Literature DB >> 6816125

Heat sink protection against lasering endotracheal cuffs.

F E LeJeune, C Guice, F LeTard, H Marice.   

Abstract

The ignition of an endotracheal cuff by CO2 laser impaction has proved troublesome in several ignition sequences. Metal foil tape wrapped around the shaft of the tube will provide adequate protection for it, but the tape does not extend to the thin-walled cuff. When normal saline solution is used instead of air to fill the cuff, heat is rapidly conducted away from the wall and the cuff is prevented from reaching its kindling temperature. If penetration should occur, a fine jet of saline solution instantaneously sprays the field, extinguishing any early combustion. The reaction time of the surgeon or an assistant is not a factor in prevention of thermal injury, no new equipment is required and preparation time is minimal.

Entities:  

Mesh:

Substances:

Year:  1982        PMID: 6816125     DOI: 10.1177/000348948209100617

Source DB:  PubMed          Journal:  Ann Otol Rhinol Laryngol        ISSN: 0003-4894            Impact factor:   1.547


  5 in total

1.  Treatment with isoproterenol of bupivacaine toxicity.

Authors:  P Lacombe; G Blaise; F Plante; C Hollmann
Journal:  Can J Anaesth       Date:  1990-05       Impact factor: 5.063

Review 2.  [Specific aspects of anesthesiological management of laser surgery in otorhinolaryngology].

Authors:  W A Wetsch; D Beutner; K J Stuermer; S A Padosch
Journal:  Anaesthesist       Date:  2014-06       Impact factor: 1.041

3.  [Specific aspects of anesthesiological management of laser surgery in otorhinolaryngology].

Authors:  W A Wetsch; D Beutner; K J Stuermer; S A Padosch
Journal:  HNO       Date:  2014-03       Impact factor: 1.284

4.  Laser safety in head and neck cancer surgery.

Authors:  Ferhan Ahmed; Andrew J Kinshuck; Michael Harrison; Dan O'Brien; Jeffrey Lancaster; Nicholas J Roland; Shaun R Jackson; Terrence M Jones
Journal:  Eur Arch Otorhinolaryngol       Date:  2010-07-21       Impact factor: 2.503

5.  Sensitivity and Specificity of a Novel Approach to Confirm the Depth of the Endotracheal Tube: A Pilot Study.

Authors:  Yansong Li; Ying Xie; Bailong Hu; Jing Wang; Haibo Song; Xinchuan Wei
Journal:  Medicine (Baltimore)       Date:  2015-08       Impact factor: 1.817

  5 in total