| Literature DB >> 26699315 |
Shifeng Chen1, Byong Yong Yi, Xiaocheng Yang, Huijun Xu, Karl L Prado, Warren D D'Souza.
Abstract
Unlike other commercial treatment planning systems (TPS) which model the rounded leaf end differently (such as the MLC dosimetric leaf gap (DLG) or rounded leaf-tip radius), the RayStation TPS (RaySearch Laboratories, Stockholm, Sweden) models transmission through the rounded leaf end of the MLC with a step function, in which the radiation transmission through the leaf end is the square root of the average MLC transmission factor. We report on the optimization of MLC model parameters for the RayStation planning system. This (TPS) models the rounded leaf end of the MLC with the following parameters: eaf-tip offset, leaf-tip width, average transmission factor, and tongue and groove. We optimized the MLC model parameters for IMRT in the RayStation v. 4.0 planning system and for a Varian C-series linac with a 120-leaf Millennium MLC, and validated the model using measured data. The leaf-tip offset is the geometric offset due to the rounded leaf-end design and resulting divergence of the light/radiation field. The offset value is a function of the leaf-tip position, and tabulated data are available from the vendor. The leaf-tip width was iteratively evaluated by comparing computed and measured transverse dose profiles of MLC defined fields at dmax in water. In-water profile comparisons were also used to verify the MLC leaf position (leaf-tip offset). The average transmission factor and leaf tongue-and-groove width were derived iteratively by maximizing the agreement between measurements and RayStation TPS calculations for five clinical IMRT QA plans. Plan verifications were performed by comparing MapCHECK2 measurements and Monte Carlo calculations. The MLC model was validated using five test IMRT cases from the AAPM Task Group 119 report. Absolute gamma analyses (3 mm/3% and 2 mm/2%) were applied. In addition, computed output factors for MLC-defined small fields (2 × 2, 3 × 3, 4 × 4, 6× 6cm2) of both 6 MV and 18 MV photons were compared to those independently measured by the Imaging and Radiation Oncology Core (IROC), Houston, TX. 6MV and 18 MV models were both determined to have the same MLC parameters: leaf-tip offset = 0.3 cm, 2.5% transmission, and leaf tongue-and-groove width = 0.05 cm. IMRT QA analysis for five test cases in TG-119 resulted in a 100% passing rate with 3 mm/3% gamma analysis for 6 MV, and > 97.5% for 18 MV. The passing rate was > 94.6% for 6 MV and > 90.9% for 18 MV when the 2 mm/2% gamma analysis criteria was applied. These results compared favorably with those published in AAPM Task Group 119. The reported MLC model parameters serve as a reference for other users.Entities:
Mesh:
Year: 2015 PMID: 26699315 PMCID: PMC5690186 DOI: 10.1120/jacmp.v16i5.5548
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1The rounded leaf tip as shown in (a) is modeled as a region with half thickness of MLC leaf as shown in (b) in RayStation TPS. The MLC leaf tip offset (c) is the difference between light projections of leaf tip and leaf end .
Five step‐and‐shoot IMRT plans (6 MV photon) used to determine MLC tongue‐and‐groove and MLC radiation transmission factors
| A |
|
|
|
|
|
|
|---|---|---|---|---|---|---|
| Brain | N |
| 7 | 25 | 176 | 120 |
| Head & Neck | Y |
| 8 | 41 | 272 | 120 |
| Lung | N |
| 6 | 50 | 359 | 200 |
| Pancreas | N |
| 7 | 50 | 429 | 180 |
| Pelvis | Y |
| 14 | 50 | 706 | 180 |
Figure 2The geometric offset between light projection of MLC leaf end and projection of MLC leaf tip: ▪ indicates the data from vendor Varian, and solid line is the fitting curve ().
Figure 3The measured profile (CC04 ion chamber) vs. RayStation‐computed profile in the MLC leaf direction for an asymmetric field at in water for (a) 6 MV photon and (b) 18 MV photon. Both 6 MV and 18 MV photons have the same MLC parameters as following: 2.5% transmission, tongue and groove 0.05 cm, and leaf‐tip width 0.3 cm.
Gamma analysis results comparing the calculated plan dose from the TPS and the MapCHECK2 measurements for the pelvis case as a function of transmission factor. Tongue‐and‐groove , and leaf‐tip
|
|
|
|
|
|
|
|---|---|---|---|---|---|
| Gamma Passing Rate ( | 74.9 | 78.4 | 81.1 | 91.4 | 91.4 |
| Gamma Passing Rate ( | 91.8 | 94 | 95.2 | 98.2 | 98.3 |
QA results (gamma analysis) of five clinical IMRT plans (6 MV) using MLC parameters as follows: transmission factor 2.5%, tongue and groove 0.05 cm, and leaf‐tip width 0.3 cm
|
|
| |||
|---|---|---|---|---|
|
|
|
|
|
|
| Lung | 99.9% | 98.8% | 99.6% | 97.1% |
| Brain | 100% | 99.7% | 99.8% | 97.8% |
| HN | 100% | 99.0% | 99.4% | 96.4% |
| Pancreas | 100% | 100% | 99.7% | 97.2% |
| Pelvis (prostate) | 98.2% | 91.4% | 98.8% | 91.3% |
QA results (gamma analysis) of five clinical IMRT plans (18 MV) using MLC parameters as follows: transmission factor 2.5%, tongue and groove 0.05 cm, and leaf‐tip width 0.3 cm
|
|
|
|
|---|---|---|
| Lung | 97.2% | 85.5% |
| Brain | 100% | 99.0% |
| HN | 99.2% | 94.6% |
| Pancreas | 100% | 94.5% |
| Pelvis (prostate) | 95.3% | 85.5% |
QA results (gamma analysis) of five TG‐119 IMRT plans. Monte Carlo calculation is not available for 18 MV
|
|
|
| ||||
|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
| Test I1: Multitarget | 100% | 94.8% | 100% | 98.0% | 100% | 99% |
| Test I2: Mock Prostate | 100% | 98.5% | 100% | 96.2% | 98.6% | 91.3% |
| Test I3: Mock Head/neck | 100% | 98.5% | 100% | 97.5% | 97.4% | 90.9% |
| Test I4: Easy C‐shape | 100% | 100% | 100% | 98.2% | 100% | 100% |
| Test I5: Hard C‐shape | 100% | 94.6% | 100% | 97.6% | 100% | 99% |
QA results (local gamma analysis) of five TG‐119 IMRT plans measured using MapCHECK2
|
|
| |||
|---|---|---|---|---|
|
|
|
|
|
|
| Test I1: Multitarget | 97.4% | 85.5% | 97.9% | 96.4% |
| Test I2: Mock Prostate | 98.5% | 97.0% | 98.5% | 91.3% |
| Test I3: Mock Head/neck | 99.2% | 95.8% | 96.1% | 88.2% |
| Test I4: Easy C‐shape | 100% | 97.8% | 100% | 97% |
| Test I5: Hard C‐shape | 94.5% | 85.7% | 96.5% | 93.4% |
Relative output factors for MLC‐defined small fields for both beam energies 6 MV and 18 MV: IROC measurements, RayStation calculations, and their ratios. The output factors were defined at 10 cm depth with 100 cm SSD, and normalized to field
|
|
| |||||
|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
| 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 |
|
| 0.937 | 0.946 | 0.990 | 0.968 | 0.970 | 0.998 |
|
| 0.886 | 0.896 | 0.989 | 0.927 | 0.922 | 1.005 |
|
| 0.850 | 0.863 | 0.985 | 0.883 | 0.877 | 1.007 |
|
| 0.804 | 0.821 | 0.979 | 0.805 | 0.804 | 1.001 |