Literature DB >> 26337773

Applicability of Bipolar Charge Analyzer (BOLAR) in Characterizing the Bipolar Electrostatic Charge Profile of Commercial Metered Dose Inhalers (MDIs).

Sharon Shui Yee Leung1, Adam Chong Ming Chiow1, Ari Ukkonen2, Hak-Kim Chan3.   

Abstract

PURPOSE: To investigate the applicability of Bipolar Charge Analyzer (BOLAR), a new commercial instrument developed by Dekati Ltd., in simultaneously characterizing the bipolar electrostatic charge profile and measuring the size distribution of commercial metered dose inhalers (MDIs).
METHODS: Intal Forte(®) (sodium cromoglycate), Tilade(®) (nedocromil sodium), Ventolin(®) (salbutamol sulphate), and QVAR(®) (beclomethasone dipropionate) were used as model MDIs in this study. Three individual actuations of each MDI were introduced into the BOLAR at an air flow rate of 60 l/min. Charge and mass profiles for each actuation were determined.
RESULTS: The BOLAR was found to have better performance in collecting valid charge data (≥80%) than valid mass data (≥50%). In all tested products, both positively and negatively charged particles were found in five defined size fractions between zero and 11.6 μm, with the charge magnitude decreased with increasing particle size. The net charge profiles obtained from the BOLAR qualitatively agreed with the results reported previously. In all suspension type MDIs, negligible masses were detected in the smallest size fraction (<0.95 μm), for which the charge was most likely caused by the propellant and excipients. QVAR was the only solution MDI tested and the charge and mass profiles were significantly different from the suspension-type MDIs. Its mass profile was found to follow closely with the charge profile.
CONCLUSIONS: Positively and negatively charged MDI particles of different size fractions and their corresponding charge-to-mass profiles were successfully characterized by the BOLAR.

Entities:  

Keywords:  BOLAR; aerosols; bipolar charge; electrostatic charge; particle deposition

Mesh:

Substances:

Year:  2015        PMID: 26337773     DOI: 10.1007/s11095-015-1786-z

Source DB:  PubMed          Journal:  Pharm Res        ISSN: 0724-8741            Impact factor:   4.200


  15 in total

1.  Electrostatic charge characteristics of aerosols produced from metered dose inhalers.

Authors:  Philip Chi Lip Kwok; William Glover; Hak-Kim Chan
Journal:  J Pharm Sci       Date:  2005-12       Impact factor: 3.534

2.  On the deposition of unipolarly charged particles in the human respiratory tract.

Authors:  C Melandri; V Prodi; G Tarroni; M Formignani; T De Zaiacomo; G F Bompane; G Maestri
Journal:  Inhaled Part       Date:  1975-09

3.  The effect of ethanol on the formation and physico-chemical properties of particles generated from budesonide solution-based pressurized metered-dose inhalers.

Authors:  Bing Zhu; Daniela Traini; Hak-Kim Chan; Paul M Young
Journal:  Drug Dev Ind Pharm       Date:  2013-11       Impact factor: 3.225

Review 4.  Electrostatics of pharmaceutical inhalation aerosols.

Authors:  Philip Chi Lip Kwok; Hak-Kim Chan
Journal:  J Pharm Pharmacol       Date:  2009-12       Impact factor: 3.765

5.  Measurements of electrodynamic effects on the deposition of MDI and DPI aerosols in a replica cast of human oral-pharyngeal-laryngeal airways.

Authors:  Mohammed Ali; Malay K Mazumder; Ted B Martonen
Journal:  J Aerosol Med Pulm Drug Deliv       Date:  2009-03       Impact factor: 2.849

6.  Precipitation of submicron charged particles in human lung airways.

Authors:  C P Yu; K Chandra
Journal:  Bull Math Biol       Date:  1977       Impact factor: 1.758

7.  The Effect of Active Pharmaceutical Ingredients on Aerosol Electrostatic Charges from Pressurized Metered Dose Inhalers.

Authors:  Yang Chen; Paul M Young; David F Fletcher; Hak Kim Chan; Edward Long; David Lewis; Tanya Church; Daniela Traini
Journal:  Pharm Res       Date:  2015-03-19       Impact factor: 4.200

8.  An indication on the biological variability of aerosol total deposition in humans.

Authors:  G Tarroni; C Melandri; V Prodi; T De Zaiacomo; M Formignani; P Bassi
Journal:  Am Ind Hyg Assoc J       Date:  1980-11

9.  The influence of actuator materials and nozzle designs on electrostatic charge of pressurised metered dose inhaler (pMDI) formulations.

Authors:  Yang Chen; Paul M Young; David F Fletcher; Hak Kim Chan; Edward Long; David Lewis; Tanya Church; Daniela Traini
Journal:  Pharm Res       Date:  2013-12-03       Impact factor: 4.200

10.  The influence of flow rate on the aerosol deposition profile and electrostatic charge of single and combination metered dose inhalers.

Authors:  Susan Hoe; Daniela Traini; Hak-Kim Chan; Paul M Young
Journal:  Pharm Res       Date:  2009-10-06       Impact factor: 4.200

View more
  2 in total

1.  Measuring The Bipolar Charge Distributions of Fine Particle Aerosol Clouds of Commercial PMDI Suspensions Using a Bipolar Next Generation Impactor (bp-NGI).

Authors:  Martin Rowland; Alessandro Cavecchi; Frank Thielmann; Janusz Kulon; Jag Shur; Robert Price
Journal:  Pharm Res       Date:  2018-11-26       Impact factor: 4.200

2.  Effect of Relative Humidity on Bipolar Electrostatic Charge Profiles of dry Powder Aerosols.

Authors:  Jiaqi Yu; Jennifer Wong; Ari Ukkonen; Jonna Kannosto; Hak-Kim Chan
Journal:  Pharm Res       Date:  2017-05-24       Impact factor: 4.200

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.