X-ray beam characteristics of the Varian Clinac 6-100 linear accelerator.

C W Coffey, J L Beach, D J Thompson, M Mendiondo.   

Abstract

Beam characteristics and dosimetry measurements of the 6 MV x-ray beam from a Varian Associates Clinac 6-100 linear accelerator are presented. Percentage depth dose tables are given as a function of field size for square fields. Tissue-maximum ratios and scatter-maximum ratios are tabulated as a function of depth and field size. Investigation of surface dose and depth of maximum dose reveal a dependence on field size. Other beam parameters measured are presented include field flatness, symmetry, SSD dependence, and penumbra. Treatment planning parameters including field size output factors and wedge factors are discussed.