Jorge L Alió1, Alfredo Vega-Estrada2, Ana B Plaza-Puche3.
Abstract
BACKGROUND: To evaluate the refractive outcomes and the optical performance as well as the quality of life in patients implanted with a new diffractive multifocal intraocular lens (IOL).Entities:
Year: 2015 PMID: 26605358 PMCID: PMC4657369 DOI: 10.1186/s40662-015-0012-8
Source DB: PubMed Journal: Eye Vis (Lond) ISSN: 2326-0254
Visual and refractive outcomes comparison between groups
|
|
|
|
|
|
|
|---|---|---|---|---|---|
|
|
| ||||
|
| 0.73 (0.38) | 0.21 (0.15) | 0.22 (0.17) | 0.22 (0.20) | <0.01 |
| 0.30 to 1.50 | 0.00 to 0.62 | 0.00 to 0.70 | 0.00 to 0.93 | ||
|
| -0.41 (2.52) | -0.04 (0.59) | 0.16 (0.65) | 0.10 (0.98) | 0.53 |
| -5.00 to +3.50 | -1.00 to +1.50 | -0.75 to +2.00 | -3.00 to +2.00 | ||
|
| -0.78 (0.54) | -0.55 (0.47) | -0.70 (0.55) | -0.81 (0.54) | 0.23 |
| -1.75 to 0.00 | -1.25 to 0.00 | -1.50 to 0.00 | -2.25 to 0.00 | ||
|
| 0.33 (0.31) | 0.04 (0.05) | 0.07 (0.16) | 0.04 (0.06) | <0.01 |
| 0.00 to 1.00 | 0.00 to 0.12 | 0.00 to 0.70 | 0.00 to 0.20 | ||
|
| 0.69(0.22) | 0.24 (0.12) | 0.31 (0.22) | 0.24 (0.15) | <0.01 |
| 0.40 to 1.00 | 0.00 to 0.40 | 0.00 to 0.90 | 0.00 to 0.60 | ||
|
| 0.36 (0.26) | 0.13 (0.08) | 0.14 (0.14) | 0.08 (0.08) | <0.01 |
| 0.10 to 1.0 | 0.00 to 0.30 | 0.00 to 0.60 | 0.00 to 0.30 | ||
|
| 2.73 (0.24) | 0.75 (0.61) | 0.88 (0.81) | 0.81 (0.65) | <0.01 |
| 2.50 to +3.00 | 0.00 to +1.75 | 0.00 to 2.50 | 0.00 to +1.50 |
Comparative table shows the preoperative and postoperative visual conditions of patients included in this study. The corresponding p-values for the comparison between preoperative and postoperative follow up are shown for each parameter evaluated.
*Abbreviations: SD, standard deviation; D, diopters; UDVA, uncorrected distance visual acuity; CDVA, corrected distance visual acuity; UNVA, uncorrected near visual acuity; CNVA, corrected near visual acuity.
Comparison of the pre and postoperative variables analyzed in the study
|
|
| |||||
|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
|
| |||||
|
| 67.30 (3.77) | 68.37 (9.11) | 0.65 | --- | --- | --- |
| 58 to 71 | 46 to 80 | |||||
|
| 0.73 (0.38) | 1.40 (0.74) | <0.01 | 0.22 (0.17) | 0.13 (0.14) | 0.02 |
| 0.30 to 1.50 | 0.10 to 2.00 | 0.00 to 0.70 | 0.00 to 0.50 | |||
|
| -0.41 (2.52) | -0.61 (4.70) | 0.92 | +0.16 (0.65) | +0.03 (0.48) | 0.4 |
| -5.00 to +3.50 | -7.25 to +12.00 | -0.75 to +2.00 | -1.00 to +1.25 | |||
|
| -0.78 (0.54) | -0.80 (0.46) | 0.93 | -0.70 (0.55) | -0.59 (0.56) | 0.37 |
| -1.75 to 0.00 | -1.75 to 0.00 | -1.50 to 0.00 | -2.25 to 0.00 | |||
|
| 0.33 (0.31) | 0.24 (0.22) | 0.51 | 0.07 (0.16) | 0.02 (0.04) | 0.02 |
| 0.00 to 1.00 | 0.00 to 0.82 | 0.00 to 0.70 | 0.00 to 0.20 | |||
|
| 0.69 (0.22) | 0.71 (0.34) | 0.64 | 0.31 (0.22) | 0.57 (0.16) | <0.01 |
| 0.40 to 1.00 | 0.10 to 1.10 | 0.00 to 0.90 | 0.30 to 0.90 | |||
|
| --- | --- | --- | 0.26 (0.18) | 0.60 (0.12) | <0.01 |
| 0.00 to 0.80 | 0.40 to 0.89 | |||||
|
| 0.36 (0.27) | 0.26 (0.19) | 0.15 | 0.14 (0.14) | 0.15 (0.07) | 0.77 |
| 0.10 to 1.20 | 0.10 to 0.70 | 0.00 to 0.60 | 0.10 to 0.30 | |||
|
| 20.80 (1.87) | 21.74 (4.98) | 0.5 | --- | --- | --- |
| 18.0 to 23.50 | 14.0 to 40 | |||||
Comparative table shows the preoperative and postoperative conditions at 3 months after cataract surgery. The corresponding p-values for the comparison between groups are shown for each parameter evaluated.
*Abbreviations: SD, standard deviation; D, diopters; UDVA, uncorrected distance visual acuity; CDVA, corrected distance visual acuity; UCNVA, uncorrected near visual acuity; CNVA, corrected near visual acuity.
Figure 1Defocus curve comparison between groups. Comparison of mean defocus curve between the patients implanted with the SeeLens MF IOL and the monofocal IOL.
Figure 2Contrast sensitivity function in the multifocal IOL group. Mean postoperative contrast sensitivity function (CSF) in patients implanted with the multifocal IOL in logarithmic scale under photopic and scotopic conditions at 3 and 6 months after surgery. Normal values for the same age group are shown in light gray.
Figure 3Contrast sensitivity function comparison between groups. It shows comparison of the postoperative contrast sensitivity function in both groups of patients under photopic and scotopic conditions.
Figure 4Internal aberrations in the multifocal IOL group. It shows evolution of the internal aberrations throughout the follow up period. There is a significant reduction for the root mean square (RMS) of the internal high order aberrations six months after implantation of the SeeLens MF IOL.
Results of the VF-14 quality of life questionnaire in the multifocal IOL group
|
|
|
|---|---|
| 1. Reading small print, such as medicine bottle labels, a telephone book, or food labels | 1.00 ± 0.93 |
| 2. Reading a newspaper or a book | 0.50 ± 0.53 |
| 3. Reading a large-print book or large-print newspaper or numbers on a telephone | 0.13 ± 0.35 |
| 4. Recognizing people when they are close to you | 0.33 ± 0.71 |
| 5. Seeing steps, stairs or curbs | 0.11 ± 0.33 |
| 6. Reading traffic signs, street signs or store signs | 0.11 ± 0.33 |
| 7. Doing fine handwork like sewing, knitting, crocheting, carpentry | 0.75 ± 0.89 |
| 8. Writing checks or filling out forms | 0.63 ± 0.74 |
| 9. Playing games such as bingo, dominos, card games, or mahjong | 0.00 ± 0.00 |
| 10. Taking part in sports like bowling, handball, tennis, golf | 0.00 ± 0.00 |
| 11. Cooking | 0.00 ± 0.00 |
| 12. Watching television | 0.22 ± 0.44 |
| 13. Driving during the day | 0.20 ± 0.45 |
| 14. Driving at night | 1.20 ± 0.45 |
Mean values of the VF-14 QOL questionnaire items at 6 months postoperatively. Grading scale: 0, no difficulty; 1, a little difficulty; 2, moderate difficulty; 3, quite difficult; 4, impossible to perform.