A tailored ML-EM algorithm for reconstruction of truncated projection data using few view angles.

Yanfei Mao1, Gengsheng L Zeng.   

Abstract

Dedicated cardiac single photon emission computed tomography (SPECT) systems have the advantage of high speed and sensitivity at no loss, or even a gain, in resolution. The potential drawbacks of these dedicated systems are data truncation by the small field of view (FOV) and the lack of view angles. Serious artifacts, including streaks outside the FOV and distortion in the FOV, are introduced to the reconstruction when using the traditional emission data maximum-likelihood expectation-maximization (ML-EM) algorithm to reconstruct images from the truncated data with a small number of views. In this note, we propose a tailored ML-EM algorithm to suppress the artifacts caused by data truncation and insufficient angular sampling by reducing the image updating step sizes for the pixels outside the FOV. As a consequence, the convergence speed for the pixels outside the FOV is decelerated. We applied the proposed algorithm to truncated analytical data, Monte Carlo simulation data and real emission data with different numbers of views. The computer simulation results show that the tailored ML-EM algorithm outperforms the conventional ML-EM algorithm in terms of streak artifacts and distortion suppression for reconstruction from truncated projection data with a small number of views.

Entities:  

Mesh:

Year:  2013        PMID: 23689102      PMCID: PMC3745016          DOI: 10.1088/0031-9155/58/12/N157

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  14 in total

1.  GATE: a simulation toolkit for PET and SPECT.

Authors:  S Jan; G Santin; D Strul; S Staelens; K Assié; D Autret; S Avner; R Barbier; M Bardiès; P M Bloomfield; D Brasse; V Breton; P Bruyndonckx; I Buvat; A F Chatziioannou; Y Choi; Y H Chung; C Comtat; D Donnarieix; L Ferrer; S J Glick; C J Groiselle; D Guez; P F Honore; S Kerhoas-Cavata; A S Kirov; V Kohli; M Koole; M Krieguer; D J van der Laan; F Lamare; G Largeron; C Lartizien; D Lazaro; M C Maas; L Maigne; F Mayet; F Melot; C Merheb; E Pennacchio; J Perez; U Pietrzyk; F R Rannou; M Rey; D R Schaart; C R Schmidtlein; L Simon; T Y Song; J M Vieira; D Visvikis; R Van de Walle; E Wieërs; C Morel
Journal:  Phys Med Biol       Date:  2004-10-07       Impact factor: 3.609

2.  Maximum likelihood reconstruction for emission tomography.

Authors:  L A Shepp; Y Vardi
Journal:  IEEE Trans Med Imaging       Date:  1982       Impact factor: 10.048

Review 3.  New trends in camera and software technology in nuclear cardiology.

Authors:  Ernest V Garcia; Tracy L Faber
Journal:  Cardiol Clin       Date:  2009-05       Impact factor: 2.213

4.  Tiny a priori knowledge solves the interior problem in computed tomography.

Authors:  Hiroyuki Kudo; Matias Courdurier; Frédéric Noo; Michel Defrise
Journal:  Phys Med Biol       Date:  2008-04-09       Impact factor: 3.609

5.  Comparison of simultaneous dual-isotope multipinhole SPECT with rotational SPECT in a group of patients with coronary artery disease.

Authors:  Peter P Steele; Dennis L Kirch; John E Koss
Journal:  J Nucl Med       Date:  2008-06-13       Impact factor: 10.057

6.  Null-space function estimation for the interior problem.

Authors:  Gengsheng L Zeng; Grant T Gullberg
Journal:  Phys Med Biol       Date:  2012-03-16       Impact factor: 3.609

7.  EM reconstruction algorithms for emission and transmission tomography.

Authors:  K Lange; R Carson
Journal:  J Comput Assist Tomogr       Date:  1984-04       Impact factor: 1.826

8.  Compressed sensing based interior tomography.

Authors:  Hengyong Yu; Ge Wang
Journal:  Phys Med Biol       Date:  2009-04-15       Impact factor: 3.609

9.  Novel solid-state-detector dedicated cardiac camera for fast myocardial perfusion imaging: multicenter comparison with standard dual detector cameras.

Authors:  Fabio P Esteves; Paolo Raggi; Russell D Folks; Zohar Keidar; J Wells Askew; Shmuel Rispler; Michael K O'Connor; Liudmilla Verdes; Ernest V Garcia
Journal:  J Nucl Cardiol       Date:  2009-08-18       Impact factor: 5.952

10.  A general local reconstruction approach based on a truncated hilbert transform.

Authors:  Yangbo Ye; Hengyong Yu; Yuchuan Wei; Ge Wang
Journal:  Int J Biomed Imaging       Date:  2007
View more
  2 in total

1.  Segmented slant hole collimator for stationary cardiac SPECT: Monte Carlo simulations.

Authors:  Yanfei Mao; Zhicong Yu; Gengsheng L Zeng
Journal:  Med Phys       Date:  2015-09       Impact factor: 4.071

2.  The impact of system matrix dimension on small FOV SPECT reconstruction with truncated projections.

Authors:  Chung Chan; Joyoni Dey; Yariv Grobshtein; Jing Wu; Yi-Hwa Liu; Rachel Lampert; Albert J Sinusas; Chi Liu
Journal:  Med Phys       Date:  2016-01       Impact factor: 4.071