| Literature DB >> 26924448 |
Mohammed S M Elbaz1, Rob J van der Geest1, Emmeline E Calkoen2, Albert de Roos3, Boudewijn P F Lelieveldt1,4, Arno A W Roest2, Jos J M Westenberg1.
Abstract
PURPOSE: To evaluate viscous energy loss and the association with three-dimensional (3D) vortex ring formation in left ventricular (LV) blood flow during diastolic filling. THEORY AND METHODS: Thirty healthy volunteers were compared with 32 patients with corrected atrioventricular septal defect as unnatural mitral valve morphology and inflow are common in these patients. 4DFlow MRI was acquired from which 3D vortex ring formation was identified in LV blood flow at peak early (E)-filling and late (A)-filling and characterized by its presence/absence, orientation, and position from the lateral wall. Viscous energy loss was computed over E-filling, A-filling, and complete diastole using the Navier-Stokes energy equations.Entities:
Keywords: 4DFlow MRI; atrioventricular septal defect; cardiac vortex flow; diastolic function; viscous energy loss
Mesh:
Year: 2016 PMID: 26924448 PMCID: PMC5297883 DOI: 10.1002/mrm.26129
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668
Figure 1a: Example of vortex ring isosurface (in green) at E‐peak filling phase as identified using Lambda2‐based vortex core detection 16 from a healthy volunteer. b: Streamlines superimposed on vortex ring isosurface (a) in a four‐chamber view. c: Definition of vortex ring orientation and normalized radial position relative to the LV long‐axis: The LV long‐axis is defined as the line from the mid of the mitral valvular opening to the LV apex. The normalized radial position was calculated as the distance of the vortex ring center (marked by “*”) to the LV long‐axis normalized by the basal endocardial radius. Vortex ring orientation was measured as the angle between the fitting plane of the vortex ring isosurface and the LV long‐axis (LA). d: Example vortex ring isosurface (in light green) at E‐peak filling phase from an AVSD‐repaired patient with an inclined E‐peak vortex ring orientation, showing a more elliptical vortex ring compared with controls. This deviated shape can be due to the restricted valve opening known to occur in AVSD patients after repair 17. e: Streamlines superimposed on vortex ring isosurface presented in (d) showing the inclined inflow and vortex ring orientation in this patient in a four‐chamber view.
Characteristics of Healthy Controls and Corrected‐AVSD Patientsa
| Controls | Corrected‐AVSD patients | |
|---|---|---|
| Age (years) | 23 [13–38] | 26 ± 12 |
| Male (%) | 14 (46) | 9 |
| Heart rate (bpm) | 68 [60–78] | 76 ± 13 |
| Diastasis (ms) | 116 ± 89 | 26 [0–67] |
| Stroke volume (mL) | 89 ± 23 | 85 ± 19 |
| Cardiac output (L/min) | 6.01 ± 1.3 | 6.39 ± 1.4 |
| EDV (mL) | 146 ± 42 | 155 ± 33 |
| Early filling fraction (%) | 76 ± 5 | 73 ± 13 |
| Ejection fraction (%) | 61 ± 5 | 56 [52–58] |
| E/A ratio peak flow rate | 2.5 ± 0.8 | 2.1 [1.7–2.6] |
| VFT index | 2.6 ± 0.6 (N = 30) | 2.4 [1.9–3.1] (N = 32) |
Normally distributed data are presented as mean ± SD, while non‐normally distributed data are presented as median [interquartile range].
excluding two cases without A‐peak.
indicates P < 0.01.
indicates P < 0.001.
EDV = end‐diastolic volume; VFT = vortex formation time.
Quantitative Analysis of Viscous Energy Loss and Kinetic Energy during LV Filling
| Controls (N = 30) | Patients (N = 32) |
| |
|---|---|---|---|
|
| 6.6 [5.5–9.0] | 12.2 ± 7.2 | <0.001 |
|
| 3.9 ± 2.2 | 8.1 ± 3.8a | <0.001 |
|
| 14.7 ± 4.6 | 18.3 [13.9–28.6] | 0.002 |
|
| 5.1 ± 2.8 | 10.4 ± 5.3 a | <0.001 |
|
| 3.4 ± 1.6 | 1.9 [1.3–2.9] a | 0.003 |
|
| 2.7 [2.3‐3.2] | 4.9 ± 2.0 | <0.001 |
|
| 0.24 [0.19–0.28] | 0.37 [0.26–0.53] | <0.001 |
|
| 25.1 ± 7.8 | 34.1 ± 18.8 | 0.03 |
|
| 9.3 [7.0–13.2] | 21.4 ± 11.2 a | <0.001 |
|
| 57.2 ± 16.6 | 55.2 [46.4–76.5] | 0.65 |
|
| 15.9 ± 8.0 | 28.9 ± 16.0 a | <0.001 |
|
| 5.0 ± 1.9 | 5.5 ± 2.7 | 0.73 |
|
| 1.3 ± 0.5 | 2.26 [1.3–3.2] a | <0.001 |
Two patients had no A‐wave and were excluded.
Figure 2Temporal evolution of viscous energy loss rate ( ), kinetic energy (KE), and inflow rate over LV diastole of a typical healthy subject (a) and a patient who did not present E‐vortex ring formation but rather a complex irregular flow instead (b). Viscous energy loss characterized by E‐ and A‐peaks is significantly elevated (with more than two‐fold increase) in the patient (b) compared with the healthy control subject (a).
Figure 3Correlation between average viscous energy loss rate during E‐filling (ELE‐avg) (a) and late filling (ELA‐avg) (b) with corresponding kinetic energy (KE), i.e., KEE‐avg and KEA‐avg, respectively, in controls (blue) and patients (red).
Quantitative Energy Parameters in Patients Presented with Vortex Ring Core vs. Those without Vortex Ring
| Patients with vortex ring (E‐filling: N = 26, A‐filling: N = 19) | Patients without vortex ring (E‐filling: N = 6, A‐filling: N = 13) |
| |
|---|---|---|---|
|
| 9.3 [7.5–12.4] | 21.0 ± 11.6 | 0.003 |
|
| 1.6 ± 1.1 | 2.2 ± 0.9 | 0.04 |
|
| 17 [13.8–22.5] | 41.8 ± 20.0 | 0.001 |
|
| 9.4 [6–13.3] | 12.1 ± 5.5 | 0.16 |
|
| 4.2 ± 1.3 | 7.60 ± 2.2 | 0.005 |
|
| 26.8 [22.8–34.9]d | 54.4 ± 31.8 | 0.03 |
|
| 19.0 ± 11.4 | 25.5 ± 10.1 | 0.13 |
|
| 57.8 ± 0.2d | 101.6 ± 59.2 | 0.05 |
|
| 24.9 [16.0–40.8] | 34.5 ± 16.2 | 0.07 |
P < 0.01 compared with controls.
Two patients had no A‐wave and were excluded.
P < 0.05 compared with controls.
P > 0.05 compared with controls.
Viscous Energy Loss in Corrected‐AVSD Patients with Normal or Abnormal E‐Peak Vortex Ring Orientation and/or Radial Position
| Phase | N | Vortex Orientation (degrees) |
|
|
| |
|---|---|---|---|---|---|---|
| Patients with Orientation within 95%CI | E‐filling | 9 | 71 ± 10 | 8.4 ± 1.8 | 8.3 ± 3.9 | 3.8 [2.6–4.0] |
| A‐filling | 10 | 71 ± 7 | 9.7 ± 2.6 | 6.7 ± 3.9 | 3.7 ± 0.8 | |
| Patients with Orientation below 95%CI | E‐filling | 14 | 37 ± 11 | 10.0 ± 3.2 | 8.3 ± 4.2 | 4.4 ± 1.30 |
| A‐filling | 9 | 35 ± 13 | 10.0 ± 3.7 | 8.3 ± 4.0 | 4.8 ± 1.1 | |
| Patients with Orientation above 95%CI | E‐filling | 3 | 118 ± 16 | 16.1 ± 5.2 | 8.1 ± 5.3 | 5.8 ± 1.6 |
| A‐filling | 0 | – | – | – |
P > 0.05.
P ≤ 0.01 when compared with controls.
0.01 < P < 0.05 when compared with controls.
Controls' Vortex Ring Parameters and Normal Limitsa
| Phase | Vortex ring parameter | 95%CI range | |||
|---|---|---|---|---|---|
| Lower limit (2.5%) | Upper limit (97.5%) | N (total=30) | Noutside | ||
| E‐filling peak | Orientation | 55° | 87° | 28 | 2 |
| Normalized Radial Position | 0.14 | 0.39 | 28 | 2 | |
| A‐filling peakb | Orientation | 57° | 84° | 25 | 2 |
| Normalized Radial Position | 0.05 | 0.39 | 25 | 2 | |
The 95% confidence interval (CI) represents the interval [2.5%–97.5%]. N is the number of subjects that presented a vortex ring core within the 95%CI range. Noutside is the number of subjects with detected vortex ring core but outside the 95%CI range. Radial position is normalized to the basal endocardial radius (measured on a short‐axis slice).
Three subjects did not present vortex ring core at peak late filling.
Figure 4Normal limits [95%CI (confidence interval)] of the total viscous energy loss integrated over complete diastole and normalized by stroke volume (ELdiastole) as derived from thirty healthy controls (in blue circles). Solid red horizontal line represents the 2.5% (lower limit) and dashed red horizontal line represents 97.5% (upper limit). Five (out of the six) patients who showed no vortex ring formation during E‐filling presented elevated viscous energy loss considerably beyond the upper limit of the healthy controls. Likewise, two (out of the three) patients who presented E‐peak vortex ring orientation above the 95%CI showed significant increase in viscous energy loss beyond the upper limit with the remaining patient approaching the upper limit.
Figure 5a: Correlation between total viscous energy loss integrated over diastole (ELdiastole) with vortex formation time (VFT) in controls (blue) and patients (red). b: Correlation between total viscous energy loss integrated over diastole but not normalized by stroke volume (abs_ELdiastole) with stroke volume in controls (blue) and patients (red).