| Literature DB >> 35433523 |
Mojgan Alizadeh1, Mehran Mohseni2, Bagher Farhood2, Akbar Aliasgharzadeh2, Habiballah Moradi2, Ahmad Ramazani-Moghaddam-Arani3.
Abstract
Background: The dose values obtained from procedures of diagnostic radiology are relatively low. To accurately and precisely measure the dose values in this dose range, it is necessary to know the characteristics of dosimeters. Objective: The aim of this study was to evaluate several thermoluminescent characteristics of GR-200, TLD-700H and TLD-100 for low dose measurement. Material andEntities:
Keywords: GR-200; Radiation Dosage; Radiology; TLD-100; TLD-700H; Thermoluminescent Dosimetry
Year: 2022 PMID: 35433523 PMCID: PMC8995761 DOI: 10.31661/jbpe.v0i0.1247
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
Reading parameters used for various thermoluminescent dosimeters (TLDs).
| Parameters of Time-temperature profile | GR-200 | TLD-100 and TLD-700H |
|---|---|---|
| Preheating temperature | 135 °C | 50° |
| Preheating time | 10 s | 5 s |
| Preheating speed | 10 °C/s | 10 °C/s |
| Maximum temperature | 240 °C | 300 °C |
| Acquisition time | 21 s | 26 s |
| Annealing temperature | 240 °C | 300 °C |
Figure 1The responses of GR-200 (a), TLD-700H (b) and TLD-100 (c) as a function of absorbed dose in 0.05-10 mGy dose range at 1 day after irradiation. The standard deviations in a number of points are low; hence, the error bars are often not seen clearly.
Goodness of the linear fit to dose-response data of GR-200, TLD-700H and TLD-100.
| Parameters | GR-200 | TLD-700 | TLD-100 |
|---|---|---|---|
| SSE | 3.4×105 | 533.1 | 2453.0 |
| R-square | 0.999 | 0.995 | 0.992 |
| Adjusted R-square | 0.999 | 0.994 | 0.992 |
| RMSE | 195.50 | 7.70 | 16.51 |
SSE: Sum of Squares Error, RMSE: Root-Mean-Square Error
Figure 2The dose-response curves of GR-200 (a), TLD-700H (b) and TLD-100 (c) for various values of current tube in 0. 5-4 mGy dose range at 1 day after irradiation. The standard deviations in a number of points are low; hence, the error bars are often not seen clearly.
Figure 3The dose-response curves of GR-200 (a), TLD-700H (b) and TLD-100 (c) for various photon energies in 0. 5-4 mGy dose range at 1 day after irradiation. The standard deviations in a number of points are low; hence, the error bars are often not seen clearly.