Performance evaluation of the Sysmex XS-1000i automated haematology analyser.

T Ghys1, R Malfait, J VAN den Bossche.   

Abstract

The Sysmex XS-1000i is a compact new, fully automated haematology analyser, designed to generate complete blood counts with five-part leucocyte differential. In our study, a Sysmex XS-1000i instrument was evaluated according to Clinical Laboratory Standards Institute (CLSI) and International Council for Standardization in Haematology (ICSH) guidelines. Precision, carry-over and linearity were determined. Using a total of 700 patient samples, results from the Sysmex XS-1000i were compared with those from a Sysmex XE-2100, an Abbott Cell Dyn 4000 and the manual reference leucocyte differential. Using quality control material, total and within-run imprecision was less than 3% except for platelets. The system demonstrated good linearity over the entire reporting range and no carry-over (<0.5%). The Sysmex XS-1000i showed good correlation with XE-2100, CD-4000 and the manual reference leucocyte differential. Overall flagging sensitivity and specificity were 91% and 48%, respectively. In conclusion, the Sysmex XS-1000i demonstrated good analytical performance, is able to generate a complete blood count with five-part differential on low blood volumes and has considerable back-up capacity.

Entities:  

Mesh:

Year:  2009        PMID: 18616754     DOI: 10.1111/j.1751-553X.2008.01081.x

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  5 in total

1.  Performance Evaluation of Fully Automated 5-Part Differential Hematology Analyzer ELite 580 (ERBA).

Authors:  Shanaz Khodaiji; Dia Mansukhani
Journal:  Indian J Hematol Blood Transfus       Date:  2021-03-16       Impact factor: 0.900

2.  Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers.

Authors:  Shubham Rastogi; Sukesh Chandran Nair; Pandiyan Murugan; Asady Sukanya Sukumar; Joy J Mammen; Saravanan Mullai
Journal:  Pract Lab Med       Date:  2022-04-18

3.  Initial performance evaluation of the UniCel® DxH 800 Coulter® cellular analysis system.

Authors:  B D Hedley; M Keeney; I Chin-Yee; W Brown
Journal:  Int J Lab Hematol       Date:  2011-02       Impact factor: 2.877

4.  Assessment of noninvasive, percutaneous hemoglobin measurement in pregnant and early postpartum women.

Authors:  Atsushi Yoshida; Keiko Saito; Kenji Ishii; Isao Azuma; Hidenori Sasa; Kenichi Furuya
Journal:  Med Devices (Auckl)       Date:  2014-01-15

5.  Diagnostic Accuracy of the HemoCue Hb 301, STAT-Site MHgb and URIT-12 Point-of-Care Hemoglobin Meters in a Central Laboratory and a Community Based Clinic in Durban, South Africa.

Authors:  Manjeetha Jaggernath; Rumallen Naicker; Savathree Madurai; Mark A Brockman; Thumbi Ndung'u; Huub C Gelderblom
Journal:  PLoS One       Date:  2016-04-05       Impact factor: 3.240

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.