Comparison of internal and total optical aberrations for 2 aberrometers: iTrace and OPD scan.

Jung Bin Won1, Sun Woong Kim, Eung Kweon Kim, Byung Jin Ha, Tae-im Kim.   

Abstract

PURPOSE: To compare and evaluate the total and internal aberrations measured by two aberrometers: the laser ray tracing aberrometer (iTrace, Tracey Technology) and the automatic retinoscope aberrometer (OPD Scan, Nidek).
METHODS: A total of 54 healthy eyes were enrolled in the study. Following pupil dilation, aberrations were measured with the iTrace and OPD Scan. We compared the aberrations obtained from measurements obtained at pupillary diameters of 4 mm and 6 mm with the OPD Scan and iTrace. Aberrations of internal optics and total aberrations were compared for the two aberrometers. For each aberrometer and each eye, the averaged Zernike data were used to calculate various root-mean-square (RMS) data. These parameters, together with the refractive parameters, were then analyzed and complimented by paired t-tests.
RESULTS: At a pupil diameter of 4 mm, the number of total aberrations in the entire eye showed significant differences for the mean values of spherical aberrations (Z4,0) obtained with the OPD Scan and iTrace aberrometers (p=0.001). Aberrations of the internal optics showed significant differences in the mean values of total RMS, coma (Z3,-1), and trefoil (Z3,3) between the iTrace and OPD Scan (p<0.001, p=0.01, p<0.001) for the same pupil diameter of 4 mm. At a pupil diameter of 6 mm, the two instruments showed a similar number of total aberrations. Aberrations of the internal optics showed significant differences in the mean values of total RMS, spherical aberration (Z4,0), and coma (Z3,-1) between the two devices (p<0.001, p=0.01, p<0.001).
CONCLUSIONS: The iTrace and OPD Scan showed the largest number of differences for aberrations of internal optics rather than total aberrations for both pupil diameters. These results suggest that in healthy eyes, the two aberrometers may vary in some details. The aberrometers showed more agreement at a pupil diameter of 6 mm compared to 4 mm.

Entities:  

Mesh:

Year:  2008        PMID: 19096236      PMCID: PMC2629919          DOI: 10.3341/kjo.2008.22.4.210

Source DB:  PubMed          Journal:  Korean J Ophthalmol        ISSN: 1011-8942


  8 in total

1.  Statistical variation of aberration structure and image quality in a normal population of healthy eyes.

Authors:  Larry N Thibos; Xin Hong; Arthur Bradley; Xu Cheng
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2002-12       Impact factor: 2.129

2.  Ocular higher-order aberrations in individuals screened for refractive surgery.

Authors:  Li Wang; Douglas D Koch
Journal:  J Cataract Refract Surg       Date:  2003-10       Impact factor: 3.351

3.  Test-retest reliability of clinical Shack-Hartmann measurements.

Authors:  Xu Cheng; Nikole L Himebaugh; Pete S Kollbaum; Larry N Thibos; Arthur Bradley
Journal:  Invest Ophthalmol Vis Sci       Date:  2004-01       Impact factor: 4.799

4.  Ocular optical aberrations after photorefractive keratectomy for myopia and myopic astigmatism.

Authors:  T Seiler; M Kaemmerer; P Mierdel; H E Krinke
Journal:  Arch Ophthalmol       Date:  2000-01

5.  Comparison of higher-order wavefront aberrations with 3 aberrometers.

Authors:  Chung-Ling Liang; Suh-Hang Hank Juo; Cheng-Jong Chang
Journal:  J Cataract Refract Surg       Date:  2005-11       Impact factor: 3.351

6.  Clinical comparison of 6 aberrometers. Part 2: statistical comparison in a test group.

Authors:  Jos J Rozema; Dirk E M Van Dyck; Marie-José Tassignon
Journal:  J Cataract Refract Surg       Date:  2006-01       Impact factor: 3.351

Review 7.  Wavefront aberration of the eye: a review.

Authors:  W N Charman
Journal:  Optom Vis Sci       Date:  1991-08       Impact factor: 1.973

8.  In vivo and in vitro repeatability of Hartmann-Shack aberrometry.

Authors:  Alireza Mirshahi; Jens Bühren; Daniel Gerhardt; Thomas Kohnen
Journal:  J Cataract Refract Surg       Date:  2003-12       Impact factor: 3.351

  8 in total
  7 in total

1.  Higher order aberration comparison between two aspherical intraocular lenses: MC6125AS and Akreos advanced optics.

Authors:  Mohammad Taher Rajabi; Sara Korouji; Mahgol Farjadnia; Mohammad Naderan; Mohammad Bagher Rajabi; Bahram Khosravi; Seyed Mehdi Tabatabaie
Journal:  Int J Ophthalmol       Date:  2015-06-18       Impact factor: 1.779

2.  Comparison of two types of visual quality analyzer for the measurement of high order aberrations.

Authors:  Jing Hao; Lin Li; Fang Tian; Hong Zhang
Journal:  Int J Ophthalmol       Date:  2016-02-18       Impact factor: 1.779

3.  Comparative analysis of the visual performance after implantation of the toric implantable collamer lens in stable keratoconus: a 4-year follow-up after sequential procedure (CXL+TICL implantation).

Authors:  Farideh Doroodgar; Feazollah Niazi; Azad Sanginabadi; Sana Niazi; Alireza Baradaran-Rafii; Cyrus Alinia; Eznollah Azargashb; Mohammad Ghoreishi
Journal:  BMJ Open Ophthalmol       Date:  2017-09-28

Review 4.  From Presbyopia to Cataracts: A Critical Review on Dysfunctional Lens Syndrome.

Authors:  Joaquín Fernández; Manuel Rodríguez-Vallejo; Javier Martínez; Ana Tauste; David P Piñero
Journal:  J Ophthalmol       Date:  2018-06-27       Impact factor: 1.909

5.  Changes and Diurnal Variation of Visual Quality after Orthokeratology in Myopic Children.

Authors:  Hao-Chen Guo; Wan-Qing Jin; An-Peng Pan; Qin-Mei Wang; Jia Qu; A-Yong Yu
Journal:  J Ophthalmol       Date:  2018-10-15       Impact factor: 1.909

6.  Comparison of Ocular Monochromatic Higher-order Aberrations in Normal Refractive Surgery Candidates of Arab and South Asian Origin.

Authors:  Gaurav Prakash; Dhruv Srivastava; Sounak Choudhuri; Ruthchel Bacero
Journal:  Middle East Afr J Ophthalmol       Date:  2016 Jan-Mar

7.  Precision and agreement of higher order aberrations measured with ray tracing and Hartmann-Shack aberrometers.

Authors:  Zequan Xu; Yanjun Hua; Wei Qiu; Guoqiang Li; Qiang Wu
Journal:  BMC Ophthalmol       Date:  2018-01-27       Impact factor: 2.209

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.