Literature DB >> 25976906

Longevity of implantable cardioverter-defibrillators for cardiac resynchronization therapy in current clinical practice: an analysis according to influencing factors, device generation, and manufacturer.

Maurizio Landolina1, Antonio Curnis2, Giovanni Morani3, Antonello Vado4, Ernesto Ammendola5, Antonio D'onofrio6, Giuseppe Stabile7, Martino Crosato8, Barbara Petracci9, Carlo Ceriotti10, Luca Bontempi2, Martina Morosato3, Gian Paolo Ballari4, Maurizio Gasparini10.   

Abstract

AIMS: Device replacement at the time of battery depletion of implantable cardioverter-defibrillators (ICDs) may carry a considerable risk of complications and engenders costs for healthcare systems. Therefore, ICD device longevity is extremely important both from a clinical and economic standpoint. Cardiac resynchronization therapy defibrillators (CRT-D) battery longevity is shorter than ICDs. We determined the rate of replacements for battery depletion and we identified possible determinants of early depletion in a series of patients who had undergone implantation of CRT-D devices. METHODS AND
RESULTS: We retrieved data on 1726 consecutive CRT-D systems implanted from January 2008 to March 2010 in nine centres. Five years after a successful CRT-D implantation procedure, 46% of devices were replaced due to battery depletion. The time to device replacement for battery depletion differed considerably among currently available CRT-D systems from different manufacturers, with rates of batteries still in service at 5 years ranging from 52 to 88% (log-rank test, P < 0.001). Left ventricular lead output and unipolar pacing configuration were independent determinants of early depletion [hazard ratio (HR): 1.96; 95% 95% confidence interval (CI): 1.57-2.46; P < 0.001 and HR: 1.58, 95% CI: 1.25-2.01; P < 0.001, respectively]. The implantation of a recent-generation device (HR: 0.57; 95% CI: 0.45-0.72; P < 0.001), the battery chemistry and the CRT-D manufacturer (HR: 0.64; 95% CI: 0.47-0.89; P = 0.008) were additional factors associated with replacement for battery depletion.
CONCLUSION: The device longevity at 5 years was 54%. High left ventricular lead output and unipolar pacing configuration were associated with early battery depletion, while recent-generation CRT-Ds displayed better longevity. Significant differences emerged among currently available CRT-D systems from different manufacturers.
© The Author 2015. Published by Oxford University Press on behalf of the European Society of Cardiology.

Entities:  

Keywords:  Battery; CRT; Implantable defibrillator; Longevity

Mesh:

Year:  2015        PMID: 25976906      PMCID: PMC4535557          DOI: 10.1093/europace/euv109

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


This is the largest multicentre analysis of consecutive patients who had undergone implantation of a recent CRT-D device from all manufacturers. This analysis provides real-world industry-independent longevity data of currently available CRT-D devices and identifies the possible determinants of early battery depletion.

Introduction

Implantable cardioverter-defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-D) have been shown to be cost-effective therapeutic options,[1] and have become standard treatment for the prevention of sudden cardiac death and the management of selected heart failure patients.[2] Patients who receive ICD and CRT-D devices must undergo device replacement at the time of battery depletion or in the case of other device-related or clinical events. Compared with ICDs, CRT-Ds are exposed to an increased risk of complications (lead dislodgment, infection, diaphragmatic stimulation, etc.), and are the most demanding antiarrhythmic devices in terms of battery consumption because of the need for continuous biventricular pacing and because pacing thresholds tend to be higher in the left ventricle.[3] Since device replacement is associated with a notable risk of complications[4] and engenders costs for healthcare systems, device lifespan is a crucial determinant of the cost-effectiveness of therapy. Projection on battery longevity provided by manufacturers is based on intensive laboratory testing under controlled conditions and might be different from device longevity in real life. Other information provided by the industry comes from product performance reports, and is based on the analysis of returned devices. However, cases of premature battery depletion may be under-reported. Moreover, events such as infection, removal for system upgrade, heart transplantation, or patient death are not measured, and are estimated in analyses only by applying statistical adjustments and correction factors so as not to overstate the number of devices in service. Industry-independent analyses on longevity are scarce and have generally been performed on single-centre series that do not include currently available ICD models.[5-9] The aim of the present study was to measure the rate of replacements for battery depletion and to identify possible determinants of early depletion in a population of consecutive patients who had undergone implantation of CRT-D devices from different manufacturers.

Methods

Patient population and study design

Data on all patients who had received a CRT-D system according to international recommendations were prospectively collected in the hospital databases of nine Italian implanting centres. At the time of implantation, all patients signed a written informed consent for data storage and analysis. In March 2014, data on all consecutive CRT-D systems implanted from January 2008 to March 2010 were retrieved for analysis, in order to estimate the proportion of devices in service more than 4 years after implantation. All patients in the present analysis underwent implantation of a CRT-D, by means of standard techniques. After implantation, patients returned for regular clinic visits every 6 months and no remote monitoring system was adopted. In all patients, the pacing output was programmed to ensure myocardial capture while avoiding phrenic nerve stimulation. Usually, voltage outputs were programmed to twice the threshold voltage assessed at pre-discharge and at periodical in-office visits. In all devices with algorithms for automatic threshold-capture determination and pacing output adjustment, the feature was programmed on. Baseline data included the date of implantation, the manufacturer and model of the device and all leads implanted. Follow-up data included the pacing output and percentage pacing at the time of the last visit and the total number of shocks delivered. In the present analysis, the study database was searched for all device-related events resulting in surgical intervention for device replacement. If pulse generator removal was required, the reason was verified and categorized as due to normal battery depletion or other causes. The endpoint of this analysis was the rate of replacements for battery depletion. The service life of the device was defined as the time from implantation to surgical replacement. Patients were censored at the time of death or the last outpatient follow-up visit. In the analysis of the time to battery depletion, removals for other causes were not counted as events and patients were censored at the time of their occurrence.

Statistical analysis

Descriptive statistics are reported as means ± SD for normally distributed continuous variables, or medians with 25th to 75th percentiles in the case of skewed distribution. Categorical variables are reported as percentages. Differences in proportions were compared by applying χ2 analysis or Fisher's exact test, as appropriate. The pacing output was described in terms of pulse amplitude (<2.5 V, from 2.5 to 4 V, more than 4 V) and duration (<0.5 ms, from 0.5 to 1.0 ms, more than 1.0 ms). Biventricular pacing was quantified as: <90%, from 90 to 95%, or >95%. Similarly, the total number of delivered shocks was divided into groups (0, from 1 to 10, from 10 to 20, >20). Kaplan–Meier analysis was used to analyse device longevity, and the log-rank test was applied to evaluate differences between trends (level of significance adjusted for multiple testing by Bonferroni correction). As differences in terms of service life were expected, the population was stratified by manufacturer and device generation. In the analysis of device generation, defibrillators from each manufacturer were divided into recent- and earlier-generation. We identified as recent-generation the most recent device families released onto the market (for the most part after 2007), and as earlier-generation all devices belonging to previous device families. Hazard ratios (HRs) and their 95% confidence intervals (CIs) were computed by means of Cox regression models, in which device data were considered as fixed covariates and pulse generator replacements were considered as time-dependent covariates. After checking for collinearity, we included in the multivariate Cox models any variable with a P–value of <0.05 on univariate analysis. A P-value of <0.05 was considered significant for all tests. All statistical analyses were performed by means of STATISTICA software, version 7.1 (StatSoft, Inc., Tulsa, OK, USA).

Results

Study population

From January 2008 to March 2010, a total of 1726 heart failure patients received a CRT-D at the nine study centres. Details of the devices implanted are summarized in Table . The devices were from five manufacturers: 49 (3%) from Biotronik, 608 (35%) from Boston Scientific, 798 (46%) from Medtronic, 99 (6%) from Sorin, and 172 (10%) from St Jude Medical. They belonged to defibrillator families released onto the market from 2003 to 2010, and were equipped with different battery types. The devices of earlier-generation (released before 2007) and recent-generation families (released since 2007) were 708 and 1018, respectively. Details (as reported in the device manuals) and numbers of devices in analysis Li/MnO2, lithium manganese dioxide; Li/SVO, lithium silver-vanadium oxide; Li/CFX-SVO, silver-vanadium oxide and carbon monofluoride. aConsidered as ‘Recent-generation’ for the purpose of the analysis. bNot reported in the device manual. The procedure was a de novo implantation in 1071 (62%) patients, a replacement of a previous CRT-D system in 472 (27%) patients, and an upgrade from a previous dual-chamber ICD in the remaining 183 (11%) patients. Two hundred and thirty-three (13%) patients were in atrial fibrillation and did not receive an atrial lead. A transvenous unipolar left ventricular lead was adopted in 385 (22%) systems, a transvenous bipolar lead in 1328 (77%) systems, and an epicardial unipolar lead in 13 (1%) systems. Unipolar leads were evenly distributed between the earlier-generation (155, 22%) and recent-generation groups (230, 23%). The right ventricular lead had an integrated bipolar design in 779 (45%) systems and a true-bipolar design in the remaining 947 (55%) systems. All devices were programmed to deliver true biventricular pacing (right and left ventricular pacing).

Follow-up

During a median follow-up of 43 months (25th to 75th percentiles, 18–53; total follow-up, 5201 person-years), 479 (28%) devices were replaced/removed for any cause (Table ). Specifically, 401 (23%) were replaced because of battery depletion, 40 CRT-Ds were removed because of device-related infection, and 7 devices were removed at the time of heart transplantation. Moreover, 31 pulse generators were replaced at the time of lead failure or elective replacement of a non-malfunctioning safety advisory lead. Length of observation period and number (proportion) of patients with reported events During follow-up, 274 (16%) patients died before device replacement and 146 (8%) patients, evenly distributed among device manufacturers, did not complete the study period because they chose to continue follow-up in another device clinic and were therefore censored from the analysis. The Kaplan–Meier estimate of time to device replacement for battery depletion in the overall study population is reported in Figure A. The actuarial probability of survival free from battery depletion was 81% at 4 years and 54% at 5 years. Kaplan–Meier estimates of survival from device replacement for battery depletion in the overall study population (A) and for battery depletion stratified by device manufacturer (B). The Kaplan–Meier estimates of time to device replacement for battery depletion, stratified by device manufacturer, demonstrated considerable differences in system longevity (overall log-rank test, P < 0.001; Figure B). Specifically, the actuarial rate of batteries still in service at 5 years ranged from 42% for Medtronic CRT-D to 66% for Boston Scientific generators. The comparison of battery longevity between generations revealed significant improvements in recent devices from all manufacturers (log-rank test, P < 0.02 for all comparisons; Figure ), except for Biotronik devices, ostensibly because of the small size of the subgroups in analysis. Kaplan–Meier estimates of survival from device replacement for battery depletion, stratified by generation and by device manufacturer. Since longevity estimates may be imprecise when the sample is small, the comparison of longevity among recent-generation CRT-D from different manufacturers was performed only for subgroups with at least 100 devices in analysis; the results are reported in Figure . The rate of batteries still in service at 5 years was 52% for Medtronic, 75% for St Jude Medical, and 88% for Boston Scientific generators (log-rank test, P < 0.01 for pairwise comparisons). Kaplan–Meier estimates of survival from device replacement for battery depletion for recent-generation devices, stratified by manufacturer.

Predictors of battery depletion

Table shows data on pacing output and the percentage of biventricular pacing at the time of the last visit and the total number of shocks delivered. There were significant differences among manufacturers in terms of programmed pacing output. Specifically, the proportion of patients with left ventricular pulse amplitude <2.5 V and duration <0.5 ms was higher in the Medtronic group. Similarly, patients with Medtronic devices were less likely to receive a shock. Pacing output, percentage of biventricular pacing, and total number of shocks delivered, by device manufacturer High pacing output: pulse amplitude >2.5 V and duration >0.5 ms. *P<0.05 vs. Medtronic. The device replacement procedure for battery depletion was independent of the percentage of biventricular pacing and the burden of defibrillator therapy (Table ). However, it showed a significant association with the left ventricular lead output and with a unipolar pacing configuration. Moreover, the implantation of a recent-generation device, equipped with lithium manganese dioxide or hybrid silver-vanadium oxide, and carbon monofluoride batteries and a CRT-D from Boston Scientific were independent protective factors against early battery depletion. The multivariate analysis limited to recent-generation devices, confirmed following variables as independent factors associated with battery depletion: the left ventricular lead output (HR: 3.09, 95% CI: 2.18–4.38; P < 0.001) and the Boston Scientific CRT-Ds (HR: 0.13, 95% CI: 0.04–0.40; P < 0.001). Univariate and multivariate analysis of factors associated with replacement for battery depletion in the overall population Li/MnO2, lithium manganese dioxide; Li/SVO, lithium silver-vanadium oxide; Li/CFX-SVO, silver-vanadium oxide and carbon monofluoride. aPulse amplitude >2.5 V and duration >0.5 ms.

Discussion

Our analysis showed that 5 years after a successful CRT-D implantation procedure, 46% of devices were replaced on account of battery depletion, and left ventricular lead output and unipolar pacing configuration turned out to be independent determinants of early depletion. Recent-generation CRT-D (mostly released onto the market after 2007) displayed significantly greater longevity than those of early generations. Nonetheless, large differences in longevity seem to exist among currently available CRT-D systems from different manufacturers. This study constitutes the largest multicentre analysis of consecutive patients who had undergone implantation of a recent CRT-D device from all manufacturers, and aimed to measure the rate of replacement for battery depletion. The devices in analysis constituted a representative sample (1726/8670, 20%) of all CRT-D implantation procedures performed in Italy during the observation period.[10,11] In the present study, we were able to obtain data on patients who had undergone implantation more than 4 years earlier. Having noticed that a substantial proportion of defibrillator recipients outlive their first device, Hauser suggested that technological research should be aimed at providing a defibrillator that lasts a lifetime.[12] Although this goal could seem more achievable for CRT-D systems, because of the short life expectancy of their recipients,[13] the extension of CRT recommendations to mildly symptomatic patients[2] has made it more difficult to narrow the mismatch between the longevity of CRT-D and patient survival. Longevity of defibrillators is crucial on account of the major drawbacks associated with replacement procedures,[3,4] even from the viewpoint of the patients.[14] From the perspective of public health systems, system revisions are a major source of incremental costs, and extending system longevity has been shown to significantly improve cost-effectiveness estimates.[15,16] Our analysis revealed large discrepancies in device lifespan in a cohort of patients with CRT-D systems from different manufacturers. In previous studies comparing longevity of ICDs (the majority produced before 2007) across manufacturers,[5-8,16] the authors unanimously reported better longevity for the Medtronic devices included in their analyses and available at that time on the market. In contrast, on including CRT-D devices released onto the market from 2003 to 2010 and still available today, we observed shorter longevity in Medtronic CRT-D. Our findings confirmed recent data from the single-centre study by Alam et al.[9] Device longevity is based on battery technology, the efficiency of the electronic circuitry and the availability of specific algorithms for automatic pacing output adjustment, and varies not only among manufacturers but also among generations of devices from the same manufacturer. Indeed, on comparing the longevity of CRT-D belonging to different generations, we found significant improvements in recent devices. This result confirms previous findings by Schaer et al.[6] and Thijssen et al.[7] In contrast, Horlbeck et al.[8] did not observe better longevity in newer defibrillators. However, while in previous studies analysis by device generation was based on the time of implantation (i.e. before 2002 or since 2002), we stratified defibrillators according to the time of release onto the market (before or since 2007), in order to account for possible delays in the adoption of new technology in clinical practice. Owing to these discrepancies in the performance of different CRT-D generations, comparison among manufacturers might be influenced by the mix of models in analysis. Nonetheless, our direct comparison of recent generations of CRT-D and the multivariate analysis limited to recent-generation devices confirmed the significantly longer lifespan of Boston Scientific defibrillators. The recent-generation Boston Scientific defibrillators considered were equipped with high-capacity lithium manganese dioxide batteries, which resulted in a markedly longer lifespan than that of previous generations. In contrast, the Medtronic and St Jude Medical CRT-Ds released onto the market after 2007 and included in the present analysis were equipped with the same lithium silver-vanadium oxide battery used in early generations. However, although the chemistry was the same, St Jude Medical adopted different battery models for its recent CRT-Ds. This, together with a possible improved efficiency of the electronic circuitry, may explain the improved performance of its recent-generation devices. The majority of the recent Medtronic CRT-Ds was endowed with algorithms for automatic pacing output adjustment, which have been demonstrated to reduce pacing output in comparison with the standard manual management approach.[17] This may have determined the observed improvement in longevity. The importance of pacing output optimization is supported by the significant positive association that we noticed between left ventricular lead output and early depletion, confirming previous findings by Thijssen et al.[7] and Alam et al.[9] This can easily be explained by the considerable battery drainage caused by the high pulse amplitude and duration required for consistent capture of the left ventricle, and by the need for continuous biventricular pacing. Moreover, the higher battery drainage with unipolar left ventricular pacing configurations can be ascribed to the lower pacing impedance and thus to the higher current consumption.[18] As previously reported by Alam et al.,[9] despite longer service life, Boston Scientific CRT-Ds had higher programmed pacing output, most probably owing to the absence of algorithms for automatic pacing output adjustment. Moreover, Boston Scientific CRT-Ds were most likely to deliver a shock, ostensibly because of the longer lifespan of each device and thus the longer exposure to the risk of arrhythmias. Nonetheless, our results suggest that battery depletion was independent of the burden of defibrillator therapy, confirming previous findings.[5-7,9] A more relevant factor might be the capacitor reformation interval,[6] but we did not include this parameter in our model. In addition, the percentage of biventricular pacing was not responsible for the observed longevity differences in CRT-D devices. This finding, specifically attributable to the fact that the percentage of pacing with CRT-D is generally close to 100%, might not apply to single- and dual-chamber defibrillators, which display higher variability in pacing burden.[6-8]

Limitations

Some limitations of this study should be noted. First, the baseline clinical characteristics of the patients were not analysed. However, the results should have not been affected by any bias, as previous studies had failed to show any association between clinical characteristics and defibrillator lifespans.[8] Secondly, it should be mentioned that not all pacing parameters were included in analysis (e.g. rate response) and data on pacing output and the percentage of biventricular pacing were only gathered at the last interrogation of the device and not throughout the entire period of its functioning. However, we assumed that these values could represent reliable surrogates of measurements over the lifespan of the device. Thirdly, although the rates of events were statistically different among the manufacturers, the proportion of replacements among devices displaying a longer lifespan was low. Thus, longer follow-up periods are warranted in order to estimate the actual lifespan of these devices. Fourthly, in this study no information was available regarding the time from device production to implantation. However, the results should have not been affected by any bias, as all devices had comparable shelf life and were implanted before the ‘use by’ date, in accordance with the manufacturer's instructions. Finally, although the majority of devices in the present analysis is still available today, newer devices have been released and our results may not apply to these. For example, the more recent Medtronic and St Jude Medical devices are equipped with hybrid silver-vanadium oxide and carbon monofluoride batteries, and St Jude Medical and Boston Scientific defibrillators are now endowed with algorithms for automatic pacing output adjustment. Moreover, new pacing algorithms, as the novel Medtronic adaptive CRT,[19] may improve the longevity of the device by delivering single-chamber left ventricular pacing.

Conclusions

In a large population of consecutive patients with CRT-D, the device longevity at 5 years was 54%. Recent-generation CRT-Ds displayed better longevity than those of earlier generations. Moreover, differences emerged among currently available CRT-D systems from different manufacturers, with rates of batteries still in service at 5 years ranging from 52 to 88%. According to our results the next step, already undertaken by manufacturers, towards the goal of providing a lifetime CRT-D system should be a device with latest-generation high-capacity batteries and with algorithms for the optimization of left ventricular pacing.

Funding

Funding to pay the Open Access publication charges for this article was provided by Boston Scientific Italia. Conflict of interest: This was an independent study. No external funding was achieved for this project. M.L. has a speakers' bureau appointment with St Jude Medical, Medtronic, and Boston Scientific and an advisory board relationship with St Jude Medical and Medtronic. M.G. has an advisory board relationship with Boston Scientific. The other authors report no conflicts.
Table 1

Details (as reported in the device manuals) and numbers of devices in analysis

ManufacturerDevice familyYear of market release (CE mark)Battery
Devices in analysis
ManufacturerModelChemistryCapacity
Biotronik, n = 49 (3%)Lumax 3002006LitronikLiS 3182 (K7/R6)Li/MnO21.28Ah3
GreatbatchGB 2491Li/CFx-SVO1.72Ah
Lumax 3402006LitronikLiS 3182 (K7/R6)Li/MnO21.28Ah26
GreatbatchGB 2491Li/CFx-SVO1.72Ah
Lumax 540a2008LitronikLiS 3192 R7Li/MnO21.72Ah20
GreatbatchGB 2491Li/CFx-SVO1.72Ah
Boston Scientific, n = 608 (35%)Renewal2004Greatbatch2000Li/SVO2.00Ah288
Livian2007Greatbatch2500Li/CFx-SVO1.86Ah29
Cognisa2008Boston Scientific401988Li/MnO21.84Ah291
Medtronic, n = 798 (46%)InSync III Marquis2003Medtronic161253Li/SVO0.90Ah67
InSync Sentry2004Medtronic161253Li/SVO0.89Ah7
InSync Maximo2005Medtronic161253Li/SVO0.89Ah21
Concerto2006Medtronic161455Li/SVO1.00Ah171
Consultaa2008Medtronic161455Li/SVO1.00Ah447
Maximo IIa2008Medtronic161455Li/SVO1.00Ah69
Protectaa2010Medtronic161455Li/SVO1.00Ah16
Sorin, n = 99 (6%)Ovatio2005GreatbatchWGL 2393Li/SVO0.87Ah30
Paradyma2008GreatbatchGB 2593Li/CFx-SVO1.96Ah69
St Jude Medical, n = 172 (10%)Atlas2003Greatbatch2255Li/SVOb40
Greatbatch2150Li/SVOb
Epic2006Greatbatch2150Li/SVOb26
Promotea2007Greatbatch2555Li/SVOb106
Greatbatch2356Li/SVOb

Li/MnO2, lithium manganese dioxide; Li/SVO, lithium silver-vanadium oxide; Li/CFX-SVO, silver-vanadium oxide and carbon monofluoride.

aConsidered as ‘Recent-generation’ for the purpose of the analysis.

bNot reported in the device manual.

Table 2

Length of observation period and number (proportion) of patients with reported events

nTotal follow-up, person-yearsMedian follow-up [25th–75th percentiles], monthsAll-cause replacements, patients (%)Battery depletion, patients (%)Death, patients (%)Incomplete follow-up, patients (%)
Biotronik4917149 [34–56]10 (20)10 (20)6 (12)2 (4)
Boston Scientific608186644 [16–55]132 (22)109 (18)112 (18)56 (9)
Medtronic798232841 [18–52]274 (34)228 (29)115 (14)64 (8)
Sorin9929947 [18–54]22 (22)20 (20)14 (14)8 (8)
St Jude Medical17253744 [18–55]41 (24)34 (20)27 (16)16 (9)
Table 3

Pacing output, percentage of biventricular pacing, and total number of shocks delivered, by device manufacturer

BiotronikBoston ScientificMedtronicSorinSt Jude Medical
Patients with high right atrial lead output (%)05641
Patients with high right ventricular lead output (%)25415
Patients with high left ventricular lead output (%)1331*1831*26*
% of patients with biventricular pacing: <90%/90–95%/≥95%10/13/7712/10/779/14/7616/18/6511/13/75
% of patients with shocks delivered: 0/1–10/10–20/>2064/36/0/0*69/29/1/1*85/13/0/183/17/0/074/26/0/0*

High pacing output: pulse amplitude >2.5 V and duration >0.5 ms.

*P<0.05 vs. Medtronic.

Table 4

Univariate and multivariate analysis of factors associated with replacement for battery depletion in the overall population

Univariable analysis
Multivariable analysis
HR95% CIP-ValueHR95% CIP-Value
Biotronik0.750.40–1.410.369
Boston Scientific0.540.43–0.67<0.0010.640.47–0.890.008
Medtronic1.001.00
Sorin0.830.53–1.300.415
St Jude Medical0.740.52–1.050.089
Recent generation0.500.40–0.61<0.0010.570.45–0.72<0.001
Battery chemistry: Li/SVO1.001.00
Battery chemistry: Li/CFx-SVO0.420.24–0.720.0020.280.16–0.50<0.001
Battery chemistry: Li/MnO20.200.13–0.33<0.0010.370.22–0.64<0.001
High right atrial lead outputa0.700.39–1.240.219
High right ventricular lead outputa1.380.83–2.310.217
High left ventricular lead outputa1.741.39–2.18<0.0011.961.57–2.46<0.001
Unipolar left ventricular lead1.711.37–2.13<0.0011.581.25–2.01<0.001
True-bipolar right ventricular lead1.471.21–1.79<0.0011.000.78–1.300.978
Percentage of biventricular pacing1.200.91–1.580.207
Shocks delivered1.580.59–4.200.365

Li/MnO2, lithium manganese dioxide; Li/SVO, lithium silver-vanadium oxide; Li/CFX-SVO, silver-vanadium oxide and carbon monofluoride.

aPulse amplitude >2.5 V and duration >0.5 ms.

  17 in total

1.  Implantable cardioverter-defibrillator longevity under clinical circumstances: an analysis according to device type, generation, and manufacturer.

Authors:  Joep Thijssen; C Jan Willem Borleffs; Johannes B van Rees; SumChe Man; Mihály K de Bie; Jeroen Venlet; Enno T van der Velde; Lieselot van Erven; Martin J Schalij
Journal:  Heart Rhythm       Date:  2011-11-15       Impact factor: 6.343

2.  The growing mismatch between patient longevity and the service life of implantable cardioverter-defibrillators.

Authors:  Robert G Hauser
Journal:  J Am Coll Cardiol       Date:  2005-06-21       Impact factor: 24.094

3.  2013 ESC guidelines on cardiac pacing and cardiac resynchronization therapy: the task force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA).

Authors:  Michele Brignole; Angelo Auricchio; Gonzalo Baron-Esquivias; Pierre Bordachar; Giuseppe Boriani; Ole-A Breithardt; John Cleland; Jean-Claude Deharo; Victoria Delgado; Perry M Elliott; Bulent Gorenek; Carsten W Israel; Christophe Leclercq; Cecilia Linde; Lluís Mont; Luigi Padeletti; Richard Sutton; Panos E Vardas
Journal:  Europace       Date:  2013-06-24       Impact factor: 5.214

4.  Cost-effectiveness of implantable cardioverter-defibrillators.

Authors:  Gillian D Sanders; Mark A Hlatky; Douglas K Owens
Journal:  N Engl J Med       Date:  2005-10-06       Impact factor: 91.245

5.  Real-world data on the lifespan of implantable cardioverter-defibrillators depending on manufacturers and the amount of ventricular pacing.

Authors:  Fritz W Horlbeck; Fritz Mellert; Jens Kreuz; Georg Nickenig; Joerg O Schwab
Journal:  J Cardiovasc Electrophysiol       Date:  2012-08-21

6.  Long-term performance of coronary sinus leads used for cardiac resynchronization therapy.

Authors:  Assunta Iuliano; Gergana Shopova; Antonio De Simone; Francesco Solimene; Pietro Turco; Natale Marrazzo; Vincenzo La Rocca; Carmine Ciardiello; Marco Agrusta; Giuseppe Stabile
Journal:  Pacing Clin Electrophysiol       Date:  2007-01       Impact factor: 1.976

7.  Battery longevity in cardiac resynchronization therapy implantable cardioverter defibrillators.

Authors:  Mian Bilal Alam; Muhammad Bilal Munir; Rohit Rattan; Susan Flanigan; Evan Adelstein; Sandeep Jain; Samir Saba
Journal:  Europace       Date:  2013-10-06       Impact factor: 5.214

8.  Impact of extending device longevity on the long-term costs of implantable cardioverter-defibrillator therapy: a modelling study with a 15-year time horizon.

Authors:  Giuseppe Boriani; Frieder Braunschweig; Jean Claude Deharo; Francisco Leyva; Andrzej Lubinski; Carlo Lazzaro
Journal:  Europace       Date:  2013-05-21       Impact factor: 5.214

9.  Investigation of a novel algorithm for synchronized left-ventricular pacing and ambulatory optimization of cardiac resynchronization therapy: results of the adaptive CRT trial.

Authors:  David O Martin; Bernd Lemke; David Birnie; Henry Krum; Kathy Lai-Fun Lee; Kazutaka Aonuma; Maurizio Gasparini; Randall C Starling; Goran Milasinovic; Tyson Rogers; Alex Sambelashvili; John Gorcsan; Mahmoud Houmsse
Journal:  Heart Rhythm       Date:  2012-07-14       Impact factor: 6.343

10.  Pacemakers and implantable cardioverter defibrillators: device longevity is more important than smaller size: the patient's viewpoint.

Authors:  David M Wild; John D Fisher; Soo G Kim; Kevin J Ferrick; Jay N Gross; Eugen C Palma
Journal:  Pacing Clin Electrophysiol       Date:  2004-11       Impact factor: 1.976

View more
  13 in total

1.  Plasticity of left ventricular function with cardiac resynchronization therapy.

Authors:  Hong-Xia Niu; Yi-Ran Hu; Wei Hua; Min Gu; Jing Wang; Michael R Gold; Shu Zhang
Journal:  J Interv Card Electrophysiol       Date:  2019-05-28       Impact factor: 1.900

2.  Association Between Industry Payments to Physicians and Device Selection in ICD Implantation.

Authors:  Amarnath R Annapureddy; Shady Henien; Yongfei Wang; Karl E Minges; Joseph S Ross; Erica S Spatz; Nihar R Desai; Pamela N Peterson; Frederick A Masoudi; Jeptha P Curtis
Journal:  JAMA       Date:  2020-11-03       Impact factor: 56.272

3.  Device runtime and costs of cardiac resynchronization therapy pacemakers - a health claims data analysis.

Authors:  Moritz Hadwiger; Nikolaos Dagres; Gerhard Hindricks; Helmut L'hoest; Ursula Marschall; Alexander Katalinic; Fabian-Simon Frielitz
Journal:  Ger Med Sci       Date:  2022-03-04

Review 4.  ENDURALIFE-Powered Cardiac Resynchronisation Therapy Defibrillator Devices for Treating Heart Failure: A NICE Medical Technology Guidance.

Authors:  James Michael Evans; Andrew Cleves; Helen Morgan; Liesl Millar; Grace Carolan-Rees
Journal:  Appl Health Econ Health Policy       Date:  2018-04       Impact factor: 2.561

5.  The economic impact of battery longevity in implantable cardioverter-defibrillators for cardiac resynchronization therapy: the hospital and healthcare system perspectives.

Authors:  Maurizio Landolina; Giovanni Morani; Antonio Curnis; Antonello Vado; Antonio D'Onofrio; Valter Bianchi; Giuseppe Stabile; Martino Crosato; Barbara Petracci; Carlo Ceriotti; Luca Bontempi; Martina Morosato; Gian Paolo Ballari; Maurizio Gasparini
Journal:  Europace       Date:  2017-08-01       Impact factor: 5.214

6.  Favorable Trend of Implantable Cardioverter-Defibrillator Service Life in a Large Single-Nation Population: Insights From 10-Year Analysis of the Italian Implantable Cardioverter-Defibrillator Registry.

Authors:  Stefano Poli; Giuseppe Boriani; Massimo Zecchin; Domenico Facchin; Maurizio Gasparini; Maurizio Landolina; Renato Pietro Ricci; Corrado Lanera; Dario Gregori; Alessandro Proclemer
Journal:  J Am Heart Assoc       Date:  2019-07-25       Impact factor: 5.501

7.  Longevity of implantable cardioverter defibrillators: a comparison among manufacturers and over time.

Authors:  Simon von Gunten; Beat A Schaer; Sing-Chien Yap; Tamas Szili-Torok; Michael Kühne; Christian Sticherling; Stefan Osswald; Dominic A M J Theuns
Journal:  Europace       Date:  2015-11-25       Impact factor: 5.214

8.  Battery drain in daily practice and medium-term projections on longevity of cardioverter-defibrillators: an analysis from a remote monitoring database.

Authors:  Giuseppe Boriani; Philippe Ritter; Mauro Biffi; Matteo Ziacchi; Igor Diemberger; Cristian Martignani; Cinzia Valzania; Sergio Valsecchi; Luigi Padeletti; Fredrik Gadler
Journal:  Europace       Date:  2016-02-03       Impact factor: 5.214

9.  Economic impact of longer battery life of cardiac resynchronization therapy defibrillators in Sweden.

Authors:  Fredrik Gadler; Yao Ding; Nathalie Verin; Martin Bergius; Jeffrey D Miller; Gregory M Lenhart; Mason W Russell
Journal:  Clinicoecon Outcomes Res       Date:  2016-10-31

10.  Persistence of ICD indication at the time of replacement in patients with initial implant for primary prevention indication: Effect on subsequent ICD therapies.

Authors:  Gabriele Dell'Era; Anna Degiovanni; Eraldo Occhetta; Andrea Magnani; Miriam Bortnik; Gabriella Francalacci; Laura Plebani; Eleonora Prenna; Sergio Valsecchi; Paolo Marino
Journal:  Indian Pacing Electrophysiol J       Date:  2016-11-14
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.