Cook and Chill: Effect of Temperature on the Performance of Nonequilibrated Blood Glucose Meters.

Sherine Deakin1, Dominic Steele2, Sarah Clarke2, Cathryn Gribben2, Anne-Marie Bexley2, Remmert Laan3, David Kerr3.   

Abstract

BACKGROUND: Exposure to extreme temperature can affect the performance of blood glucose monitoring systems. The aim was to determine the non-equilibrated performance of these systems at extreme high and low temperatures that can occur in daily life.
METHODS: The performances of 5 test systems, (1) Abbott FreeStyle Freedom Lite, (2) Roche AccuChek Aviva, (3) Bayer Contour, (4) LifeScan OneTouch Verio, and (5) Sanofi BG Star, were compared after "cooking" (50°C for 1 hour) or "chilling" (-5°C for 1 hour) with room temperature controls (23°C) using whole blood with glucose concentrations of 50, 100, and 200 mg/dl.
RESULTS: The equilibration period (time from the end of incubation to when the test system is operational) was between 1 and 8 minutes, and each test system took between 15 and 30 minutes after incubation to obtain stable measurements at room temperature. Incubating the strips at -5°C or 50°C had little effect on the glucose measurement, whereas incubating the meters introduced bias in performance between 0 and 15 minutes but not subsequently, compared to room temperature controls and at all 3 glucose levels.
CONCLUSIONS: Compensating technologies embedded within blood glucose monitoring systems studied here perform well at extreme temperatures. People with diabetes need to be alerted to this feature to avoid perceptions of malperformance of their devices and the possible inability to get blood glucose readings on short notice (eg, during time of suspected rapid change or before an unplanned meal).
© 2015 Diabetes Technology Society.

Entities:  

Keywords:  SMBG; glucose; meters; temperature

Mesh:

Substances:

Year:  2015        PMID: 26292960      PMCID: PMC4667317          DOI: 10.1177/1932296815598775

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  10 in total

1.  Diabetes in the desert: what do patients know about the heat?

Authors:  Adrienne A Nassar; Raymond D Childs; Mary E Boyle; Kimberly A Jameson; Margaret Fowke; Ken R Waters; Michael J Hovan; Curtiss B Cook
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

Review 2.  Stability and performance of rapid-acting insulin analogs used for continuous subcutaneous insulin infusion: a systematic review.

Authors:  David Kerr; Erik Wizemann; Jakob Senstius; Mette Zacho; Francisco Javier Ampudia-Blasco
Journal:  J Diabetes Sci Technol       Date:  2013-11-01

Review 3.  Geoenvironmental diabetology.

Authors:  Curtiss B Cook; Kay E Wellik; Margaret Fowke
Journal:  J Diabetes Sci Technol       Date:  2011-07-01

Review 4.  Factors affecting blood glucose monitoring: sources of errors in measurement.

Authors:  Barry H Ginsberg
Journal:  J Diabetes Sci Technol       Date:  2009-07-01

5.  Effect of ambient temperature on analytical performance of self-monitoring blood glucose systems.

Authors:  Kari Nerhus; Pål Rustad; Sverre Sandberg
Journal:  Diabetes Technol Ther       Date:  2011-06-29       Impact factor: 6.118

6.  Heat stress from enclosed vehicles: moderate ambient temperatures cause significant temperature rise in enclosed vehicles.

Authors:  Catherine McLaren; Jan Null; James Quinn
Journal:  Pediatrics       Date:  2005-07       Impact factor: 7.124

Review 7.  When it rains, it pours: future climate extremes and health.

Authors:  Jonathan A Patz; Maggie L Grabow; Vijay S Limaye
Journal:  Ann Glob Health       Date:  2014-11-25       Impact factor: 2.462

8.  Personal medicines storage in New Zealand.

Authors:  Campbell Hewson; Chong Chi Shen; Clare Strachan; Pauline Norris
Journal:  J Prim Health Care       Date:  2013-06-01

Review 9.  Managing diabetes in the heat: potential issues and concerns.

Authors:  Sydney A Westphal; Raymond D Childs; Karen M Seifert; Mary E Boyle; Margaret Fowke; Paul Iñiguez; Curtiss B Cook
Journal:  Endocr Pract       Date:  2010 May-Jun       Impact factor: 3.443

Review 10.  Health risks at the Hajj.

Authors:  Qanta A Ahmed; Yaseen M Arabi; Ziad A Memish
Journal:  Lancet       Date:  2006-03-25       Impact factor: 79.321

  10 in total
  3 in total

1.  Assessment of System Accuracy, Intermediate Measurement Precision, and Measurement Repeatability of a Blood Glucose Monitoring System Based on ISO 15197.

Authors:  Nina Jendrike; Annette Baumstark; Stefan Pleus; Jochen Mende; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-12-14

2.  Evaluation of Analytical Performance of Three Blood Glucose Monitoring Systems: System Accuracy, Measurement Repeatability, and Intermediate Measurement Precision.

Authors:  Stefan Pleus; Nina Jendrike; Annette Baumstark; Jochen Mende; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-10-05

Review 3.  Interferences and Limitations in Blood Glucose Self-Testing: An Overview of the Current Knowledge.

Authors:  Michael Erbach; Guido Freckmann; Rolf Hinzmann; Bernhard Kulzer; Ralph Ziegler; Lutz Heinemann; Oliver Schnell
Journal:  J Diabetes Sci Technol       Date:  2016-08-22
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.