Evaluation of optimized breath-hold and free-breathing 3D ultrashort echo time contrast agent-free MRI of the human lung.

Neville D Gai1, Ashkan Malayeri1, Harsh Agarwal2, Robert Evers1, David Bluemke1.   

Abstract

PURPOSE: To evaluate an optimized stack of radials ultrashort echo time (UTE) 3D magnetic resonance imaging (MRI) sequence for breath-hold and free-breathing imaging of the human lung.
MATERIALS AND METHODS: A 3D stack of ultrashort echo time radials trajectory was optimized for coronal and axial lower-resolution breath-hold and higher-resolution free-breathing scans using Bloch simulations. The sequence was evaluated in 10 volunteers, without the use of contrast agents. Signal-to-noise ratio (SNR) mean and 95% confidence interval (CI) were determined from separate signal and noise images in a semiautomated fashion. The four scanning schemes were evaluated for significant differences in image quality using Student's t-test. Ten clinical patients were scanned with the sequence and findings were compared with concomitant computed tomography (CT) in nine patients. Breath-hold 3D spokes images were compared with 3D stack of radials in five volunteers. A Mann-Whitney U-test was performed to test significance in both cases.
RESULTS: Breath-hold imaging of the entire lung in volunteers was performed with SNR (mean = 42.5 [CI]: 35.5-49.5; mean = 34.3 [CI]: 28.6-40) in lung parenchyma for coronal and axial scans, respectively, which can be used as a quick scout scan. Longer respiratory triggered free-breathing scan enabled high-resolution UTE scanning with mean SNR of 14.2 ([CI]: 12.9-15.5) and 9.2 ([CI]: 8.2-10.2) for coronal and axial scans, respectively. Axial free-breathing scans showed significantly higher image quality (P = 0.008) than the three other scanning schemes. The mean score for comparison with CT was 1.67 (score 0: n = 0; 1: n = 3; 2: n = 6). There was no significant difference between CT and MRI (P = 0.25). 3D stack of radials images were significantly better than 3D spokes images (P < 0.001).
CONCLUSION: The optimized 3D stack of radials trajectory was shown to provide high-quality MR images of the lung parenchyma without the use of MRI contrast agents. The sequence may offer the possibility of breath-hold imaging and provides greater flexibility in trading off slice thickness and parallel imaging for scan time.
© 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  3D stack of radials; breath-hold; free-breathing; lung parenchyma

Mesh:

Substances:

Year:  2015        PMID: 26458867      PMCID: PMC4833722          DOI: 10.1002/jmri.25073

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  30 in total

1.  Projection reconstruction techniques for reduction of motion effects in MRI.

Authors:  G H Glover; J M Pauly
Journal:  Magn Reson Med       Date:  1992-12       Impact factor: 4.668

2.  Lung parenchyma: projection reconstruction MR imaging.

Authors:  C J Bergin; J M Pauly; A Macovski
Journal:  Radiology       Date:  1991-06       Impact factor: 11.105

3.  Selection of a convolution function for Fourier inversion using gridding [computerised tomography application].

Authors:  J I Jackson; C H Meyer; D G Nishimura; A Macovski
Journal:  IEEE Trans Med Imaging       Date:  1991       Impact factor: 10.048

4.  Imaging of pulmonary pathologies: focus on magnetic resonance imaging.

Authors:  Hans-Ulrich Kauczor; Julia Ley-Zaporozhan; Sebastian Ley
Journal:  Proc Am Thorac Soc       Date:  2009-08-15

Review 5.  Non-Cartesian parallel imaging reconstruction.

Authors:  Katherine L Wright; Jesse I Hamilton; Mark A Griswold; Vikas Gulani; Nicole Seiberlich
Journal:  J Magn Reson Imaging       Date:  2014-01-10       Impact factor: 4.813

Review 6.  MRI of the lung: state of the art.

Authors:  Mark Wielpütz; Hans-Ulrich Kauczor
Journal:  Diagn Interv Radiol       Date:  2012-03-20       Impact factor: 2.630

7.  Respiratory Motion-Resolved Compressed Sensing Reconstruction of Free-Breathing Radial Acquisition for Dynamic Liver Magnetic Resonance Imaging.

Authors:  Hersh Chandarana; Li Feng; Justin Ream; Annie Wang; James S Babb; Kai Tobias Block; Daniel K Sodickson; Ricardo Otazo
Journal:  Invest Radiol       Date:  2015-11       Impact factor: 6.016

8.  Myocardial perfusion MRI with an undersampled 3D stack-of-stars sequence.

Authors:  Liyong Chen; Ganesh Adluru; Matthias C Schabel; Chris J McGann; Edward V R Dibella
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

9.  Comparison of lung T2* during free-breathing at 1.5 T and 3.0 T with ultrashort echo time imaging.

Authors:  Jiangsheng Yu; Yiqun Xue; Hee Kwon Song
Journal:  Magn Reson Med       Date:  2011-02-24       Impact factor: 4.668

10.  Optimized 3D ultrashort echo time pulmonary MRI.

Authors:  Kevin M Johnson; Sean B Fain; Mark L Schiebler; Scott Nagle
Journal:  Magn Reson Med       Date:  2012-12-04       Impact factor: 4.668

View more
  4 in total

1.  From low-dose to no-dose: thin-section magnetic resonance imaging for evaluation of pulmonary nodules.

Authors:  Tommaso D'Angelo; Thomas J Vogl; Julian L Wichmann
Journal:  J Thorac Dis       Date:  2018-04       Impact factor: 2.895

2.  Long T2 suppression in native lung 3-D imaging using k-space reordered inversion recovery dual-echo ultrashort echo time MRI.

Authors:  Neville D Gai; Ashkan A Malayeri; David A Bluemke
Journal:  MAGMA       Date:  2017-03-10       Impact factor: 2.310

3.  Hyperpolarized 129Xe for investigation of mild cystic fibrosis lung disease in pediatric patients.

Authors:  Robert P Thomen; Laura L Walkup; David J Roach; Zackary I Cleveland; John P Clancy; Jason C Woods
Journal:  J Cyst Fibros       Date:  2016-07-29       Impact factor: 5.482

4.  Ultra-short echo time images quantify high liver iron.

Authors:  Eamon K Doyle; Kristin Toy; Bertin Valdez; Jonathan M Chia; Thomas Coates; John C Wood
Journal:  Magn Reson Med       Date:  2017-06-22       Impact factor: 4.668

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.