The aims of this study were to quantify patient radiation exposure for a single interventional procedure during transition from an adult catheterization laboratory to a next-generation imaging system with pediatric settings, and to compare this radiation data to published benchmarks. Radiation exposure occurs with any X-ray-directed pediatric catheterization. Technologies and imaging techniques that limit dose while preserving image quality benefit patient care. Patient radiation dose metrics, air kerma, and dose-area product (DAP) were retrospectively obtained for patients <20 kg who underwent patent ductus arteriosus (PDA) closure on a standard imaging system (Group 1, n = 11) and a next-generation pediatric imaging system (Group 2, n = 10) with air-gap technique. Group 2 radiation dose metrics were then compared to published benchmarks. Patient demographics, procedural technique, PDA dimensions, closure devices, and fluoroscopy time were similar for the two groups. Air kerma and DAP decreased by 65-70% in Group 2 (p values <0.001). The average number of angiograms approached statistical significance (p value = 0.06); therefore, analysis of covariance (ANCOVA) was conducted that confirmed significantly lower dose measures in Group 2. This degree of dose reduction was similar when Group 2 data (Kerma 28 mGy, DAP 199 µGy m(2)) was compared to published benchmarks for PDA closure (Kerma 76 mGy, DAP 500 µGy m(2)). This is the first clinical study documenting the radiation reduction capabilities of a next-generation pediatric imaging platform. The true benefit of this dose reduction will be seen in patients requiring complex and often recurrent catheterizations.

Entities:  

Keywords:  Angiography; Pediatric catheterization; Radiation reduction

Mesh:

Year:  2015        PMID: 26215767     DOI: 10.1007/s00246-015-1233-2

Source DB:  PubMed          Journal:  Pediatr Cardiol        ISSN: 0172-0643            Impact factor:   1.655


  16 in total

1.  Radiation dose reduction without compromise of image quality in cardiac angiography and intervention with the use of a flat panel detector without an antiscatter grid.

Authors:  J Partridge; G McGahan; S Causton; M Bowers; M Mason; M Dalby; A Mitchell
Journal:  Heart       Date:  2005-09-13       Impact factor: 5.994

2.  Patient-specific dose and radiation risk estimation in pediatric cardiac catheterization.

Authors:  Klaus Bacher; Evelien Bogaert; Régine Lapere; Daniël De Wolf; Hubert Thierens
Journal:  Circulation       Date:  2004-12-20       Impact factor: 29.690

3.  The ICRP 2007 recommendations.

Authors:  C Streffer
Journal:  Radiat Prot Dosimetry       Date:  2007-10-12       Impact factor: 0.972

4.  Radiation reduction in pediatric and adult congenital patients during cardiac catheterization.

Authors:  Daniel A Mauriello; Kenneth A Fetterly; Ryan J Lennon; Guy S Reeder; Nathaniel W Taggart; Donald J Hagler; Frank Cetta; Allison K Cabalka
Journal:  Catheter Cardiovasc Interv       Date:  2014-06-20       Impact factor: 2.692

5.  Use of a dose-dependent follow-up protocol and mechanisms to reduce patients and staff radiation exposure in congenital and structural interventions.

Authors:  Jaclynn M Sawdy; Tanya Maria Kempton; Vincent Olshove; Mark Gocha; Joanne L Chisolm; Sharon L Hill; Amy Kirk; John P Cheatham; Ralf J Holzer
Journal:  Catheter Cardiovasc Interv       Date:  2011-07-01       Impact factor: 2.692

6.  Dose optimization in pediatric cardiac x-ray imaging.

Authors:  Amber J Gislason; Andrew G Davies; Arnold R Cowen
Journal:  Med Phys       Date:  2010-10       Impact factor: 4.071

7.  Characterization of radiation exposure and effect of a radiation monitoring policy in a large volume pediatric cardiac catheterization lab.

Authors:  George R Verghese; Doff B McElhinney; Keith J Strauss; Lisa Bergersen
Journal:  Catheter Cardiovasc Interv       Date:  2011-12-08       Impact factor: 2.692

8.  Acute chromosomal DNA damage in human lymphocytes after radiation exposure in invasive cardiovascular procedures.

Authors:  Maria Grazia Andreassi; Angelo Cioppa; Samantha Manfredi; Cataldo Palmieri; Nicoletta Botto; Eugenio Picano
Journal:  Eur Heart J       Date:  2007-06-28       Impact factor: 29.983

9.  Optimization of radiographic parameters for paediatric cardiac angiography.

Authors:  D G W Onnasch; A Schemm; H-H Kramer
Journal:  Br J Radiol       Date:  2004-06       Impact factor: 3.039

Review 10.  The ALARA concept in pediatric cardiac catheterization: techniques and tactics for managing radiation dose.

Authors:  Henri Justino
Journal:  Pediatr Radiol       Date:  2006-09
View more
  4 in total

1.  Intraoperative Completion Angiogram May Be Superior to Transesophageal Echocardiogram for Detection of Pulmonary Artery Residual Lesions in Congenital Heart Surgery.

Authors:  Luke Lamers; Erick E Jimenez; Catherine Allen; Derreck Hoyme; Entela Bua Lushaj; Petros V Anagnostopoulos
Journal:  Pediatr Cardiol       Date:  2018-03-10       Impact factor: 1.655

2.  Radiation Dose Reduction in Congenital Heart Disease Patients During Cardiac Catheterization by a Novel Protocol.

Authors:  Selman Gokalp; Ibrahim Cansaran Tanidir; Erkut Ozturk; Yakup Ergul; Alper Guzeltas
Journal:  Turk Arch Pediatr       Date:  2021-07-01

Review 3.  Recent advances in cardiac catheterization for congenital heart disease.

Authors:  Sok-Leng Kang; Lee Benson
Journal:  F1000Res       Date:  2018-03-26

4.  Multicenter Assessment of Radiation Exposure during Pediatric Cardiac Catheterizations Using a Novel Imaging System.

Authors:  Luke J Lamers; Brian H Morray; Alan Nugent; Michael Speidel; Petch Suntharos; Lourdes Prieto
Journal:  J Interv Cardiol       Date:  2019-10-31       Impact factor: 2.279

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.