Evaluating the image quality of cone beam CT acquired during rotational delivery.

S A Yoganathan1, K J Maria Das1, K Maria Midunvaleja2, D Gowtham Raj1, Arpita Agarwal1, J Velmurugan2, Shaleen Kumar1.   

Abstract

OBJECTIVE: The aim of this work was to evaluate the quality of kilovoltage (kV) cone beam CT (CBCT) images acquired during arc delivery.
METHODS: Arc plans were delivered on a Catphan(®) 600 phantom (The Phantom Laboratory Inc., Salem, NY), and kV CBCT images were acquired during the treatment. The megavoltage (MV) scatter effect on kV CBCT image quality was evaluated using parameters such as Hounsfield unit (HU) accuracy, spatial resolution, contrast-to-noise ratio (CNR) and spatial non-uniformity (SNU). These CBCT images were compared with reference scans acquired with the same acquisition parameters without MV "beam on". This evaluation was carried out for different photon beams (6 and 15 MV), arc types (half vs full arc), static field sizes (10 × 10 and 25 × 25 cm(2)) and source-to-imager distances (SID) (150 and 170 cm). RESULTS AND
CONCLUSION: HU accuracy, CNR and SNU were considerably affected by MV scatter, and this effect was increased with increasing field size and decreasing photon energy, whereas the spatial resolution was almost unchanged. The MV scatter effect was observed to be more for full-rotation arc delivery than for half-arc delivery. In addition, increasing the SID resulted in decreased MV scatter effect and improved the image quality. ADVANCES IN KNOWLEDGE: Nowadays, volumetric modulated arc therapy (VMAT) is increasingly used in clinics, and this arc therapy enables us to acquire CBCT imaging simultaneously. But, the main issue of concurrent imaging is the "MV scatter" effect on CBCT imaging. This study aims to experimentally quantify the effect of MV scatter on CBCT image quality.

Entities:  

Mesh:

Year:  2015        PMID: 26226396      PMCID: PMC4730980          DOI: 10.1259/bjr.20150425

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  18 in total

1.  Cone-beam computed tomography with a flat-panel imager: initial performance characterization.

Authors:  D A Jaffray; J H Siewerdsen
Journal:  Med Phys       Date:  2000-06       Impact factor: 4.071

2.  Flat-panel cone-beam computed tomography for image-guided radiation therapy.

Authors:  David A Jaffray; Jeffrey H Siewerdsen; John W Wong; Alvaro A Martinez
Journal:  Int J Radiat Oncol Biol Phys       Date:  2002-08-01       Impact factor: 7.038

3.  Grids or air gaps for scatter reduction in digital radiography: a model calculation.

Authors:  U Neitzel
Journal:  Med Phys       Date:  1992 Mar-Apr       Impact factor: 4.071

4.  Verification of in-treatment tumor position using kilovoltage cone-beam computed tomography: a preliminary study.

Authors:  Keiichi Nakagawa; Hideomi Yamashita; Kenshiro Shiraishi; Hiroshi Igaki; Atsuro Terahara; Naoki Nakamura; Kuni Ohtomo; Shigeki Saegusa; Takashi Shiraki; Takashi Oritate; Kiyoshi Yoda
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-11-15       Impact factor: 7.038

5.  Combined kV and MV imaging for real-time tracking of implanted fiducial markers.

Authors:  R D Wiersma; Weihua Mao; L Xing
Journal:  Med Phys       Date:  2008-04       Impact factor: 4.071

6.  Determining appropriate imaging parameters for kilovoltage intrafraction monitoring: an experimental phantom study.

Authors:  D Wallace; J A Ng; P J Keall; R T O'Brien; P R Poulsen; P Juneja; J T Booth
Journal:  Phys Med Biol       Date:  2015-06-09       Impact factor: 3.609

7.  Prostate gland motion assessed with cine-magnetic resonance imaging (cine-MRI).

Authors:  Michel J Ghilezan; David A Jaffray; Jeffrey H Siewerdsen; Marcel Van Herk; Anil Shetty; Michael B Sharpe; Syed Zafar Jafri; Frank A Vicini; Richard C Matter; Donald S Brabbins; Alvaro A Martinez
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-06-01       Impact factor: 7.038

8.  Dosimetric effect of respiratory motion in external beam radiotherapy of the lung.

Authors:  James Mechalakos; Ellen Yorke; Gikas S Mageras; Agung Hertanto; Andrew Jackson; Ceferino Obcemea; Kenneth Rosenzweig; C Clifton Ling
Journal:  Radiother Oncol       Date:  2004-05       Impact factor: 6.280

9.  Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy.

Authors:  Yvette Seppenwoolde; Hiroki Shirato; Kei Kitamura; Shinichi Shimizu; Marcel van Herk; Joos V Lebesque; Kazuo Miyasaka
Journal:  Int J Radiat Oncol Biol Phys       Date:  2002-07-15       Impact factor: 7.038

10.  Measurement of the interplay effect in lung IMRT treatment using EDR2 films.

Authors:  Ross I Berbeco; Cynthia J Pope; Steve B Jiang
Journal:  J Appl Clin Med Phys       Date:  2006-11-28       Impact factor: 2.102

View more
  5 in total

1.  Deep learning-based image quality improvement for low-dose computed tomography simulation in radiation therapy.

Authors:  Tonghe Wang; Yang Lei; Zhen Tian; Xue Dong; Yingzi Liu; Xiaojun Jiang; Walter J Curran; Tian Liu; Hui-Kuo Shu; Xiaofeng Yang
Journal:  J Med Imaging (Bellingham)       Date:  2019-10-24

2.  Cone beam computed tomography and its image guidance technology during percutaneous nucleoplasty procedures at L5/S1 lumbar level.

Authors:  Anna Maria Ierardi; Filippo Piacentino; Francesca Giorlando; Alberto Magenta Biasina; Alessandro Bacuzzi; Raffaele Novario; Gianpaolo Carrafiello
Journal:  Skeletal Radiol       Date:  2016-10-03       Impact factor: 2.199

3.  Image quality of 4D in-treatment CBCT acquired during lung SBRT using FFF beam: a phantom study.

Authors:  Jihun Kim; Ki Chang Keum; Ho Lee; Chae-Seon Hong; Kwangwoo Park; Jin Sung Kim
Journal:  Radiat Oncol       Date:  2020-09-25       Impact factor: 3.481

4.  Image quality evaluation of intra-irradiation cone-beam computed tomography acquired during one- and two-arc prostate volumetric-modulated arc therapy delivery: A phantom study.

Authors:  Hiraku Iramina; Ayaka Kitamura; Mitsuhiro Nakamura; Takashi Mizowaki
Journal:  J Appl Clin Med Phys       Date:  2020-11-16       Impact factor: 2.102

5.  Kilovoltage projection streaming-based tracking application (KiPSTA): First clinical implementation during spine stereotactic radiation surgery.

Authors:  Jihun Kim; Yang-Kyun Park; David Edmunds; Kevin Oh; Gregory C Sharp; Brian Winey
Journal:  Adv Radiat Oncol       Date:  2018-06-07
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.