| Literature DB >> 27046200 |
Manjeetha Jaggernath1, Rumallen Naicker2, Savathree Madurai2, Mark A Brockman3,4,5, Thumbi Ndung'u1,6,5,7, Huub C Gelderblom1,8.
Abstract
In South Africa, various point-of-care hemoglobin meters are used. However, the regulatory framework for approval, implementation and oversight of use of point-of-care hemoglobin meters is suboptimal. We assessed the diagnostic accuracy of the HemoCue Hb 301, STAT-Site MHgb and URIT-12 point-of-care hemoglobin meters, compared to a central laboratory based reference assay, in a central laboratory and a community based clinic in Durban, South Africa. Differences in performance of the point-of-care assays, compared to the reference assay, were more pronounced in the community based clinic. Results were reasonable for the HemoCue Hb 301, but poor for the STAT-Site MHgb and the URIT-12. Poor test performance of point-of-care hemoglobin meters, and inadequate evaluations and oversight in South Africa, leads to suboptimal clinical care and clinical research, and increased costs. There is a need for proper evaluation and quality assurance of point-of-care tests, the results of which should be made widely available to key stakeholders.Entities:
Mesh:
Substances:
Year: 2016 PMID: 27046200 PMCID: PMC4821624 DOI: 10.1371/journal.pone.0152184
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1Correlation of the 3 point-of-care assays with values within the dynamic range of the reference Hemoglobin meter in 60 samples in a central laboratory (phase 1), and 100 samples in a community based clinical setting (phase 2).
In phase 1 of the study the STAT-Site had a high failure rate. It seems that this was related to uneven migration of the sample through the sample strip, resulting in the sample migration time exceeding the maximum time specified by the manufacturer.
Fig 2Bland-Altman plots comparing the reference laboratory test with the 3 point-of-care assays in phase 1 (left column) and phase 2 (right column).
In phase 1 of the study the STAT-Site had a high failure rate. It seems that this was related to uneven migration of the sample through the sample strip, resulting in the sample migration time exceeding the maximum time specified by the manufacturer.
Phase 2 (community based clinical setting) sensitivity, specificity, and predictive values for the 3 point-of-care tests to detect or exclude anemia.
| Females (12 g/dl cut-off) | ||||||||
| Point-of-care tests | reference | |||||||
| Outcome | Positive | Negative | Subtotal | sensitivity | specificity | PPV | NPV | |
| HemoCue | Positive | 26 | 8 | 34 | 72% | 70% | 76% | 66% |
| Negative | 10 | 19 | 29 | |||||
| Total | 36 | 27 | 63 | |||||
| Stat-Site | Positive | 30 | 11 | 41 | 83% | 59% | 73% | 73% |
| Negative | 6 | 16 | 22 | |||||
| Total | 36 | 27 | 63 | |||||
| URIT | Positive | 36 | 25 | 61 | 100% | 7% | 59% | 100% |
| Negative | 0 | 2 | 2 | |||||
| Total | 36 | 27 | 63 | |||||
| Males (13 g/dl cut-off) | ||||||||
| Point-of-care tests | reference | |||||||
| Outcome | Positive | Negative | Subtotal | sensitivity | specificity | PPV | NPV | |
| HemoCue | Positive | 2 | 10 | 12 | 50% | 70% | 17% | 92% |
| Negative | 2 | 23 | 25 | |||||
| Total | 4 | 33 | 37 | |||||
| Stat-Site | Positive | 4 | 10 | 14 | 100% | 70% | 29% | 100% |
| Negative | 0 | 23 | 23 | |||||
| Total | 4 | 33 | 37 | |||||
| URIT | Positive | 4 | 31 | 35 | 100% | 6% | 11% | 100% |
| Negative | 0 | 2 | 2 | |||||
| Total | 4 | 33 | 37 | |||||
| Females and males (using the respective cut-off values) | ||||||||
| Point-of-care tests | reference | |||||||
| Outcome | Positive | Negative | Subtotal | sensitivity | specificity | PPV | NPV | |
| HemoCue | Positive | 28 | 18 | 46 | 70% | 70% | 61% | 78% |
| Negative | 12 | 42 | 54 | |||||
| Total | 40 | 60 | 100 | |||||
| Stat-Site | Positive | 34 | 21 | 55 | 85% | 65% | 62% | 87% |
| Negative | 6 | 39 | 45 | |||||
| Total | 40 | 60 | 100 | |||||
| URIT | Positive | 40 | 56 | 96 | 100% | 7% | 42% | 100% |
| Negative | 0 | 4 | 4 | |||||
| Total | 40 | 60 | 100 | |||||
Sensitivity = % of patients with anemia, according to the reference laboratory test, that were identified as anemic by the point-of-care test; specificity = % of patients without anemia, according to the reference laboratory test, that were identified as non-anemic by the point-of-care test; PPV, positive predictive value = % of patients that were identified as anemic by the point-of-care test that were confirmed as anemic by the reference laboratory test; NPV, negative predictive value = % of patients that were identified as non-anemic by the point-of-care test that were confirmed as non-anemic by the reference laboratory test.
Fig 3Correlation of the 3 point-of-care assays with values within the dynamic range of the reference hemoglobin meter in 100 samples in a community based clinical setting in phase 2 of the study.
Dots in the blue quadrant indicate that the point-of-care assay missed anemia, i.e., misclassified a finger prick sample as non-anemic where the venous blood sample was classified as anemic according to the reference assay in a central laboratory. Dots in the red quadrant indicate that the point-of-care assay misclassified as anemic a sample that was non-anemic according to the reference assay.