OBJECTIVE: Derive filtered tungsten X-ray spectra used in digital mammography systems by means of Monte Carlo simulations.Entities:
Keywords: HVL; Mammography; Monte Carlo; X-ray spectra
Year: 2015 PMID: 26811553 PMCID: PMC4725397 DOI: 10.1590/0100-3984.2014.0108
Source DB: PubMed Journal: Radiol Bras ISSN: 0100-3984
Figure 1Geometric model schematic drawing.
Figure 2Unfiltered tungsten X-ray spectra used in the Monte Carlo simulations.
Figure 3Experimental setup showing the compression plate in contact with detector.
HVL values for each voltage and W/Rh target/filter combination.
| HVL (mmAl) | ||
|---|---|---|
| kV | Experimental | Monte Carlo |
| 26 | 0.511 ± 3% | 0.513 ± 3% |
| 27 | 0.518 ± 3% | 0.527 ± 3% |
| 28 | 0.528 ± 3% | 0.535 ± 3% |
| 29 | 0.537 ± 3% | 0.552 ± 3% |
| 30 | 0.545 ± 3% | 0.565 ± 3% |
| 31 | 0.552 ± 3% | 0.574 ± 3% |
| 32 | 0.562 ± 3% | 0.585 ± 3% |
Estimated values.
Experimental HVL values uncertainty calculation.
| Uncertainty sources | Uncertainty (%) |
|---|---|
| Calibration | 2.0 |
| Detector resolution | 0.5 |
| Angular dependence | 0.5 |
| Temperature and humidity | 0.5 |
| Energy dependence | 2.0 |
| Readings accuracy | 0.3 |
| Measurement position | 0.3 |
| Combined standard uncertainty (k = 1) | 3.0 |
Standard deviation of the average of three measurements.
± 2 mm for 700 mm.
Figure 4HVL values for each voltage and W/Rh target/filter combination.
Figure 5Tungsten target X-ray spectra( and Monte Carlo obtained W/Rh target/filter combination X-ray spectra for 29 kV.