Simon von Gunten1, Beat A Schaer2, Sing-Chien Yap3, Tamas Szili-Torok3, Michael Kühne1, Christian Sticherling1, Stefan Osswald1, Dominic A M J Theuns3.
Abstract
AIMS: Longevity of implantable cardioverter defibrillators (ICDs) is crucial for patients and healthcare systems as replacements impact on infection rates and cost-effectiveness. Aim was to determine longevity using very large databases of two teaching hospitals with a high number of replacements and a rather homogeneous distribution among manufacturers. METHODS ANDEntities:
Keywords: Cardiac resynchronization therapy; Implantable cardioverter-defibrillator; Longevity; Survival
Mesh:
Year: 2015 PMID: 26609076 PMCID: PMC4880113 DOI: 10.1093/europace/euv296
Source DB: PubMed Journal: Europace ISSN: 1099-5129 Impact factor: 5.214
Baseline characteristics of the 3436 patients
| Male gender | 2721 (79%) |
| Age (mean ± SD) | 59 ± 14 |
| Follow-up (mean ± SD) | 60.3 ± 44.4 |
| Ejection fraction (mean ± SD) | 32 ± 13 |
| QRS width (mean ± SD) | 127 ± 35 |
| Primary prevention | 1975 (57%) |
| Ischaemic cardiomyopathy | 2525 (74%) |
| Myocardial infarction | 1825 (53%) |
| Percutaneous coronary intervention | 1069 (31%) |
| Coronary artery bypass | 769 (22%) |
| Diabetes mellitus | 694 (20%) |
| Creatinine level (μmol/L, mean ± SD) | 103 ± 69 |
| NYHA Class I | 792 (23%) |
| NYHA Class II | 1680 (49%) |
| NYHA Class III | 907 (27%) |
| NYHA Class IV | 39 (1%) |
| ACE/ARB therapy | 2640 (77%) |
| Beta-blocker therapy | 2621 (76%) |
| Diuretic therapy | 2029 (59%) |
| Statin therapy | 1978 (58%) |
| Amiodarone therapy | 685 (20%) |
| Digoxin therapy | 461 (13%) |
SD, standard deviation; ACE, angiotensin converting enzyme inhibitor; ARB, angiotensin receptor blocker.
Manufacturers, ICD, pacing modes (upper row) and replaced ICDs (lower row), shown in alphabetical order
| Manufacturer | VVI | DDD | CRT | Total | % Replaced |
|---|---|---|---|---|---|
| Biotronik | 645 | 346 | 228 | 1219 | |
| 149 | 126 | 23 | 298 | 24.5% | |
| Boston Scientific | 413 | 275 | 259 | 947 | |
| 118 | 113 | 80 | 311 | 31.8% | |
| Intermedics | 21 | 0 | 0 | 21 | |
| 19 | 0 | 0 | 19 | 90.5% | |
| Medtronic | 449 | 182 | 267 | 898 | |
| 181 | 78 | 97 | 356 | 39.6% | |
| St. Jude Medical | 625 | 388 | 526 | 1539 | |
| 94 | 85 | 116 | 295 | 19.2% | |
| Sorin ELA | 10 | 144 | 4 | 158 | |
| 1 | 48 | 0 | 49 | 31.0% | |
| Total | |||||
| Devices | 2158 | 1340 | 1284 | 4782 | |
| Replaced | 557 | 455 | 316 | 1328 | |
| % Replaced | 25.8% | 34.0% | 24.6% | 27.8% | |
| Cameron Health | 99 | 99 | |||
| 11 | 11 | 11.1% | |||
ICD, implantable cardioverter defibrillator; VVI, single-chamber ICD; DDD, dual-chamber ICD; CRT, cardiac resynchronization therapy ICD.
Comparison of longevity of devices implanted until December 2005 and thereafter (highlighted is the best performance in the corresponding group) according to the manufacturer and pacing mode
| Before 2006 | Thereafter | |||
|---|---|---|---|---|
| 5-year longevity (%) | 6-year longevity (%) | 5-year longevity (%) | 6-year longevity (%) | |
| All ICD models | ||||
| All manufacturers** | 63.9 | 44.9 | 80.6 | 61.6 |
| Biotronik** | 44.0 | 10.5 | 81.4 | 42.1 |
| Boston** | 65.1 | 45.7 |
|
|
| Medtronic|| |
|
| 85.8 | 72.6 |
| St. Jude Medical** | 64.3 | 49.8 | 74.1 | 60.7 |
| Sorin|| | 59.8 | 27.8 | 77.5 | 77.5 |
| Intermedics | 0 | 0 | n.a. | n.a. |
| Cameron Health | n.a. | n.a. | 47.9 | n.a. |
| VVI | ||||
| All manufacturers** | 73.7 | 56.4 | 92.1 | 76.0 |
| Biotronik** | 59.8 | 15.2 | 89.1 | 45.6 |
| Boston** | 74.3 | 53.3 |
|
|
| Medtronic|| |
|
| 91.7 | 85.9 |
| St. Jude Medical** | 70.9 | 60.1 | 94.3 | 92.6 |
| Sorin | n.a. | n.a. | 80.0 | 80.0 |
| Intermedics | 0 | 0 | n.a. | n.a. |
| Cameron Health | n.a. | n.a. | 47.9 | n.a. |
| DDD | ||||
| All manufacturers** | 58.2 | 40.8 | 76.1 | 50.9 |
| Biotronik** | 26.6 | 4.2 | 60.0 | 26.3 |
| Boston# | 65.8 | 52.2 |
|
|
| Medtronic|| |
|
| 89.3 | 76.5 |
| St. Jude Medical|| | 54.7 | 46.0 | 78.7 | 35.3 |
| Sorin|| | 59.8 | 27.8 | n.a. | n.a. |
| CRT | ||||
| All manufacturers** | 47.1 | 21.2 | 66.3 | 43.0 |
| Biotronik** | 0 | 0 | 76.2 | 44.9 |
| Boston** | 43.5 | 17.5 |
|
|
| Medtronic** | 39.2 | 7.4 | 74.1 | 46.3 |
| St. Jude Medical|| |
|
| 45.3 | 26.5 |
n.a., not applicable, i.e. not manufactured in this period, not implanted in the two hospitals, or time point not reached; ICD, implantable cardioverter defibrillator; VVI, single-chamber ICD; DDD, dual-chamber ICD; CRT, cardiac resynchronization therapy ICD.
|| P = n.s., #P ≤ 0.05, **P ≤ 0.001.