Toward improved target conformity for two spot scanning proton therapy delivery systems using dynamic collimation.

Alexandra Moignier1, Edgar Gelover1, Blake R Smith1, Dongxu Wang1, Ryan T Flynn1, Maura L Kirk2, Liyong Lin2, Timothy D Solberg2, Alexander Lin2, Daniel E Hyer1.   

Abstract

PURPOSE: To quantify improvement in target conformity in brain and head and neck tumor treatments resulting from the use of a dynamic collimation system (DCS) with two spot scanning proton therapy delivery systems (universal nozzle, UN, and dedicated nozzle, DN) with median spot sizes of 5.2 and 3.2 mm over a range of energies from 100 to 230 MeV.
METHODS: Uncollimated and collimated plans were calculated with both UN and DN beam models implemented within our in-house treatment planning system for five brain and ten head and neck datasets in patients previously treated with spot scanning proton therapy. The prescription dose and beam angles from the clinical plans were used for both the UN and DN plans. The average reduction of the mean dose to the 10-mm ring surrounding the target between the uncollimated and collimated plans was calculated for the UN and the DN. Target conformity was analyzed using the mean dose to 1-mm thickness rings surrounding the target at increasing distances ranging from 1 to 10 mm.
RESULTS: The average reductions of the 10-mm ring mean dose for the UN and DN plans were 13.7% (95% CI: 11.6%-15.7%; p < 0.0001) and 11.5% (95% CI: 9.5%-13.5%; p < 0.0001) across all brain cases and 7.1% (95% CI: 4.4%-9.8%; p < 0.001) and 6.3% (95% CI: 3.7%-9.0%; p < 0.001), respectively, across all head and neck cases. The collimated UN plans were either more conformal (all brain cases and 60% of the head and neck cases) than or equivalent (40% of the head and neck cases) to the uncollimated DN plans. The collimated DN plans offered the highest conformity.
CONCLUSIONS: The DCS added either to the UN or DN improved the target conformity. The DCS may be of particular interest for sites with UN systems looking for a more economical solution than upgrading the nozzle to improve the target conformity of their spot scanning proton therapy system.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26936726      PMCID: PMC5360164          DOI: 10.1118/1.4942375

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  16 in total

1.  A novel technique for measuring the low-dose envelope of pencil-beam scanning spot profiles.

Authors:  Liyong Lin; Christopher G Ainsley; Thierry Mertens; Olivier De Wilde; Patrick T Talla; James E McDonough
Journal:  Phys Med Biol       Date:  2013-06-05       Impact factor: 3.609

2.  Experimental characterization of two-dimensional pencil beam scanning proton spot profiles.

Authors:  Liyong Lin; Christopher G Ainsley; James E McDonough
Journal:  Phys Med Biol       Date:  2013-08-16       Impact factor: 3.609

3.  A pencil beam algorithm for proton dose calculations.

Authors:  L Hong; M Goitein; M Bucciolini; R Comiskey; B Gottschalk; S Rosenthal; C Serago; M Urie
Journal:  Phys Med Biol       Date:  1996-08       Impact factor: 3.609

4.  Creating a spread-out Bragg peak in proton beams.

Authors:  David Jette; Weimin Chen
Journal:  Phys Med Biol       Date:  2011-05-10       Impact factor: 3.609

5.  Effects of spot size and spot spacing on lateral penumbra reduction when using a dynamic collimation system for spot scanning proton therapy.

Authors:  Daniel E Hyer; Patrick M Hill; Dongxu Wang; Blake R Smith; Ryan T Flynn
Journal:  Phys Med Biol       Date:  2014-10-21       Impact factor: 3.609

6.  Is there a single spot size and grid for intensity modulated proton therapy? Simulation of head and neck, prostate and mesothelioma cases.

Authors:  Lamberto Widesott; Antony J Lomax; Marco Schwarz
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

7.  A method for modeling laterally asymmetric proton beamlets resulting from collimation.

Authors:  Edgar Gelover; Dongxu Wang; Patrick M Hill; Ryan T Flynn; Mingcheng Gao; Steve Laub; Mark Pankuch; Daniel E Hyer
Journal:  Med Phys       Date:  2015-03       Impact factor: 4.071

8.  Monte Carlo study of the potential reduction in out-of-field dose using a patient-specific aperture in pencil beam scanning proton therapy.

Authors:  Stephen J Dowdell; Benjamin Clasie; Nicolas Depauw; Peter Metcalfe; Anatoly B Rosenfeld; Hanne M Kooy; Jacob B Flanz; Harald Paganetti
Journal:  Phys Med Biol       Date:  2012-04-19       Impact factor: 3.609

9.  Impact of spot size on plan quality of spot scanning proton radiosurgery for peripheral brain lesions.

Authors:  Dongxu Wang; Blake Dirksen; Daniel E Hyer; John M Buatti; Arshin Sheybani; Eric Dinges; Nicole Felderman; Mindi TenNapel; John E Bayouth; Ryan T Flynn
Journal:  Med Phys       Date:  2014-12       Impact factor: 4.071

10.  Theoretical Benefits of Dynamic Collimation in Pencil Beam Scanning Proton Therapy for Brain Tumors: Dosimetric and Radiobiological Metrics.

Authors:  Alexandra Moignier; Edgar Gelover; Dongxu Wang; Blake Smith; Ryan Flynn; Maura Kirk; Liyong Lin; Timothy Solberg; Alexander Lin; Daniel Hyer
Journal:  Int J Radiat Oncol Biol Phys       Date:  2015-10-08       Impact factor: 7.038

View more
  14 in total

1.  Technical Note: Optimization of spot and trimmer position during dynamically collimated proton therapy.

Authors:  Blake R Smith; Daniel E Hyer; Ryan T Flynn; Wesley S Culberson
Journal:  Med Phys       Date:  2019-03-05       Impact factor: 4.071

2.  Experimental and Monte Carlo characterization of a dynamic collimation system prototype for pencil beam scanning proton therapy.

Authors:  Blake R Smith; Mark Pankuch; Daniel E Hyer; Wesley S Culberson
Journal:  Med Phys       Date:  2020-09-09       Impact factor: 4.071

3.  Trimmer sequencing time minimization during dynamically collimated proton therapy using a colony of cooperating agents.

Authors:  Blake R Smith; Daniel E Hyer; Ryan T Flynn; Patrick M Hill; Wesley S Culberson
Journal:  Phys Med Biol       Date:  2019-10-21       Impact factor: 3.609

Review 4.  Empowering Intensity Modulated Proton Therapy Through Physics and Technology: An Overview.

Authors:  Radhe Mohan; Indra J Das; Clifton C Ling
Journal:  Int J Radiat Oncol Biol Phys       Date:  2017-10-01       Impact factor: 7.038

Review 5.  Neutron dose and its measurement in proton therapy-current State of Knowledge.

Authors:  Roger Antoine Hälg; Uwe Schneider
Journal:  Br J Radiol       Date:  2020-01-21       Impact factor: 3.039

6.  The dosimetric enhancement of GRID profiles using an external collimator in pencil beam scanning proton therapy.

Authors:  Blake R Smith; Nicholas P Nelson; Theodore J Geoghegan; Kaustubh A Patwardhan; Patrick M Hill; Jen Yu; Alonso N Gutiérrez; Bryan G Allen; Daniel E Hyer
Journal:  Med Phys       Date:  2022-02-21       Impact factor: 4.071

7.  Investigating aperture-based approximations to model a focused dynamic collimation system for pencil beam scanning proton therapy.

Authors:  Nicholas P Nelson; Wesley S Culberson; Daniel E Hyer; Blake R Smith; Ryan T Flynn; Patrick M Hill
Journal:  Biomed Phys Eng Express       Date:  2022-02-18

8.  Design of a focused collimator for proton therapy spot scanning using Monte Carlo methods.

Authors:  Theodore J Geoghegan; Nicholas P Nelson; Ryan T Flynn; Patrick M Hill; Suresh Rana; Daniel E Hyer
Journal:  Med Phys       Date:  2020-04-06       Impact factor: 4.071

9.  An investigation into the robustness of dynamically collimated proton therapy treatments.

Authors:  Blake R Smith; Daniel E Hyer; Wesley S Culberson
Journal:  Med Phys       Date:  2020-05-16       Impact factor: 4.071

10.  Innovations and the Use of Collimators in the Delivery of Pencil Beam Scanning Proton Therapy.

Authors:  Daniel E Hyer; Laura C Bennett; Theodore J Geoghegan; Martin Bues; Blake R Smith
Journal:  Int J Part Ther       Date:  2021-06-25
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.