Yansong Li1, Ying Xie, Bailong Hu, Jing Wang, Haibo Song, Xinchuan Wei.
Abstract
This study aimed to explore the sensitivity and specificity an approach to confirm the depth of the endotracheal tube. It was a prospective, diagnostic study (Registration number: chiCTR-TRC-14005170) conducted in the Department of Anesthesiology of West China Hospital. A total of 100 patients underwent general anesthesia were included. The distance between the upper margin of the cuff and the manubriosternal joint (CM) and the distance between the upper margin of the cuff and the carina (CC) were measured. The accuracy of the confirmed approach against fiberoptic bronchoscopy (FOB) was evaluated. Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) were calculated. The correlation and agreement between CM and CC were explored.The sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) were determined to be 94.74% (95% CI: 87.07-98.55%), 79.17% (95% CI: 57.85-92.87%), 93.51% (95% CI: 85.49-97.86%), 82.61% (95% CI: 61.22%-95.05%), respectively. The κ value was 0.75 (95% CI: 0.60-0.90). There was a significant correlation between CM and CC (P < 0.001) with a correlation coefficient of 0.91(95% CI: 0.87-0.94). Bland and Altman plots also demonstrated a good agreement between the CM and CC. This novel method may be used to confirm the depth of the endotracheal tubes.Entities:
Mesh:
Year: 2015 PMID: 26313810 PMCID: PMC4602919 DOI: 10.1097/MD.0000000000001460
Source DB: PubMed Journal: Medicine (Baltimore) ISSN: 0025-7974 Impact factor: 1.817
FIGURE 1The left picture shows cuff without saline. The right picture shows cuff with saline. The distance between the upper margin of the cuff and the suprasternal notch was measured. The white arrow indicates the wire reinforcement. Cephalad is to the right, and caudad is to the left.
FIGURE 2CS: the distance between the upper margin of the cuff and the suprasternal notch; SM: the distance between the suprasternal notch and manubriosternal joint; CM: the distance between the upper margin of the cuff and the manubriosternal joint; CT: the distance between the upper margin of the cuff and the tip of the endotracheal tube; TC: the distance between the tip of the ETT and the carina; CC: the distance between the upper margin of the cuff and the carina.
FIGURE 3FOB = fiberoptic bronchoscopy.
The Characteristics of the Participated Patients
2 × 2 Table of Comparison of the Novel Approach versus that of FOB
Test Characteristics of the Approach for Proper Depth of the Endotracheal Tube
FIGURE 4The correlation between CM and CC (R2 = .831, r = 0.91, P < 0.001). CM: the distance between the upper margin of the cuff and the manubriosternal joint; CC: the distance between the upper margin of the cuff and the carina; b: the agreement between CM and CC. [bias: 0.37cm (95% CI: 0.31–0.44) standard deviation (SD): 0.32, limits of agreement: −0.26, 1.00; and percentage of error: 7.4%] ○: female, ●: male.