 15.4 to 21.0 | 27.4 to 42.0 |
| IQR: | 31.0 | 23 | 50.5 |
| range: | 0 to 194 | 0 to 174 | 0 to 315 |
| Δ scan length [cm] | |||
| Median: | 1.4 | 1.3 | 1.5 |
| 95% CI for the median: | 1.1 to 1.6 | 1.0 to 1.5 | 1.2 to 2.0 |
| IQR: | 1.7 | 1.7 | 2.8 |
| Δ table height [cm] | |||
| Mean: | 1.1 | 1.1 | 1.1 |
| 95% CI for the median: | 0.9 to 1.3 | 0.9 to 1.3 | 0.9 to 1.3 |
| IQR: | 1.3 | 1.3 | 1.3 |
| Δ tube voltage | |||
| 0 kV (n) | 160 | 146 | 180 |
| 20 kV (n) | 20 | 34 | 0 |
| Δ mean mAs | |||
| Median: | 6.4 | 7.9 | 8.8 |
| 95% CI for the median: | 5.4 to 7.8 | 6.1 to 9.1 | 7.1 to 10.4 |
| IQR: | 16.9 | 12.8 | 12.8 |
| Δ max. effective diameter [cm] | |||
| Median: | 0.5 | 0.5 | 0.5 |
| 95% CI for the median: | 0.4 to 0.6 | 0.4 to 0.6 | 0.4 to 0.6 |
| IQR: | 0.7 | 0.7 | 0.7 |
Abbreviations: CTDIvol = volume computed tomography dose index, DLP = dose-length-product, CI = confidence interval, IQR = interquartile range.
Fig 2Box- and Whisker-plots of the percentage differences in CTDIvol and DLP between follow-up examinations.
Correlation analysis between the differences in the radiation exposure and the differences in the parameters that varied between the follow up examinations.
| correlation between | Contrast-enhanced CT of the chest | Non-enhanced CT of the liver | Contrast-enhanced CT of the abdomen |
|---|---|---|---|
| Δ CTDIvol and Δ table height | |||
| ρ | -0.5 | -0.63 | -0.68 |
| 95% CI of ρ | -0.6 to -0.38 | -0.71 to -0.53 | -0.75 to -0.59 |
| p | <0.0001 | <0.0001 | <0.0001 |
| Δ CTDIvol and Δ tube voltage | |||
| ρ | 0.52 | 0.42 | 0.172 |
| 95% CI of r | 0.41 to 0.62 | 0.29 to 0.53 | 0.027 to 0.31 |
| p | <0.0001 | <0.0001 | 0.02 |
| Δ CTDIvol and Δ mean mAs | |||
| ρ | 0.93 | 0.96 | 0.93 |
| 95% CI of r | 0.91 to 0.95 | 0.95 to 0.97 | 0.91 to 0.95 |
| p | <0.0001 | <0.0001 | <0.0001 |
| Δ CTDIvol and Δ max. effective diameter | |||
| ρ | 0.43 | 0.41 | 0.39 |
| 95% CI of ρ | 0.31 to 0.55 | 0.28 to 0.53 | 0.26 to 0.51 |
| p | <0.0001 | <0.0001 | <0.0001 |
| Δ DLP and Δ scan length | |||
| ρ | 0.26 | 0.55 | 0.41 |
| 95% CI of ρ | 0.12 to 0.39 | 0.44 to 0.65 | 0.28 to 0.52 |
| p | 0.0005 | <0.0001 | <0.0001 |
| Δ DLP and Δ table height | |||
| ρ | -0.45 | -0.49 | -0.62 |
| 95% CI of ρ | -0.56 to -0.33 | -0.6 to -0.37 | -0.7 to -0.52 |
| p | <0.0001 | <0.0001 | <0.0001 |
| Δ DLP and Δ tube voltage | |||
| ρ | 0.51 | 0.37 | 0.03 |
| 95% CI of r | 0.39 to 0.61 | 0.24 to 0.49 | 0.02 to 0.31 |
| p | <0.0001 | <0.0001 | 0.03 |
| Δ DLP and Δ mean mAs | |||
| ρ | 0.82 | 0.67 | 0.87 |
| 95% CI of r | 0.77 to 0.87 | 0.56 to 0.75 | 0.83 to 0.90 |
| p | <0.0001 | <0.0001 | <0.0001 |
| Δ DLP and Δ max. effective diameter | |||
| ρ | 0.3 | 0.39 | 0.43 |
| 95% CI of ρ | 0.16 to 0.42 | 0.26 to 0.51 | 0.3 to 0.54 |
| p | 0.0001 | <0.0001 | <0.0001 |
Abbreviations: CTDIvol = volume computed tomography dose index, DLP = dose-length-product, CI = confidence interval.
Note:
* Calculated only for cases in which the tube voltage was kept constant between follow up examinations.
Multiple linear regression analysis to predict differences in the radiation exposure between follow-up examinations.
| Contrast-enhanced CT of the chest | Non-enhanced CT of the liver | Contrast-enhanced CT of the abdomen | |||||
|---|---|---|---|---|---|---|---|
| CTDIvol | DLP | CTDIvol | DLP | CTDIvol | DLP | ||
| Predictors | |||||||
| Δ scan length [cm] | |||||||
| coefficient | n.a. | 6.6 | n.a. | 8.4 | n.a. | 7.5 | |
| p | n.a. | <0.0001 | n.a. | <0.0001 | n.a. | <0.0001 | |
| Δ table height [cm] | |||||||
| coefficient | -0.32 | -12.4 | -0.43 | -9.0 | -0.43 | -20.7 | |
| p | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 | |
| Δ abdominal diameter [cm] | |||||||
| coefficient | 0.49 | 19.0 | 0.63 | 14.7 | 0.61 | 31.7 | |
| p | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 | |
| Multiple correlation coefficient | 0.57 | 0.6 | 0.79 | 0.82 | 0.81 | 0.83 | |
| Coefficient of determination R2 | 0.33 | 0.35 | 0.62 | 0.67 | 0.66 | 0.68 | |
Abbreviations: CTDIvol = volume computed tomography dose index, DLP = dose-length-product, n.a. = not applicable.
Note:
* Not applicable due to the fact that CTDIvol is per definition independent from the scan length.
Fig 3a-f: Scout views of three follow-up CT examinations of a 54 year old male patient with superimposed modulation of the tube current.
While the differences in the radiation dose between the chest CT examinations can be explained by differences in the table height (TH) and the scan length (SL), the difference in dose between the abdominal CT scans d and f, cannot be explained by these factors due to the fact that in d and f the table height as well as the scan length had been virtually the same. However, the positioning of the thighs was different between these examinations, which resulted in a different tube current modulation and consecutively in different radiation dose.
Influence of the phantom’s positioning on the CT table on the radiation dose.
Except of the phantom’s positioning all other imaging parameters were kept constant between the measurements.
| Chest CT | Abdominal CT | |||
|---|---|---|---|---|
| CTDIvol | DLP | CTDIvol | DLP | |
| baseline phantom examination | 4.87 mGy | 178 mGy*cm | 5.29 mGy | 275 mGy*cm |
| eccentric positioning of the phantom on the table | 3.38 mGy | 124 mGy*cm | 4.26 mGy | 222 mGy*cm |
| lifting one side of the phantom by approximately 20° | 3.37 mGy | 123 mGy*cm | 4.26 mGy | 222 mGy*cm |
| lifting one side of the phantom by approximately 90° | 3.60 mGy | 132 mGy*cm | 4.22 mGy | 220 mGy*cm |
Abbreviations: CTDIvol = volume computed tomography dose index, DLP = dose-length-product.