Literature DB >> 26644156

Development of an Automated Bone Mineral Density Software Application: Facilitation Radiologic Reporting and Improvement of Accuracy.

I-Ta Tsai1, Meng-Yuan Tsai2, Ming-Ting Wu2, Clement Kuen-Huang Chen3.   

Abstract

The conventional method of bone mineral density (BMD) report production by dictation and transcription is time consuming and prone to error. We developed an automated BMD reporting system based on the raw data from a dual energy X-ray absorptiometry (DXA) scanner for facilitating the report generation. The automated BMD reporting system, a web application, digests the DXA's raw data and automatically generates preliminary reports. In Jan. 2014, 500 examinations were randomized into an automatic group (AG) and a manual group (MG), and the speed of report generation was compared. For evaluation of the accuracy and analysis of errors, 5120 examinations during Jan. 2013 and Dec. 2013 were enrolled retrospectively, and the context of automatically generated reports (AR) was compared with the formal manual reports (MR). The average time spent for report generation in AG and in MG was 264 and 1452 s, respectively (p < 0.001). The accuracy of calculation of T and Z scores in AR is 100 %. The overall accuracy of AR and MR is 98.8 and 93.7 %, respectively (p < 0.001). The mis-categorization rate in AR and MR is 0.039 and 0.273 %, respectively (p = 0.0013). Errors occurred in AR and can be grouped into key-in errors by technicians and need for additional judgements. We constructed an efficient and reliable automated BMD reporting system. It facilitates current clinical service and potentially prevents human errors from technicians, transcriptionists, and radiologists.

Entities:  

Keywords:  Bone mineral density; DXA; Open source; Radiology reporting; Workflow

Mesh:

Year:  2016        PMID: 26644156      PMCID: PMC4879035          DOI: 10.1007/s10278-015-9848-7

Source DB:  PubMed          Journal:  J Digit Imaging        ISSN: 0897-1889            Impact factor:   4.056


  6 in total

1.  Error in radiology.

Authors:  P Goddard; A Leslie; A Jones; C Wakeley; J Kabala
Journal:  Br J Radiol       Date:  2001-10       Impact factor: 3.039

2.  Quantitative computed tomography (QCT) as a radiology reporting tool by using optical character recognition (OCR) and macro program.

Authors:  Young Han Lee; Ho-Taek Song; Jin-Suck Suh
Journal:  J Digit Imaging       Date:  2012-12       Impact factor: 4.056

3.  The effect of voice recognition software on comparative error rates in radiology reports.

Authors:  S McGurk; K Brauer; T V Macfarlane; K A Duncan
Journal:  Br J Radiol       Date:  2008-07-15       Impact factor: 3.039

4.  Informatics in radiology: use of a macro scripting editor to facilitate transfer of dual-energy X-ray absorptiometry reports into an existing departmental voice recognition dictation system.

Authors:  Michael Iv; Mahesh R Patel; Adam Santos; Young S Kang
Journal:  Radiographics       Date:  2011-05-05       Impact factor: 5.333

Review 5.  Consensus development conference: diagnosis, prophylaxis, and treatment of osteoporosis.

Authors: 
Journal:  Am J Med       Date:  1993-06       Impact factor: 4.965

Review 6.  Executive summary of the 2013 International Society for Clinical Densitometry Position Development Conference on bone densitometry.

Authors:  John T Schousboe; John A Shepherd; John P Bilezikian; Sanford Baim
Journal:  J Clin Densitom       Date:  2013 Oct-Dec       Impact factor: 2.617

  6 in total
  2 in total

1.  Structured reporting has the potential to reduce reporting times of dual-energy x-ray absorptiometry exams.

Authors:  Su Hwan Kim; Lara M Sobez; Judith E Spiro; Adrian Curta; Felix Ceelen; Eric Kampmann; Martin Goepfert; Raphael Bodensohn; Felix G Meinel; Wieland H Sommer; Nora N Sommer; Franziska Galiè
Journal:  BMC Musculoskelet Disord       Date:  2020-04-16       Impact factor: 2.362

2.  BatchBMD as an Efficient and Accurate Dual-Energy X-ray Absorptiometry Report Generator.

Authors:  Chun-Hsiang Chan; Wen-Chi Huang; Yi-Chien Lu; Hsing-Fen Hsiao; Wing P Chan
Journal:  Diagnostics (Basel)       Date:  2021-12-20
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.