| Literature DB >> 27017271 |
Abu Naser Mohon1, Lydia Da-Yeong Lee2, Abebe Genetu Bayih3, Asongna Folefoc3, Dylan Guelig4, Robert A Burton4, Paul LaBarre4, Wilson Chan3, Bonnie Meatherall5, Dylan R Pillai6.
Abstract
Microscopy and field adaptable rapid diagnostic tests (RDTs) are not sensitive and specific in certain conditions such as poor training of microscopists, lack of electricity, or lower sensitivity in the detection of non-falciparum malaria. More sensitive point-of-care testing (POCT) would reduce delays in diagnosis and initiation of therapy. In the current study, we have evaluated the efficacy of noninstrumented nucleic acid amplification (NINA) coupled with loop-mediated isothermal amplification (LAMP) for detection of traveler's malaria (n=140) in comparison with microscopy, nested PCR, and the only Food and Drug Administration-approved rapid diagnostic test. NINA-LAMP was 100% sensitive and 98.6% specific when compared to nested PCR. For non-falciparum detection, NINA-LAMP sensitivity was 100% sensitive compared to nested PCR, whereas RDT sensitivity was 71%. LAMP is highly sensitive and specific for symptomatic malaria diagnosis regardless of species.Entities:
Keywords: LAMP; Malaria; Point of care test
Mesh:
Year: 2015 PMID: 27017271 PMCID: PMC4862928 DOI: 10.1016/j.diagmicrobio.2015.11.009
Source DB: PubMed Journal: Diagn Microbiol Infect Dis ISSN: 0732-8893 Impact factor: 2.803