Literature DB >> 29785725

Density scaling of phantom materials for a 3D dose verification system.

Kensuke Tani1, Yukio Fujita2, Akihisa Wakita3, Ryohei Miyasaka4, Ryuzo Uehara5, Takumi Kodama6, Yuya Suzuki7, Ako Aikawa3, Norifumi Mizuno8, Jiro Kawamori8, Hidetoshi Saitoh1.   

Abstract

In this study, the optimum density scaling factors of phantom materials for a commercially available three-dimensional (3D) dose verification system (Delta4) were investigated in order to improve the accuracy of the calculated dose distributions in the phantom materials. At field sizes of 10 × 10 and 5 × 5 cm2 with the same geometry, tissue-phantom ratios (TPRs) in water, polymethyl methacrylate (PMMA), and Plastic Water Diagnostic Therapy (PWDT) were measured, and TPRs in various density scaling factors of water were calculated by Monte Carlo simulation, Adaptive Convolve (AdC, Pinnacle3 ), Collapsed Cone Convolution (CCC, RayStation), and AcurosXB (AXB, Eclipse). Effective linear attenuation coefficients (μeff ) were obtained from the TPRs. The ratios of μeff in phantom and water ((μeff )pl,water ) were compared between the measurements and calculations. For each phantom material, the density scaling factor proposed in this study (DSF) was set to be the value providing a match between the calculated and measured (μeff )pl,water . The optimum density scaling factor was verified through the comparison of the dose distributions measured by Delta4 and calculated with three different density scaling factors: the nominal physical density (PD), nominal relative electron density (ED), and DSF. Three plans were used for the verifications: a static field of 10 × 10 cm2 and two intensity modulated radiation therapy (IMRT) treatment plans. DSF were determined to be 1.13 for PMMA and 0.98 for PWDT. DSF for PMMA showed good agreement for AdC and CCC with 6 MV x ray, and AdC for 10 MV x ray. DSF for PWDT showed good agreement regardless of the dose calculation algorithms and x-ray energy. DSF can be considered one of the references for the density scaling factor of Delta4 phantom materials and may help improve the accuracy of the IMRT dose verification using Delta4.
© 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  zzm321990IMRTzzm321990; Delta4; density scaling; dose verification; phantom

Mesh:

Year:  2018        PMID: 29785725      PMCID: PMC6036349          DOI: 10.1002/acm2.12357

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


INTRODUCTION

It is necessary to verify the agreement between dose distributions calculated by a radiation treatment planning system (RTPS) and delivered by a linear accelerator (linac) for intensity modulated radiation therapy (IMRT).1, 2 A number of approaches and systems have been developed for this verification.2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13 Recently, three‐dimensional (3D) dose verification systems consisting of a solid phantom and detector arrays have become commercially available. These 3D dose verification systems can measure the absorbed dose at thousands of measurement points, and they are efficient in verifying the 3D dose distribution. IMRT dose verifications should be evaluated more accurately. Kly et al.14 showed that institutional patient‐specific IMRT quality assurance (QA) does not necessarily detect unacceptable plans. In their study, 14% of plans accepted by institutional IMRT QA were described as fail by an audit. In other words, even if an IMRT plan is accepted by one verification system, this does not ensure that the plan will be accepted by another verification system. Although various causes can be considered for this discrepancy, commissioning of verification systems is important to ensure the evaluation certainty. Because the verification is basically comparing the dose distributions in a solid phantom measured by the detectors and that calculated by the RTPS, the appropriate density scaling factor of the solid phantom used in the verification system should be adopted in the RTPS, where the density scaling factor is defined as a density to be assigned for the phantom material in RTPS (e.g., physical density, relative electron density, or other value). A number of studies for IMRT 3D dose verification systems have been reported. One of these 3D dose verification systems (Delta4 (ScandiDos, Inc., Ashland, VA, USA)) consists of 1069 silicon diodes arranged on two orthogonal boards in polymethyl methacrylate (PMMA) or Plastic Water Diagnostic Therapy (PWDT) as shown in Fig. 1. It has been used for IMRT patient‐specific QA,13, 15, 16, 17, 18, 19, 20 commissioning of volumetric modulated arc therapy (VMAT),21 and comparisons of dose calculation algorithms.22, 23 However, these studies had an approximately 2% dose difference resulting from the difference in density scaling factors of the phantom materials. Pham et al.15 and Feygelman et al.18 evaluated Delta4 with the same photon energy, phantom material, and RTPS; the former adopted a density scaling factor of 1.19, while the latter adopted 1.14 for the PMMA phantom. Other studies13, 17, 19, 21, 22, 23 have not reported the adopted density scaling factor, and their appropriateness has not been discussed so far.
Figure 1

Appearance of a 3D dose verification system (Delta4, ScandiDos). The silicon diodes are placed at 5 mm intervals in a central 6 cm × 6 cm area and at 10 mm intervals elsewhere in a 20 cm × 20 cm area on two orthogonal boards in the cylindrical phantom materials with a diameter of 22 cm. The standard detector geometry is depicted (“X”). If the attachment for sagittal‐coronal option was used, the detector geometry could be rotated to the “+” orientation.

Appearance of a 3D dose verification system (Delta4, ScandiDos). The silicon diodes are placed at 5 mm intervals in a central 6 cm × 6 cm area and at 10 mm intervals elsewhere in a 20 cm × 20 cm area on two orthogonal boards in the cylindrical phantom materials with a diameter of 22 cm. The standard detector geometry is depicted (“X”). If the attachment for sagittal‐coronal option was used, the detector geometry could be rotated to the “+” orientation. The purpose of this study was to clarify the optimum density scaling factor for PMMA and PWDT in order to improve the accuracy of the calculated dose distributions in the phantom materials of Delta4. The density scaling factors proposed in this study (DSF) for PMMA and PWDT were determined from measurements and calculations with several algorithms. The appropriateness of the DSF was validated by dose verifications with several plans using commercially available algorithms.

MATERIALS AND METHODS

Effective linear attenuation coefficients (μ eff)

Measurements of μ eff

Tissue‐phantom ratios (TPRs) in water, PMMA, and PWDT were measured at field sizes of 10 × 10 and 5 × 5 cm2 with 6 and 10 MV x rays from linacs (Clinac iX and 21EX (Varian Medical Systems, Palo Alto, CA, USA)). Calibration slab phantoms of Delta4 were stacked, and an ionization chamber (30013, PTW, Freiburg, Germany) was set at a source‐to‐chamber distance (SCD) of 100 cm, as shown in Fig. 2. The depths (d) were 4.25, 7.05, 9.25, and 12.05 cm in PMMA and 4.95, 8.45, 11.95, and 15.45 cm in PWDT. The reference depth of the TPRs was set to the shallowest depth. The μ eff at each condition were determined from the slope of the exponential regression curve approximating the TPR curve.
Figure 2

Example of the phantom geometry used to measure s. Normally, the manufacturer provides one slab for buildup, one slab for chamber insert, and one slab for backscatter in order to measure the absorbed dose at a depth of 4.25 cm in PMMA and 4.95 cm in PWDT for the cross‐calibration of Delta4. This figure shows four PWDT buildup slabs, one PWDT chamber insert slab (SCD = 100 cm, depth = 15.45 cm), and one PMMA slab for backscatter. The manufacture provides only PMMA for the backscatter slab. In this study, several sets of the calibration slab phantoms were stacked to measure s at several depths. To calculate s in the Monte Carlo simulation and RTPS, the geometry and phantoms were modeled the same as the geometry of the measurements.

Example of the phantom geometry used to measure s. Normally, the manufacturer provides one slab for buildup, one slab for chamber insert, and one slab for backscatter in order to measure the absorbed dose at a depth of 4.25 cm in PMMA and 4.95 cm in PWDT for the cross‐calibration of Delta4. This figure shows four PWDT buildup slabs, one PWDT chamber insert slab (SCD = 100 cm, depth = 15.45 cm), and one PMMA slab for backscatter. The manufacture provides only PMMA for the backscatter slab. In this study, several sets of the calibration slab phantoms were stacked to measure s at several depths. To calculate s in the Monte Carlo simulation and RTPS, the geometry and phantoms were modeled the same as the geometry of the measurements.

Calculations of μ eff

The following dose calculation algorithms were used to calculate the TPRs: Monte Carlo (the EGSnrc Monte Carlo code system24, 25 and BEAMnrc code system26), Adaptive Convolve (AdC) (Pinnacle3 ver. 9.10, Philips Radiation Oncology Systems, Fitchburg, WI, USA), Collapsed Cone Convolution (CCC) (RayStation ver. 4.5, RaySearch Laboratories, Stockholm, Sweden), and AcurosXB (AXB) (Eclipse ver. 11, Varian Medical Systems, Palo Alto, CA, USA). The geometries of the phantoms in the calculations were modeled to be the same as the measurement. The phantoms were assigned as water but the physical densities were varied from 0.96 to 1.19 g/cm3. The reasons for assigning the material of the phantoms as water were: (a) the phantoms could not be assigned as PMMA for AdC and PWDT for AdC, CCC, and AXB, and (b) the physical density of the materials other than water could not be changed from the default physical density for CCC and AXB. For the Monte Carlo simulation, the phantom materials were generated in PEGS (Preprocessor for EGS).27 The phase space data of the particles were scored and validated by comparing between calculated and measured depth dose and off‐axis ratio in water, and they were used in all simulations. The simulations were repeated until a statistical uncertainty of less than 0.1% was obtained. For the RTPS dose calculations, the grid size was 2 mm for AXB and 1 mm for the other dose calculation algorithms because dose using AXB with a grid size of 1 mm could not be calculated under several conditions due to a shortage of computer memory resources. The reference depth of the TPRs was set to the shallowest depth. The μ eff at each condition were determined from the slope of the exponential regression curve approximating the TPR curve.

Determining DSF

DSFs were determined through comparisons of the measured and calculated TPRs. To compare the TPRs between the measurements and calculations, the ratios of μ eff in phantom and water ((μ eff)pl,water) were used. The measured (μ eff)pl,water were obtained by dividing the μ eff measured in phantoms by the μ eff measured in water. The calculated (μ eff)pl,water were obtained by dividing the μ eff calculated with various density scaling factors by the μ eff calculated with the density scaling factor of 1.0. Although the beam qualities of the linacs used in this study were consistent, the μ eff calculated with the density scaling factor of 1.0 had a small variation among the dose calculation algorithms due to the modeling accuracy. This normalization is to make the changes of the slope of the TPRs for density scaling factors independent of the modeling accuracy of the each dose calculation algorithm. The measured (μ eff)pl,water were used as the reference value in the comparisons. The calculated (μ eff)pl,water were obtained as a function of the density scaling factors. The regression line approximating the median values of the (μ eff)pl,water calculated by the dose calculation algorithms for several density scaling factors was drawn. When the regression line matched the measured (μ eff)pl,water, the density scaling factor was set to DSF regression for the x‐ray energy, field size, and phantom material. Finally, for each phantom material, the mean value of DSF regression was used to define DSF. Additionally, specific DSFs for dose calculation algorithms (sDSF) were determined. Individually, the regression line approximating the (μ eff)pl,water calculated by each dose calculation algorithm for several density scaling factor was drawn. When each regression line matched the measured (μ eff)pl,water, the density scaling factor was set to sDSF of the dose calculation algorithm for a given condition.

Dose verifications with different density scaling factors

For each phantom material, the appropriateness of DSF was verified through comparisons of the dose distributions measured with Delta4 and calculated with three different density scaling factors: the nominal physical density (PD), nominal relative electron density (ED), and DSF. The PDs of PMMA and PWDT are 1.190 and 1.039 g/cm3, and EDs of PMMA and PWDT are 1.159 and 1.003, respectively.28, 29 The following dose calculation algorithms were used for the verification: AdC (Pinnacle3 ver. 9.0 for PMMA and 9.10 for PWDT), CCC, and AXB. Three plans were used for the verifications: one was 10 × 10, which denotes a static field of 10 × 10 cm2 with static gantry angles of 45° and 315° for the “+” (sagittal‐coronal option) detector geometry and 0° for the “X” (standard) detector geometry. The others were IMRT plans using “mock head&neck” and “mock prostate” in AAPM TG‐119.30 The IMRT plans were created following the dose constraints shown in AAPM TG‐119.30 The delivery techniques for the IMRT plans were step‐and‐shoot in Pinnacle3 and VMAT in the others. Before all measurements, the dose per monitor unit (DMU) of each x‐ray energy was obtained in accordance with the standard dosimetry protocol, and the daily machine output was corrected by the daily correction factor from the built‐in Delta4 software. These dose verifications were evaluated according to the pass rate of the global gamma index (gGI) for different criteria (2%/2 mm and 1%/1 mm) and the median of the global dose deviation (gDD) with the lower dose threshold of 20%. The normalization doses for the gGI and gDD were set to the measured dose at the isocenter for 10 × 10 and 2.0 Gy for the IMRT plans.

RESULTS

μ eff of TPRs

Figure 3 shows the measured (μ eff)pl,water and changes in the calculated (μ eff)pl,water for density scaling factors at a field size of 10 × 10 cm2. The measured (μ eff)pl,water were obtained within only the actual phantoms with no change in density, hence they are drawn as a horizontal line. The measured (μ eff)pl,water at 6 and 10 MV were 1.13 and 1.13 for PMMA, and 0.98 and 0.98 for PWDT, respectively.
Figure 3

Calculated (μ eff)pl,water as a function of density scaling factor for dose calculation and measured (μ eff)pl,water, presented as a horizontal line (field size = 10 × 10 cm2). The regression line approximating the median values of the calculated (μ eff)pl,water of the dose calculation algorithms at each density scaling factor is represented as a diagonal dashed line.

Calculated (μ eff)pl,water as a function of density scaling factor for dose calculation and measured (μ eff)pl,water, presented as a horizontal line (field size = 10 × 10 cm2). The regression line approximating the median values of the calculated (μ eff)pl,water of the dose calculation algorithms at each density scaling factor is represented as a diagonal dashed line. Figure 4 shows the measured (μ eff)pl,water and changes in the calculated (μ eff)pl,water for density scaling factors at a field size of 5 × 5 cm2. The measured (μ eff)pl,water at 6 and 10 MV were 1.13 and 1.13 for PMMA, and 0.98 and 0.99 for PWDT, respectively.
Figure 4

Calculated (μ eff)pl,water as a function of density scaling factor for dose calculation and measured (μ eff)pl,water, represented as a horizontal line (field size = 5 × 5 cm2). The regression line approximating the median values of the calculated (μ eff)pl,water of the dose calculation algorithms at each density scaling factor is represented as a diagonal dashed line.

Calculated (μ eff)pl,water as a function of density scaling factor for dose calculation and measured (μ eff)pl,water, represented as a horizontal line (field size = 5 × 5 cm2). The regression line approximating the median values of the calculated (μ eff)pl,water of the dose calculation algorithms at each density scaling factor is represented as a diagonal dashed line.

DSF of PMMA and PWDT

Based on Figs. 3 and 4, DSF regression was determined as the density scaling factor when the regression line matched the measured (μ eff)pl,water. At a field size of 10 × 10 cm2, DSF regression of 6 and 10 MV x ray were 1.13 and 1.14 for PMMA, and 0.98 and 0.98 for PWDT, respectively. At a field size of 5 × 5 cm2, DSF regression of 6 and 10 MV x ray were 1.13 and 1.14 for PMMA, and 0.98 and 0.99 for PWDT, respectively. Therefore, DSF in this study was determined as 1.13 for PMMA and 0.98 for PWDT, as given in Table 1.
Table 1

Nominal physical density, nominal relative electron density, and DSF of PMMA and PWDT

Phantom materialPhysical density [g cm−3]Relative electron density DSF
PMMA1.1901.1591.13
PWDT1.0391.0030.98
Nominal physical density, nominal relative electron density, and DSF of PMMA and PWDT At a field size of 10 × 10 cm2, sDSF of Monte Carlo, AdC, CCC, and AXB for PMMA were 1.13, 1.13, 1.15, and 1.12 in 6 MV x ray, and 1.13, 1.13, 1.14, and 1.13 in 10 MV x ray, respectively. The sDSF of Monte Carlo, AdC, CCC, and AXB for PWDT were 0.97, 0.98, 0.98, and 0.97 in 6 MV x ray, and 0.98, 0.99, 0.98, and 0.98 in 10 MV x ray, respectively. At a field size of 5 × 5 cm2, sDSF of Monte Carlo, AdC, CCC, and AXB for PMMA were 1.13, 1.14, 1.15, and 1.11 in 6 MV x ray, and 1.13, 1.13, 1.12, and 1.11 in 10 MV x ray, respectively. The sDSF of Monte Carlo, AdC, CCC, and AXB for PWDT were 0.97, 0.98, 0.99, and 0.98 in 6 MV x ray, and 0.99, 0.98, 0.99, and 0.98 in 10 MV x ray, respectively. Figure 5 shows the measured and calculated dose profiles in Delta4 PMMA phantom with AdC for the several dose verifications using PD and DSF. These verification plans were 10 × 10 with 6 MV x ray, mock head&neck with 6 MV x ray, and mock prostate with 10 MV x ray. Fig. 6 shows the pass rates of gGI and median of gDD of the dose calculation algorithms with different density scaling factors for the dose verifications of 10 × 10 with 6 MV x ray. Tables 2 and 3 show the summary of the pass rates of gGI and median of gDD of the dose calculation algorithms with different density scaling factors for the dose verifications of 10 × 10 and IMRT within PMMA and PWDT.
Figure 5

Comparisons between the measured and calculated dose profiles in the Delta4 PMMA phantom with AdC for the several dose verifications using PD and . The pictures at the upper left show the measured dose distributions for each plan on one of the two orthogonal detector boards and the blue lines on the pictures show the position of the displayed profiles. [Correction added on 28 May, after first Online publication: Figure 5 position rearrangement has been updated.]

Figure 6

Pass rates of gGI (left axis, bars; : gGI with the criterion of 2%/2 mm, and : gGI with the criterion of 1%/1 mm) and median of gDD (right axis, circle and line; ) of several dose calculation algorithms with physical density (PD), relative electron density (ED), and (Plan: 10 × 10, x‐ray energy: 6 MV).

Table 2

Summary of the pass rates (%) of gGI with the criteria of 2%/2 mm and 1%/1 mm, and median (%) of gDD of several dose calculation algorithms with physical density (PD), relative electron density (ED), and DSF for PMMA

Plan/x‐ray energyAdCCCCAXB
MetricsPDEDDSFPDEDDSFPDEDDSF
10 × 10/6 MV
gGI (2%/2 mm)48.339.495.668.293.499.597.797.086.1
gGI (1%/1 mm)22.818.971.925.260.485.769.277.048.3
Median of gDD2.62.90.81.91.00.20.5−0.5−1.5
Mock head&neck/6 MV
gGI (2%/2 mm)67.254.498.285.696.299.564.583.097.0
gGI (1%/1 mm)31.927.881.831.927.881.845.259.377.8
Median of gDD2.12.40.51.60.90.21.81.10.3
Mock prostate/6 MV
gGI (2%/2 mm)79.872.7100.093.598.798.987.697.999.2
gGI (1%/1 mm)51.346.787.767.388.189.263.782.992.2
Median of gDD1.71.90.11.30.5−0.31.20.5−0.1
10 × 10/10 MV
gGI (2%/2 mm)63.452.798.799.099.796.097.594.783.8
gGI (1%/1 mm)21.619.475.882.082.161.881.764.838.2
Median of gDD2.32.50.70.5−0.2−0.9−0.2−0.9−1.7
Mock head&neck/10 MV
gGI (2%/2 mm)83.271.899.895.299.199.178.793.198.8
gGI (1%/1 mm)37.233.389.371.184.185.754.571.382.6
Median of gDD1.71.90.50.90.4−0.11.30.70.3
Mock prostate/10 MV
gGI (2%/2 mm)88.182.3100.099.499.093.495.598.799.0
gGI (1%/1 mm)57.553.797.289.684.673.778.888.891.1
Median of gDD1.51.70.30.2−0.4−0.90.60.1−0.3
Table 3

Summary of the pass rates (%) of gGI with the criteria of 2%/2 mm and 1%/1 mm, and median (%) of gDD of several dose calculation algorithms with physical density (PD), relative electron density (ED), and DSF for PWDT

Plan/x‐ray energyAdCCCCAXB
MetricsPDEDDSFPDEDDSFPDEDDSF
10 × 10/6 MV
gGI (2%/2 mm)57.3100.0100.039.196.0100.082.799.898.9
gGI (1%/1 mm)23.171.796.320.737.281.436.474.482.1
Median of gDD2.10.80.22.51.40.61.50.4−0.3
Mock head&neck/6 MV
gGI (2%/2 mm)74.790.196.073.492.898.862.478.791.4
gGI (1%/1 mm)53.463.869.852.770.081.842.754.568.2
Median of gDD1.40.60.12.11.20.72.11.30.9
Mock prostate/6 MV
gGI (2%/2 mm)74.790.196.579.497.299.972.591.998.6
gGI (1%/1 mm)53.463.870.358.578.190.848.363.979.9
Median of gDD1.40.60.11.90.90.21.91.30.7
10 × 10/10 MV
gGI (2%/2 mm)99.0100.0100.096.9100.0100.084.099.8100.0
gGI (1%/1 mm)43.296.995.246.693.797.136.069.387.1
Median of gDD1.40.4−0.11.30.50.01.40.70.2
Mock head&neck/10 MV
gGI (2%/2 mm)90.399.4100.084.795.298.563.878.387.3
gGI (1%/1 mm)62.578.383.256.975.985.837.648.260.3
Median of gDD1.10.30.11.71.00.62.21.51.1
Mock prostate/10 MV
gGI (2%/2 mm)90.399.4100.094.999.7100.076.690.596.6
gGI (1%/1 mm)62.578.383.780.093.693.953.267.878.4
Median of gDD1.10.30.11.10.3−0.11.91.31.0
Comparisons between the measured and calculated dose profiles in the Delta4 PMMA phantom with AdC for the several dose verifications using PD and . The pictures at the upper left show the measured dose distributions for each plan on one of the two orthogonal detector boards and the blue lines on the pictures show the position of the displayed profiles. [Correction added on 28 May, after first Online publication: Figure 5 position rearrangement has been updated.] Pass rates of gGI (left axis, bars; : gGI with the criterion of 2%/2 mm, and : gGI with the criterion of 1%/1 mm) and median of gDD (right axis, circle and line; ) of several dose calculation algorithms with physical density (PD), relative electron density (ED), and (Plan: 10 × 10, x‐ray energy: 6 MV). Summary of the pass rates (%) of gGI with the criteria of 2%/2 mm and 1%/1 mm, and median (%) of gDD of several dose calculation algorithms with physical density (PD), relative electron density (ED), and DSF for PMMA Summary of the pass rates (%) of gGI with the criteria of 2%/2 mm and 1%/1 mm, and median (%) of gDD of several dose calculation algorithms with physical density (PD), relative electron density (ED), and DSF for PWDT Figure 6 shows the graphs of the numerical data of 10 × 10 with 6 MV x ray in Tables 2 and 3. The pass rates of gGI and median of gDD were changed dramatically according to the changes of adopted density scaling factors. For AdC (ver.9.0) in PMMA, there was an unreasonable change at the ED. The measured dose distributions were consistent regardless of the adopted density scaling factors in the dose verifications. The calculated dose distributions should change according to the adopted density scaling factor; the calculated dose should become gradually higher when the adopted density scaling factor becomes gradually lower. However, the median of gDD using ED (1.159) was not between that using PD (1.19) and DSF (1.13). This may be due to the coarse resolution of the mass attenuation coefficient in an older version of Pinnacle3, as pointed out by Dickof.31 Except for AXB within PMMA, the pass rates of gGI increased and median of gDD moved close to 0% from the PD to DSF. The tendency was consistent regardless of the dose calculation algorithms, verification plans, and x‐ray energy within PWDT, as shown in Table 3. DSF for PWDT showed good agreement between the measured and calculated dose distributions under multiple conditions. On the other hand, as shown in Table 2, DSF for PMMA showed good agreement between those in AdC and CCC with 6 MV x ray, and AdC for 10 MV x ray for 10 × 10. The PD or ED showed good agreement between those in AXB with 6 MV x ray, and CCC and AXB with 10 MV x ray for 10 × 10. The results within PMMA varied depending on the dose calculation algorithms and x‐ray energy. Although the dose verifications of 10 × 10 for AXB within PMMA were conducted in three institutions after the absolute dose calibration for the Delta4 detectors, these results were unchanged. This removes the dependence of these results on site‐specific errors such as linac output, cross‐calibration of Delta4, or beam data in RTPS. In Tables 2 and 3, the results for CCC within PMMA and PWDT were obtained at one institution, and those for AXB within PMMA and PWDT were obtained at a different institution. Although the results of the dose verifications for CCC within PWDT were consistent for both 6 and 10 MV x ray, the results for CCC within PMMA were not consistent. The optimum density scaling factor for CCC in PMMA was DSF in 6 MV x ray and ED in 10 MV x ray. Although the results of mock head&neck for CCC with 10 MV x ray were slightly different from those of 10 × 10 and mock prostate in PMMA, it was the same tendency as that seen in the results in PWDT. However, the optimum density scaling factor of PMMA for AXB seemed to be PD or ED in 10 × 10 and DSF in IMRT. The reason may be the systematic dose difference of 1% for the IMRT plans between the linac output and dose calculation in the institution. The results of the IMRT verifications in PWDT were also higher by 1% compared to 10 × 10.

DISCUSSION

The choice of density scaling factor has a large effect on the ability to accurately calculate dose distributions in the Delta4 phantoms. This is evidenced by the fact that the average difference of the pass rates of gGI with the criterion of 2%/2 mm between PD and DSF for mock head&neck plans in 6 MV x ray were 25.8% in PMMA and 25.2% in PWDT. Furthermore, as shown in Fig. 5, the calculated dose profiles adopted with PD were clearly different from both the measured dose distributions and the calculated dose profiles adopted with DSF. The choice of density scaling factor plays a crucial role not only for patient‐specific IMRT QA in order to appropriately judge for pass or fail, but also for the IMRT commissioning for RTPS. The reason why the choice is important for the IMRT commissioning is that the calculated dose distributions would need to be adjusted by modifying several dose calculation parameters (e.g., dosimetric leaf gap, MLC transmission, tongue and groove effect, or focal spot size) compared with measured dose distributions.32, 33, 34, 35 For these modifications, the usage of a solid phantom with detector arrays or films would be inevitable in order to evaluate steep dose distributions. If the dose verifications for these modifications were conducted with an inadequate density scaling factor, these parameters may be decided as inadequate values. Therefore, the appropriate choice of the density scaling factor to improve the accuracy of the calculated dose distribution in solid phantoms is important for patient‐specific IMRT QA and IMRT commissioning for RTPS. Although the relative electron density is commonly assigned to solid phantoms in RTPS, its appropriateness for density scaling has been demonstrated for the equivalent path length in narrow photon beams.36 However, the optimum density scaling factor for 3D dose verification is not the ratio for the equivalent path length of photon interactions but the density in RTPS that gives a calculated dose distribution closely matching the measured dose distribution. Therefore, the DSFs were obtained according to the changes of depth dose that included the component of the scattered dose in the phantoms. To determine the cause for this divergence as the component of the scattered dose in the phantom or others, further study should be conducted with more types of solid phantoms and geometry conditions. In this study for the density scaling factors of Delta4 phantom materials, DSF were obtained as lower values than the nominal relative electron densities through our original method. Regarding the density scaling factors of PMMA applied in other studies of Delta4, Pham and Luo15 used 1.19 in Pinnacle3. Geurts et al. used 1.19 in TomoTherapy system.20 Kumagai et al.37 found 1.16 for 4 MV and 1.15 for 10 MV in Pinnacle3. Feygelman et al.18 used 1.14 in Pinnacle3 and their value was close to the DSF in the previous studies. TG‐2138 was the reference for the relative electron density of 1.14. TG‐2138 calculated a relative electron density of 1.137 from the physical density of 1.17 g/cm3 for PMMA. However, the physical density of PMMA was shown as 1.19 g/cm3.28, 29, 39 In additional investigations, we measured the physical density and analyzed the elemental composition of PMMA in a portion of Delta4 with the thermal conductivity method for hydrogen and carbon, and the infrared absorption spectrophotometry for oxygen. Consequently, the physical density was 1.19 g/cm3, and the elemental composition closely matched the nominal elemental composition.28, 29, 39 The DSF are the lowest density scaling factor acquired theoretically. DSF was shown to be the optimum density scaling factor for PWDT regardless of the dose calculation algorithm and x‐ray energy. Furthermore, the changes of the (μ eff)pl,water calculated by the dose calculation algorithms were consistent at each density scaling factor for PWDT. On the other hand, these changes were not consistent for PMMA. Specifically, these changes of AdC were lower than those of other dose calculation algorithms and those of AXB were higher in several conditions, as shown in Figs. 3 and 4. The reason these differences occurred in treating different densities of water for PMMA was difficult to specify because the details of the dose calculation algorithms related to treat different densities of water in the calculations opened to the public were limited. At least, the results showed the possibility that optimum density scaling factor for AdC may become higher than DSF such as ED and the one for AXB may become lower than DSF through the dose verifications. However, in the dose verifications for PMMA, DSF was the optimum density scaling factor in AdC and CCC with 6 MV x ray, and AdC with 10 MV x ray. The PD or ED may be the optimum density scaling factor in AXB with 6 MV x ray, and CCC and AXB with 10 MV x ray, nevertheless none of the (μ eff)pl,water calculated by the dose calculation algorithms matched the measured (μ eff)pl,water at PD and ED, as shown in Fig. 2. A reason for the considerable deviation may be the accuracy of the absorbed dose calculation in a higher density of water. Because DSF were obtained from the slopes of the TPRs, the DSF were not found to be an appropriate density scaling factor for the absorbed dose calculation in different densities of water. If there was some mismatch or uncertainty between the slope of the depth dose and absorbed dose calculated in a different density of water, it should be corrected by something other than the density scaling factor.

CONCLUSIONS

The difference in density scaling factors caused a bigger dosimetric difference than the pass/fail criterion. We clarified DSF of PMMA and PWDT from measurements and calculations, and validated the appropriateness of DSF. The DSF were lower than not only the PD but also the ED. DSF can be used as a reference for the density scaling factor of the Delta4 phantom material in multiple clinical institutions and may help improve the accuracy of the IMRT dose verification using Delta4.

CONFLICT OF INTEREST

The authors declare no conflicts of interest.
  32 in total

1.  Commissioning of volumetric modulated arc therapy (VMAT).

Authors:  James L Bedford; Alan P Warrington
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-02-01       Impact factor: 7.038

2.  Characterization of a novel 2D array dosimeter for patient-specific quality assurance with volumetric arc therapy.

Authors:  Sotirios Stathakis; Pamela Myers; Carlos Esquivel; Panayiotis Mavroidis; Nikos Papanikolaou
Journal:  Med Phys       Date:  2013-07       Impact factor: 4.071

3.  A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systems.

Authors:  Mohammad Hussein; Pejman Rowshanfarzad; Martin A Ebert; Andrew Nisbet; Catharine H Clark
Journal:  Radiother Oncol       Date:  2013-10-04       Impact factor: 6.280

4.  Evaluation of a 3D diode array dosimeter for helical tomotherapy delivery QA.

Authors:  Vladimir Feygelman; D Opp; K Javedan; A J Saini; G Zhang
Journal:  Med Dosim       Date:  2010-01-22       Impact factor: 1.482

5.  Patient-specific quality assurance method for VMAT treatment delivery.

Authors:  Eduard Schreibmann; Anees Dhabaan; Eric Elder; Tim Fox
Journal:  Med Phys       Date:  2009-10       Impact factor: 4.071

6.  The next step in patient-specific QA: 3D dose verification of conformal and intensity-modulated RT based on EPID dosimetry and Monte Carlo dose calculations.

Authors:  Wouter van Elmpt; Sebastiaan Nijsten; Ben Mijnheer; André Dekker; Philippe Lambin
Journal:  Radiother Oncol       Date:  2007-12-03       Impact factor: 6.280

7.  Evaluation of a new VMAT QA device, or the "X" and "O" array geometries.

Authors:  Vladimir Feygelman; Geoffrey Zhang; Craig Stevens; Benjamin E Nelms
Journal:  J Appl Clin Med Phys       Date:  2011-01-31       Impact factor: 2.102

8.  Characterization and clinical evaluation of a novel IMRT quality assurance system.

Authors:  Ramaswamy Sadagopan; Jose A Bencomo; Rafael L Martin; Gorgen Nilsson; Thomas Matzen; Peter A Balter
Journal:  J Appl Clin Med Phys       Date:  2009-05-07       Impact factor: 2.102

9.  Evaluation of a biplanar diode array dosimeter for quality assurance of step-and-shoot IMRT.

Authors:  Vladimir Feygelman; Kenneth Forster; Daniel Opp; Görgen Nilsson
Journal:  J Appl Clin Med Phys       Date:  2009-09-30       Impact factor: 2.102

10.  Determining the optimal dosimetric leaf gap setting for rounded leaf-end multileaf collimator systems by simple test fields.

Authors:  Weiguang Yao; Jonathan B Farr
Journal:  J Appl Clin Med Phys       Date:  2015-07-08       Impact factor: 2.102