| Literature DB >> 26609428 |
Tomoichiro Ogawa1, Takuya Shiba1, Hiroshi Tsuneoka1.
Abstract
Purpose. This study retrospectively analyzed the postoperative visual functions of myopic eyes implanted with multifocal intraocular lens (IOL) to evaluate the efficacy of multifocal IOL in highly myopic eyes. Methods. We studied 61 patients (96 eyes) who were implanted with multifocal IOL ZMA00 or ZMB00 (Abbott Medical Optics). The patients were stratified into two groups by axial length: 26 mm or above (AL ≥ 26 group) and below 26 mm (AL < 26 group). Postoperative corrected and uncorrected distance (5 m) and near (30 cm) visual acuity (VA), contrast sensitivity, and depth of focus were compared between two groups. Results. In the AL ≥ 26 group and the AL < 26 group, the mean ± standard deviation uncorrected distance logMAR VA at 12-month postoperative follow-up was -0.04 ± 0.11 and -0.01 ± 0.14, respectively; and the corrected distance VA was -0.17 ± 0.08 and -0.14 ± 0.07, with no significant differences between two groups (p = 0.558 and 0.101; Mann-Whitney U test). For near VA, the corresponding uncorrected VA was 0.06 ± 0.08 and 0.05 ± 0.09; and distance-corrected VA was 0.01 ± 0.06 and 0.01 ± 0.02, with no significant differences between two groups (p = 0.572, and 0.157; Mann-Whitney U test). Conclusion. The present study demonstrates that it is possible to achieve good uncorrected near and distance VA following implantation of multifocal IOL in eyes with long axial lengths.Entities:
Year: 2015 PMID: 26609428 PMCID: PMC4644551 DOI: 10.1155/2015/956046
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Demographic and clinical characteristics of patients stratified into axial length of 26 or above (AL ≥ 26 group) and axial length below 26 mm (AL < 26 group).
| AL ≥ 26 group | AL < 26 group |
| |
|---|---|---|---|
| Number of eyes | 32 | 64 | |
| ZMA00 | 18 | 36 | |
| ZMB00 | 14 | 28 | |
| Age (y) mean ± SD (range) | 57.0 ± 10.9 | 59.1 ± 8.2 | 0.703 |
| (40–71) | (41–69) | ||
| Male/female | 15/6 | 13/27 | |
| Axial length (mm) | 27.38 | 24.13 | <0.001 |
| Mean (range) | (26.06–29.24) | (22.00–25.95) |
Figure 1Uncorrected distance visual acuity during 12-month follow-up after implantation of multifocal intraocular lens. (a) Eyes were divided into axial length of 26 mm or longer (AL ≥ 26 group) and less than 26 mm (AL < 26 group). (b) Eyes in AL ≥ 26 group were further divided into 26 ≤ AL < 28 group and AL ≥ 28 group. NS Mann-Whitney U test.
Figure 2Best corrected distance visual acuity during 12-month follow-up after implantation of multifocal intraocular lens. (a) Eyes were divided into axial length of 26 mm or longer (AL ≥ 26 group) and less than 26 mm (AL < 26 group). (b) Eyes in AL ≥ 26 group were further divided into 26 ≤ AL < 28 group and AL ≥ 28 group. Mann-Whitney U test.
Figure 3Uncorrected near visual acuity during 12-month follow-up after implantation of multifocal intraocular lens. (a) Eyes were divided into axial length of 26 mm or longer (AL ≥ 26 group) and less than 26 mm (AL < 26 group). (b) Eyes in AL ≥ 26 group were further divided into 26 ≤ AL < 28 group and AL ≥ 28 group. NS Mann-Whitney U test.
Figure 4Best corrected near visual acuity during 12-month follow-up after implantation of multifocal intraocular lens. (a) Eyes were divided into axial length of 26 mm or longer (AL ≥ 26 group) and less than 26 mm (AL < 26 group). (b) Eyes in AL ≥ 26 group were further divided into 26 ≤ AL < 28 group and AL ≥ 28 group. NS Mann-Whitney U test.
Figure 5Best distance-corrected near visual acuity during 12-month follow-up after implantation of multifocal intraocular lens. (a) Eyes were divided into axial length of 26 mm or longer (AL ≥ 26 group) and less than 26 mm (AL < 26 group). (b) Eyes in AL ≥ 26 group were further divided into 26 ≤ AL < 28 group and AL ≥ 28 group. NS Mann-Whitney U test.
Postoperative refractive data. (A) Patients were stratified into axial length of 26 or above (AL ≥ 26 group) and below 26 mm (AL < 26 group). (B) Eyes in AL ≥ 26 group were further divided into 26 ≤ AL < 28 group and AL ≥ 28 group.
| A | B | |||||
|---|---|---|---|---|---|---|
| AL ≥ 26 group | AL < 26 group |
| 26 ≤ AL < 28 group | AL ≥ 28 group |
| |
| Postoperative spherical equivalent (D) | 0.14 ± 0.35 | −0.25 ± 0.44 | <0.001 | 0.18 ± 0.38 | −0.12 ± 0.34 | 0.742 |
|
| ||||||
| Postoperative absolute spherical equivalent (D) | 0.29 ± 0.23 | 0.35 ± 0.37 | 0.956 | 0.29 ± 0.29 | 0.29 ± 0.21 | 0.869 |
|
| ||||||
| Postoperative refractive error (D) | 0.29 ± 0.37 | −0.23 ± 0.36 | <0.001 | 0.44 ± 0.23 | −0.25 ± 0.40 | 0.232 |
|
| ||||||
| Postoperative absolute refractive error (D) | 0.39 ± 0.27 | 0.30 ± 0.30 | 0.068 | 0.44 ± 0.23 | 0.37 ± 0.28 | 0.477 |
|
| ||||||
| Postoperative corneal astigmatism (D) | 0.66 ± 0.44 | 0.64 ± 0.42 | 0.730 | 0.83 ± 0.33 | 0.60 ± 0.47 | 0.112 |
Data are expressed as mean ± SD (range).
Figure 6Defocus curves. (a) Eyes were divided into axial length of 26 mm or longer (AL ≥ 26 group) and less than 26 mm (AL < 26 group). Bimodal curves are shown with two peaks at spherical equivalent addition of 0.0 D and −3.0 D. Although a significant difference between two groups is observed at spherical equivalent addition of −5.0 D, no significant differences were found at the other values. (b) Eyes in AL ≥ 26 group were further divided into 26 ≤ AL < 28 group and AL ≥ 28 group. Bimodal curves are alsoobserved. Significant differences are detected at additional spherical equivalent of 0.0, −1.5, −2.0, −2.5, and −3.0 D. Mann-Whitney U test.
Figure 7Contrast sensitivity at various target sizes. (a) Eyes were divided into axial length of 26 mm or longer (AL ≥ 26 group) and less than 26 mm (AL < 26 group). In the absence of glare, significant differences in contrast sensitivity are observed between two groups at target sizes of 6.3°, 4.0°, and 2.5°. In the presence of glare, significant differences are observed between two groups at target sizes of 6.3° and 2.5°. (b) Eyes in AL ≥ 26 group were further divided into 26 ≤ AL < 28 group and AL ≥ 28 group. No significant differences are observed. Mann-Whitney U test.
Distance and near visual acuity following implantation of multifocal intraocular lenses: the present study and those of Wang et al. [4] and Alfonso et al. [5].
| This report | Wang et al. [ | Alfonso et al. [ | |
|---|---|---|---|
| UDVA | −0.04 ± 0.13 | 0.18 ± 0.12 | 0.18 ± 0.20 |
| BCDVA | −0.15 ± 0.06 | 0.05 ± 0.05 | 0.05 ± 0.09 |
| UNVA (30 cm) | 0.07 ± 0.10 | 0.10 ± 0.07 | 0.09 ± 0.10 |
| BDCNVA (30 cm) | 0.02 ± 0.05 | 0.07 ± 0.08 | 0.07 ± 0.08 |
UDVA, uncorrected distant visual acuity; BCDVA, best corrected visual acuity; UNVA, uncorrected near visual acuity; BDCNVA, best distance-corrected near visual acuity.