Literature DB >> 24297072

The influence of actuator materials and nozzle designs on electrostatic charge of pressurised metered dose inhaler (pMDI) formulations.

Yang Chen1, Paul M Young, David F Fletcher, Hak Kim Chan, Edward Long, David Lewis, Tanya Church, Daniela Traini.   

Abstract

PURPOSE: To investigate the influence of different actuator materials and nozzle designs on the electrostatic charge properties of a series of solution metered dose inhaler (pMDI) aerosols.
METHODS: Actuators were manufactured with flat and cone nozzle designs using five different materials from the triboelectric series (Nylon, Polyethylene terephthalate, Polyethylene-High density, Polypropylene copolymer and Polytetrafluoroethylene). The electrostatic charge profiles of pMDI containing beclomethasone dipropionate (BDP) as model drug in HFA-134a propellant, with different concentrations of ethanol were studied. Electrostatic measurements were taken using a modified electrical low-pressure impactor (ELPI) and the deposited drug mass assayed chemically using HPLC.
RESULTS: The charge profiles of HFA 134a alone have shown strong electronegativity with all actuator materials and nozzle designs, at an average of -1531.34 pC ± 377.34. The presence of co-solvent ethanol significantly reduced the negative charge magnitude. BDP reduced the suppressing effect of ethanol on the negative charging of the propellant. For all tested formulations, the flat nozzle design showed no significant differences in net charge between different actuator materials, whereas the charge profiles of cone designs followed the triboelectric series.
CONCLUSION: The electrostatic charging profiles from a solution pMDI containing BDP and ethanol can be significantly influenced by the actuator material, nozzle design and formulation components. Ethanol concentration appears to have the most significant impact. Furthermore, BDP interactions with ethanol and HFA have an influence on the electrostatic charge of aerosols. By choosing different combinations of actuator materials and orifice design, the fine particle fractions of formulations can be altered.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24297072     DOI: 10.1007/s11095-013-1253-7

Source DB:  PubMed          Journal:  Pharm Res        ISSN: 0724-8741            Impact factor:   4.200


  32 in total

1.  Electrostatic charge characteristics of aerosols produced from metered dose inhalers.

Authors:  Philip Chi Lip Kwok; William Glover; Hak-Kim Chan
Journal:  J Pharm Sci       Date:  2005-12       Impact factor: 3.534

2.  Regional lung deposition and bronchodilator response as a function of beta2-agonist particle size.

Authors:  Omar S Usmani; Martyn F Biddiscombe; Peter J Barnes
Journal:  Am J Respir Crit Care Med       Date:  2005-09-28       Impact factor: 21.405

Review 3.  Electrostatics and inhaled medications: influence on delivery via pressurized metered-dose inhalers and add-on devices.

Authors:  Jolyon P Mitchell; Dominic P Coppolo; Mark W Nagel
Journal:  Respir Care       Date:  2007-03       Impact factor: 2.258

4.  Measurements of electrodynamic effects on the deposition of MDI and DPI aerosols in a replica cast of human oral-pharyngeal-laryngeal airways.

Authors:  Mohammed Ali; Malay K Mazumder; Ted B Martonen
Journal:  J Aerosol Med Pulm Drug Deliv       Date:  2009-03       Impact factor: 2.849

5.  Static electrification of airborne asbestos: a study of its causes, assessment and effects on deposition in the lungs of rats.

Authors:  J H Vincent; W B Johnston; A D Jones; A M Johnston
Journal:  Am Ind Hyg Assoc J       Date:  1981-10

6.  Dynamic electrostatic charge of lactose-salbutamol sulphate powder blends dispersed from a Cyclohaler®.

Authors:  Susan Hoe; Paul M Young; Daniela Traini
Journal:  Drug Dev Ind Pharm       Date:  2011-05-10       Impact factor: 3.225

7.  An indication on the biological variability of aerosol total deposition in humans.

Authors:  G Tarroni; C Melandri; V Prodi; T De Zaiacomo; M Formignani; P Bassi
Journal:  Am Ind Hyg Assoc J       Date:  1980-11

8.  The effect of aerosol charge on the deposition and clearance of TiO2 particles in rats.

Authors:  J Ferin; T T Mercer; L J Leach
Journal:  Environ Res       Date:  1983-06       Impact factor: 6.498

9.  Effect of relative humidity on the electrostatic charge properties of dry powder inhaler aerosols.

Authors:  Philip Chi Lip Kwok; Hak-Kim Chan
Journal:  Pharm Res       Date:  2007-07-10       Impact factor: 4.200

10.  The influence of flow rate on the aerosol deposition profile and electrostatic charge of single and combination metered dose inhalers.

Authors:  Susan Hoe; Daniela Traini; Hak-Kim Chan; Paul M Young
Journal:  Pharm Res       Date:  2009-10-06       Impact factor: 4.200

View more
  3 in total

1.  The Effect of Active Pharmaceutical Ingredients on Aerosol Electrostatic Charges from Pressurized Metered Dose Inhalers.

Authors:  Yang Chen; Paul M Young; David F Fletcher; Hak Kim Chan; Edward Long; David Lewis; Tanya Church; Daniela Traini
Journal:  Pharm Res       Date:  2015-03-19       Impact factor: 4.200

2.  The effect of actuator nozzle designs on the electrostatic charge generated in pressurised metered dose inhaler aerosols.

Authors:  Yang Chen; Paul M Young; David F Fletcher; Hak Kim Chan; Edward Long; David Lewis; Tanya Church; Daniela Traini
Journal:  Pharm Res       Date:  2014-10-01       Impact factor: 4.200

3.  Applicability of Bipolar Charge Analyzer (BOLAR) in Characterizing the Bipolar Electrostatic Charge Profile of Commercial Metered Dose Inhalers (MDIs).

Authors:  Sharon Shui Yee Leung; Adam Chong Ming Chiow; Ari Ukkonen; Hak-Kim Chan
Journal:  Pharm Res       Date:  2015-09-03       Impact factor: 4.200

  3 in total