Radiation doses in CT examinations from the West China Hospital, Sichuan University and setting local diagnostic references levels.

Zhenlin Li1, Jinge Zhang1, Chunchao Xia1, Fei Zhao1, Kai Zhang1, Yuming Li1, Lei Li1, Jin Pu1, Wanlin Peng1, Keling Liu1, Yingkun Guo2.   

Abstract

BACKGROUND: Our study aims to summarize the data of radiation doses collected from consecutive CT examinations by using the Radiometrics software and contributing to the establishment of the region's diagnostic reference levels (DRLs).
METHODS: The radiation doses in 158,463 CT examinations performed on 106,275 adults between April 2017 and April 2019 were retrospectively analyzed. The median value and interquartile range (IQR) of volumetric CT dose index (CTDIvol), dose-length product (DLP), effective dose (ED), and size-specific dose estimate (SSDE) were calculated according to the scanning region.
RESULTS: The median CTDIvol (mGy) for each scanning region was 42.3 (head), 6.2 (chest), and 9.0 (abdomen). The median DLPs (mGy.cm) for single-phase, multi-phase, and all examinations were as follows: 607, 794, and 641 for the head; 220, 393, and 237 for the chest; 298, 1,141, and 570 for the abdomen. The median EDs (mSv) for single-phase, multi-phase, and all examinations are as follows: 1.6, 2.6, and 1.8 for the head; 5.1, 8.1, and 5.3 for the chest; 5.8, 20.3, and 10.4 for the abdomen.
CONCLUSIONS: Our study's results could provide a basis for the evaluation of CT scanning radiation dosage and supply evidence for the establishment of local DRLs in China's Sichuan Province. 2020 Annals of Translational Medicine. All rights reserved.

Entities:  

Keywords:  Radiation dosage; multidetector computed tomography; quality control; reference standards

Year:  2020        PMID: 32953810      PMCID: PMC7475485          DOI: 10.21037/atm-20-5443

Source DB:  PubMed          Journal:  Ann Transl Med        ISSN: 2305-5839


Introduction

The development of multislice spiral computed tomography (MSCT) has led to a noticeable quantum leap in the clinical performance of CT, enabling a faster and correct diagnosis of diseases. As imaging techniques develop so rapidly, using CT has increased greatly in the past few years (1). Many new techniques including CT angiography (CTA), CT perfusion imaging (CTP), and gemstone spectral imaging (GSI) are providing more and more valuable information for clinical diagnosis (2-5). However, the diagnostic benefits of CT imaging are associated with unavoidable radiation exposure. Exposure to the ionizing radiation of CT scans, even at relatively low doses, will bring incrementally increased risks of cancer, especially in women and young people (6-8). Although techniques of low-dose CT scanning are widely used in clinical applications nowadays, the radiation dose is still the biggest obstacle to popularizing using CT (3). With the public’s growing recognition of the importance of radiation protection, dose management has become a research hot spot in the field of CT imaging (9-11). Nowadays, nearly all CT scanners are required to report radiation doses as some particular values, including volumetric CT dose index (CTDIvol) and dose-length product (DLP) (12). These values are usually given in a dose report by CT modality automatically and effective dose (ED) could be calculated from the organ dosage and tissue-weighting factors. However, a single dose report only reflects the dose level of a specific single examination. Diagnostic reference levels (DRLs) with big data of consecutive CT examinations could better reflect the average dose level in individual hospitals or regions and guide radiology facilities in quality improvement efforts to optimize CT workflow (13). Some researchers or health oversight organizations in the United States and Europe have investigated the DRLs of local hospitals (13-15). Although the National Health Commission of China has issued the national standard of the radiation dose for adults in CT scan, there are few studies from dose data of a large sample of the radiation dose, and there is also a lack of research on classification by single phase scan and multiple phase scan. And in China, the ethnic and physical differences of the population in different regions lead to a large gap in the height and weight of local patients treated in hospitals in different regions (for example, people in north China are usually taller and stronger than those in the south), which will affect the parameter setting of CT scans and the final radiation doses. In addition, different regions in China have very different levels of economic development, leading to a large gap in the performance of CT equipment in public hospitals. Considering patients from different regions may have substantial differences in average body size, as well as different equipment and workflow, it is necessary to establish more local diagnostic reference levels (LDRLs) in China. The purpose of our study is to summarize the data of radiation doses collected from consecutive CT examinations in West China Hospital and to develop the DRLs of this region. We present the following article in accordance with the MDAR reporting checklist (available at http://dx.doi.org/10.21037/atm-20-5443).

Methods

Patients

This retrospective study was conducted according to the ethical guidelines of the Declaration of Helsinki (as revised in 2013) and approved by the Ethics Committee of West China Hospital. Written informed consent was obtained from all patients or their next of kin before the CT examination. All patients were informed of the benefits and risks of CT scans, the possibility that their data might be used in medical research, the privacy policy, and other related information. We collected the dose data of adults for the head, chest, and abdomen CT scans obtained between April 2017 and April 2019 from 10 scanners in West China Hospital.

Assessment and statistics of radiation doses

The scanning data were transferred from CT scanners to a single server of picture archiving and communication system (PACS) and found by Radimetrics (Bayer Healthcare, Whippany, NJ, USA). This software tool can identify and track the dose data from the dose report files on PACS, including the parameters of CTDIvol and DLP. Also, Radiometrics can calculate the size-specific dose estimate (SSDE) by calculating patient diameter from the middle-level scan length. And Radimetrics can also calculate ED automatically by using radiation exposure parameters and phantom libraries. ED is calculated on the organs exposed by the applied radiation multiplied by tissue-weighting factors. In radiometrics, the organ doses are first calculated using Monte Carlo probabilistic simulations that account for scattered radiation using a library that includes standardized male and female anthropomorphic mathematical phantoms. The ED is estimated according to the published ICRP103 tissue-weighting factors (16,17). The radiation dose was analyzed for single-phase and multiphase scans (including contrast-enhanced scans and angiography) separately. In multiphase scans, we averaged the SSDE and CTDIvol for each radiation event and added the DLPs and effective doses from each radiating event to obtain the total metric for the examination. And the brief scans obtained to determine the timing for the injection of iodinated contrast medium (monitoring trigger scans) were not included in the calculation. According to the clinical requirements of our hospital, abdominal contrast-enhanced scans are generally performed in 3 phases (plain scan, late arterial phase, and vein phase), while other areas are generally performed in 2 phases. Abdominal CT angiography is generally performed in 4 phases (plain scan, arterial phase, portal vein phase, and vein phase), chest angiography is usually performed in 3 phases, and angiography of other examined areas is usually performed in 2 phases.

CT modalities

Scans were obtained from 10 CT scanners at three institutions of West China Hospital (including headquarter and two branches). shows the specifications and other details of the ten scanners in this study.
Table 1

Details of the 10 CT scanners in this study

No.Manufacturer of CT scannerModel of CT scannerNumber of detector rowsInstallation time
CT 1SiemensSomatom Definition AS642013-5-14
CT 2PhilipsBrilliance 64642006-11-29
CT 3SiemensSomatom Definition AS642013-5-14
CT 4SiemensSomatom Definition Flash642013-5-14
CT 5SiemensSomatom Sensation 16162003-11-1
CT 6PhilipsBrilliance 16162006-7-11
CT 7SiemensSomatom Definition Flash642015-3-25
CT 8GERevolution2562016-9-5
CT 9SiemensSomatom Definition AS642012-12-7
CT 10SiemensSomatom Definition AS642009-12-18

Statistical analysis

These dose data were collected and analyzed on the Radimetrics platform so that we could get the median value and interquartile range (IQR) of CTDIvol, DLP, ED, and SSDE for each examined region. We summarized dose data of all patients included, and compare the results with other similar studies. This study is a collection and summary of large sample data, without using statistical test methods.

Results

Number of patients and examinations

Totally 158,463 diagnostic CT examinations performed on 106,275 adults (men: n=52,463; women: n=53,812) were enrolled in our retrospective analysis, including 29,762 examinations for head (18.8% of 158,463), 73,591 examinations for chest (46.4% of 158,463), 37,271 examinations for abdomen (23.5% of 158,463), and 17,839 examinations for chest & abdomen (11.3% of 158,463). The median age of all patients was 46.9±10.2 years.

Data of radiation doses

The statistics of radiation dose parameters are reported in , including median value and IQR. In general, the 75th percentiles are usually selected as DRLs. So, we used the 3rd quartile (Q3) as a simple point of reference from which hospitals of this area can compare their radiation doses. summarizes the Q3 of dose data in our study.
Table 2

Radiation dose data of each examination type in our study (numbers in parentheses are SSDEs)

Area and examination typeNumber of examinationsCTDIvol (mGy)DLP (mGy.cm)Effective dose (mSv)
25th percentile (Q1)50th percentile (Q2)75th percentile (Q3)25th percentile (Q1)50th percentile (Q2)75th percentile (Q3)25th percentile (Q1)50th percentile (Q2)75th percentile (Q3)
Head
   Single phase25,15334.541.550.13936078181.31.62.3
   Multiphase4,60938.649.253.06937941,0932.32.63.6
   All29,76234.542.351.34986418291.41.82.4
Chest
   Single phase63,2464.1 (6.2)6.3 (9.4)8.4 (11.9)1442203023.25.16.5
   Multiphase10,3454.3 (6.1)5.5 (7.9)9.7 (14.1)2893936975.88.114.1
   All73,5914.2 (6.2)6.2 (9.3)8.5 (12.0)1582373323.65.37.0
Abdomen
   Single phase17,3325.9 (9.3)7.6 (11.1)9.8 (13.2)1662984423.25.88.4
   Multiphase19,9397.5 (10.6)10.4 (14.8)12.5 (17.6)6981,1411,46612.620.328.8
   All37,2716.5 (9.7)9.0 (12.4)12.1 (16.7)2985701,1875.710.421.6
Chest & abdomen
   Single phase5,2997.2 (10.5)8.9 (12.4)10.9 (14.8)3644796597.89.812.2
   Multiphase12,5408.2 (12.2)11.8 (16.2)12.9 (19.3)1,0981,4311,80819.426.733.0
   All17,8397.9 (11.4)9.9 (14.0)12.4 (17.8)6621,1751,64812.021.030.6

SSDE, size-specific dose estimate; CTDIvol, volumetric CT dose index; DLP, dose-length product.

Table 3

Preliminary local CT DRLs from West China Hospital

Area and examination typeCTDIvol (mGy)DLP (mGy.cm)Effective dose (mSv)
Head
   Single phase50.18182.3
   Multiphase53.01,0933.6
   All51.38292.4
Chest
   Single phase8.43026.5
   Multiphase9.769714.1
   All8.53327.0
Abdomen
   Single-phase9.84428.4
   Multiphase12.51,46628.8
   All12.11,18721.6
Chest & abdomen
   Single phase10.965912.2
   Multiphase12.91,80833.0
   All12.41,64830.6

DRL, diagnostic reference level; CTDIvol, volumetric CT dose index; DLP, dose-length product.

SSDE, size-specific dose estimate; CTDIvol, volumetric CT dose index; DLP, dose-length product. DRL, diagnostic reference level; CTDIvol, volumetric CT dose index; DLP, dose-length product.

Comparison with other DRLs

And we compared the DRLs (Q3) of CTDIvol and DLP in our study with those in other researches or guidelines published in previous, including DRLs from the United States of America (13), UK (15), Japan (18), and China’s Jilin (19) and Jiangsu (20) Province. The results are shown in and 2 [Zhou et al.’s (20) study did not provide DRLs of CTDIvol. Therefore, data from China’s Jiangsu province were unincluded in the comparison of ]. It is also worth mentioning that a single-phase scan (plain scan) and a multiphase scan (including contrast-enhanced CT scan and CT angiography) were studied separately in our study and the national standard of China. However, some research only included plain scan data or did not explicitly indicate it, so that the figures only included single-phase scan dose data of our study.
Figure 1

DRLs of CTDIvol for adults from different studies. DRL, diagnostic reference level; CTDIvol, volumetric CT dose index.

DRLs of CTDIvol for adults from different studies. DRL, diagnostic reference level; CTDIvol, volumetric CT dose index.

Discussion

Presently, radiation doses of CT scans are usually reported as two values: CTDIvol and DLP (21), these two values are automatically calculated by CT scanners according to scanning parameters. CTDIvol is an index of average scanner output (mGy), which reflects specific scanning parameter choices in one scanning transverse slice related to radiation dose. DLP is the product of CTDIvol and scanning length (mGy 3 cm). CTDIvol and DLP are determined for a 32 cm phantom (all body CT examinations: chest, abdomen, trunk, and spine) and a 16 cm phantom (head CT examination). However, neither CTDIvol nor DLP indicates the specific factors of patients. The American Association of Physicists in Medicine (AAPM) task group 204 has introduced the concept of size-specific dose estimate (SSDE) to refine CTDIvol on patient size, derived from CTDIvol, and measures patient diameters using look-up tables for conversion factors (6,22,23). In other words, SSDE reflects an adjusted CTDIvol measurement according to patient size. Calculating SSDE requires manual or automatic measurement of cross-sectional diameters of patients on transverse CT images or localizers, and requires complex calculation procedures, the main obstacle to the expansion of the practical application of SSDE (13,24). In our study, the software tool Radimetrics automatically calculated SSDEs. What is more, it also calculates effective dose (ED) by using the library of cristy phantoms and matching patients to a particular phantom on gender, age, weight, or diameter. SSDE and ED both consider the factors of patients so that they can reflect the radiation dose level more accurately. The Radiometric software obtained the DLP and CTDIvol values by reading the dose report from the CT scanners, so these two values were accurate, while SSDE and ED were calculated, so their accuracy needs to be verified. Considering there have been many types of research directly using the results of Radimetrics, we did not analyze its accuracy (13,19). The accuracy and scientificity of its calculation method could be explored in the follow-up study. The term diagnostic reference level (DRL) was first introduced in 1996 by the International Commission on Radiological Protection (ICRP) (25). The DRL process involves collecting dose data of consecutive examinations from participating institutions, usually set at the 75th percentile of the histogram of aggregated dose data (13,26). DRLs provide a means for practices comparing their radiation dose to benchmarks derived from aggregated dose data collected on a local, regional, or national level (25). By setting this dose reference level, it becomes possible to determine the facilities or units with higher doses for evaluation of the reason for such doses and perform optimization. The DRL process has been popularized in the United States, Europe, and Japan, and is applied with good results (13-15). In China, although the national standard of DRLs for adults in CT scans has been issued by the National Health Commission (27), there are still few studies on dose data of a large sample of patients or multi-center and cross-regional studies. Zhou et al. (20) and Zhou et al. (19) investigated radiation dose data from CT scans in Jiangsu and Jilin provinces of China to help establish their local DRLs, but these studies did not collect data of single-phase and multi-phase scans separately. Multiphase scans, including angiography and contrast-enhanced scans, often have higher radiation doses than plain scans because they require separate scans before and after contrast injections. Moreover, the exposure parameters of multi-phase scanning are different from those of plain scanning, so it is necessary to set up a reference standard for its separate classification statistics. It is also hoped it could be counted separately, or a more detailed classification of scanning scheme could be established in the future multi-center and cross-region research. In our study, we collected dose data of consecutive clinical CT examinations over two years (Apr. 2017–Apr. 2019) at three institutions of West China Hospital (including headquarter and two branches, Yongning Branch and Shangjin Branch), and 158,463 examinations of 106,275 adults were enrolled. West China Hospital, namely West China Health Center, is the largest general hospital in Sichuan province of China, many CT examinations are performed in the department of radiology of West China Hospital daily. Meanwhile, more than 10 CT scanners from various manufactures (including Siemens, Philips, and GE) had been installed as of 2017 since 2003 (shown in ), covering old and new imaging equipment. Therefore, the dose data we collected was representative to a certain extent and could help develop the DRLs of this hospital and this region. Institutions in Sichuan Province or other areas of China can use our metrics as a starting point to assess their CT dose levels. We compared the local DRLs obtained in our study with DRLs from guidelines or research in the United States, UK, and Japan. Our results showed similarly or lowered CTDIvol and DLP. Further, compared with China’s national standards and the results of China’s Jilin and Jiangsu provinces, our DRLs are also relatively lower (shown in ). These results may be attributed to the continuous progress of low-dose CT scanning technology, as well as the emphasis on radiation doses in West China Hospital and the corresponding measures taken. For dose optimization, we have several suggestions on the experience of West China Hospital. First, enable real-time automatic exposure control technology when scanning, including Siemens’ Care Dose 4D (28), Care kV (29) technology, and so on. Secondly, the iterative reconstruction algorithm could significantly reduce the radiation dose on the premise of ensuring image quality, including Siemens’ IRIS technology (30) and GE’s ASIR and ASIR-V technology (31). Thirdly, the training and learning of CT technicians should be strengthened to construct CT scan workflow more standardized in the preparation of patients before the examination and the setting of scan parameters to accurately formulate the scanning plan and avoid rescanning due to operational errors. Fourthly, a complete radiation dose and image quality control system should be established to review and analyze the cases of excessive radiation doses regularly. As for the specific impact of these techniques or measures on the reduction of radiation dose, we are conducting further comparative research with the help of the Radimetrics software to explore it. It must be noted, due to the huge differences in patient ethnicity, patient body size, scanning sequence, scanning equipment, workflow and other factors between hospitals in each country or region, the differences between DRLs reported from different studies may not be as significant actually, but more about helping set up local DRLs. Moreover, the establishment of DRLs is mainly to provide reference for radiation dose control in local hospitals, but for a specific patient, the scanning parameters and radiation dose of CT scan should be optimized according to their body size, gender, age, medical history and other factors on the premise of meeting clinical diagnosis.
Figure 2

DRLs of DLP for adults from different studies. DRL, diagnostic reference level; DLP, dose-length product.

DRLs of DLP for adults from different studies. DRL, diagnostic reference level; DLP, dose-length product. Our study has some limitations. First, only two years of data in one hospital were enrolled, meaning the sample size was small. Dose data from more hospitals and more patients need to be included, only then will DRLs on regional or national levels be set up. Most hospitals in the Sichuan province of China do not yet have Radiometrics, or similar software installed and not join the data-sharing network, we are not yet able to obtain more hospitals’ data. It is expected the cloud platform of the Sichuan radiation quality control center under construction and the multi-center research will help a lot to set up the DRLs on the province level. Together, we hope West China Hospital and other hospitals will continue to update and add to these dose data for the creation of benchmarks, as only constantly updated DRLs are meaningful. Then, with some constraints, we only analyzed the examined area of the head, chest, and abdomen. More examined body parts should be included in a future study. And the categories are not detailed enough, for example, the examinations of the abdomen in our study included the upper abdomen, lower abdomen, whole abdomen, abdomen, angiography, and other types of abdomen CT scans. So, the value of our local DRLs for clinical guidance is limited, and further research is needed. The article’s supplementary files as
  25 in total

1.  Radiation oncology organization, FDA announce radiation safety initiatives.

Authors:  Tracy Hampton
Journal:  JAMA       Date:  2010-04-07       Impact factor: 56.272

2.  Radiation Doses in Consecutive CT Examinations from Five University of California Medical Centers.

Authors:  Rebecca Smith-Bindman; Michelle Moghadassi; Nicole Wilson; Thomas R Nelson; John M Boone; Christopher H Cagnon; Robert Gould; David J Hall; Mayil Krishnam; Ramit Lamba; Michael McNitt-Gray; Anthony Seibert; Diana L Miglioretti
Journal:  Radiology       Date:  2015-05-19       Impact factor: 11.105

3.  Assessment of carotid plaque composition using fast-kV switching dual-energy CT with gemstone detector: comparison with extracorporeal and virtual histology-intravascular ultrasound.

Authors:  Yuki Shinohara; Makoto Sakamoto; Keita Kuya; Junichi Kishimoto; Naoki Iwata; Yasutoshi Ohta; Shinya Fujii; Takashi Watanabe; Toshihide Ogawa
Journal:  Neuroradiology       Date:  2015-05-16       Impact factor: 2.804

4.  Improvement of image quality and radiation dose of CT perfusion of the brain by means of low-tube voltage (70 KV).

Authors:  Zhen-lin Li; Hang Li; Kai Zhang; Wang-jiang Li; Xian Chen; Bin Wu; Bin Song
Journal:  Eur Radiol       Date:  2014-06-04       Impact factor: 5.315

5.  Raw data-based iterative reconstruction in body CTA: evaluation of radiation dose saving potential.

Authors:  Anna Winklehner; Christoph Karlo; Gilbert Puippe; Bernhard Schmidt; Thomas Flohr; Robert Goetti; Thomas Pfammatter; Thomas Frauenfelder; Hatem Alkadhi
Journal:  Eur Radiol       Date:  2011-08-06       Impact factor: 5.315

6.  Simplifying size-specific radiation dose estimates in pediatric CT.

Authors:  Ranish Deedar Ali Khawaja; Sarabjeet Singh; Beth Vettiyil; Ruth Lim; Michael Gee; Sjirk Westra; Mannudeep K Kalra
Journal:  AJR Am J Roentgenol       Date:  2015-01       Impact factor: 3.959

7.  Size-specific dose estimates for adult patients at CT of the torso.

Authors:  Jodie A Christner; Natalie N Braun; Megan C Jacobsen; Rickey E Carter; James M Kofler; Cynthia H McCollough
Journal:  Radiology       Date:  2012-10-22       Impact factor: 11.105

8.  Single-stage hybrid localization: a combination of bronchoscopic lung mapping followed by post-mapping computed tomographic reconstruction and additional transthoracic needle localization in a cone beam computed tomography room.

Authors:  Shun-Mao Yang; Wei-Chun Ko; Ling-Hsuan Meng; Li-Wei Chen; Kun-Hsien Lin; Yueh-Lun Liu; Shao-En Sun; Huan-Jang Ko
Journal:  Ann Transl Med       Date:  2019-01

9.  Estimates of the potential risk of radiation-related cancer from screening in the UK.

Authors:  Amy Berrington de González
Journal:  J Med Screen       Date:  2011       Impact factor: 2.136

10.  New adaptive statistical iterative reconstruction ASiR-V: Assessment of noise performance in comparison to ASiR.

Authors:  Paolo De Marco; Daniela Origgi
Journal:  J Appl Clin Med Phys       Date:  2018-01-24       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.