AIMS: The longevity of generators is a crucial determinant of the cost-effectiveness of therapy with implantable cardioverter-defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-D). We evaluated the trend of device-measured residual battery capacity and longevity projections over 5-year follow-up. We also investigated possible factors associated with battery drain. METHODS ANDEntities:
Keywords: Battery; Cardiac resynchronization therapy; Implantable cardioverter-defibrillator; Longevity; Remote monitoring
Mesh:
Year: 2016 PMID: 26847076 PMCID: PMC5006961 DOI: 10.1093/europace/euv436
Source DB: PubMed Journal: Europace ISSN: 1099-5129 Impact factor: 5.214
Numbers of devices in analysis
| Device model | Battery capacity (Ah) | Devices in analysis (4851) |
|---|---|---|
| CRT-D | ||
| Cognis | 2.0 | 1260 |
| Energen | 1.9 | 279 |
| Punctua | 1.9 | 50 |
| Incepta | 1.9 | 436 |
| Dual-chamber ICD | ||
| Teligen | 1.7 | 774 |
| Energen | 1.8 | 231 |
| Punctua | 1.8 | 13 |
| Incepta | 1.8 | 271 |
| Single-chamber ICD | ||
| Teligen | 1.7 | 955 |
| Energen | 1.8 | 259 |
| Punctua | 1.8 | 25 |
| Incepta | 1.8 | 298 |
Factors associated with battery drain, measured as the year-to-year decrease in residual battery capacity (mAh)
| CRT-D | ICD-DR | ICD-VR | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Coefficient | SE | Coefficient | SE | Coefficient | SE | ||||
| Year | 48.06 | 1.75 | <0.001 | 8.37 | 1.09 | <0.001 | 6.56 | 0.73 | <0.001 |
| Right ventricular amplitude, V | −2.36 | 2.75 | 0.392 | 1.36 | 1.83 | 0.456 | 0.35 | 0.84 | 0.678 |
| Right ventricular width, ms | 19.64 | 7.80 | 0.012 | 1.54 | 4.87 | 0.752 | 3.10 | 3.61 | 0.391 |
| Right ventricular impedance, Ohm | – | – | – | – | – | – | – | – | – |
| Left ventricular amplitude, V | 11.80 | 1.41 | <0.001 | n.a. | n.a. | n.a. | n.a. | n.a. | n.a. |
| Left ventricular width, ms | 20.57 | 3.10 | <0.001 | n.a. | n.a. | n.a. | n.a. | n.a. | n.a. |
| Left ventricular impedance, Ohm | – | – | – | n.a. | n.a. | n.a. | n.a. | n.a. | n.a. |
| Right atrial amplitude, V | −3.22 | 2.70 | 0.234 | −1.10 | 1.85 | 0.553 | n.a. | n.a. | n.a. |
| Right atrial width, ms | 8.42 | 7.18 | 0.241 | 22.62 | 5.99 | <0.001 | n.a. | n.a. | n.a. |
| Right atrial impedance, Ohm | – | – | – | – | – | – | n.a. | n.a. | n.a. |
| Ventricular pacing, % | 0.20 | 0.07 | 0.004 | 0.17 | 0.04 | <0.001 | 0.00 | 0.07 | 0.981 |
| Atrial pacing, % | 0.09 | 0.07 | 0.222 | 0.11 | 0.03 | <0.001 | n.a. | n.a. | n.a. |
| Rate response mode, ON | 1.96 | 3.90 | 0.616 | 4.05 | 1.99 | 0.042 | −0.02 | 5.92 | 0.997 |
| Heart rate, bpm | 0.35 | 0.14 | 0.017 | 0.27 | 0.09 | 0.002 | 0.04 | 0.05 | 0.469 |
| Shock delivered, | 2.65 | 0.34 | <0.001 | 2.34 | 0.30 | <0.001 | 2.54 | 0.16 | <0.001 |
| Shock diverted, | 1.74 | 0.84 | 0.039 | 1.02 | 0.35 | 0.004 | 0.86 | 0.31 | 0.006 |