[Comparison between Coulter LH-750 and URIT-12 for measuring hemoglobin concentration].

Xue Qin1, Yan-ping Liu, Fa-quan Lin, Zhen Yi, Ming-hui Xu, Ruo-lin Li, Tai-jie Li, Shan Li.   

Abstract

OBJECTIVE: To evaluate the feasibility of URIT-12 hemoglobin analyzer for fast measurement of hemoglobin concentration.
METHODS: Hemoglobin concentration was detected in 100 random blood samples using URIT-12 hemoglobin analyzer and Coulter LH-750 hematology analyzer.
RESULTS: The two analyzers showed good correlation of the results (r=0.994) without significant difference between them (P>0.05). The linear range of URIT-12 hemoglobin analyzer was 46-240 g/L, and in the repeated measurements (20 times) of 3 batches of blood samples with low, moderate and high hemoglobin concentrations, the within-batch coefficient of variation of URIT-12 hemoglobin analyzer, from low to high concentrations, were 2.13%, 2.17%, and 2.33%, respectively. In the measurement of 4 batches of high-fat, high-bilirubin, high-globin and high-white-blood-cell blood samples, the interference rate of the former 3 samples were all less than 4% by the two devices, but that of the fourth sample was 10% by URIT-12 hemoglobin analyzer and 7% by Coulter LH-750 analyzer.
CONCLUSION: The results detected by URIT-12 hemoglobin analyzer have high accuracy and precision and is easy to operate, fast-testing and portable.

Entities:  

Mesh:

Year:  2008        PMID: 19114355

Source DB:  PubMed          Journal:  Nan Fang Yi Ke Da Xue Xue Bao        ISSN: 1673-4254


  1 in total

1.  Diagnostic Accuracy of the HemoCue Hb 301, STAT-Site MHgb and URIT-12 Point-of-Care Hemoglobin Meters in a Central Laboratory and a Community Based Clinic in Durban, South Africa.

Authors:  Manjeetha Jaggernath; Rumallen Naicker; Savathree Madurai; Mark A Brockman; Thumbi Ndung'u; Huub C Gelderblom
Journal:  PLoS One       Date:  2016-04-05       Impact factor: 3.240

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.