Radiation Dose Reduction in Congenital Heart Disease Patients During Cardiac Catheterization by a Novel Protocol.

Selman Gokalp1, Ibrahim Cansaran Tanidir1, Erkut Ozturk1, Yakup Ergul1, Alper Guzeltas1.   

Abstract

OBJECTIVE: Cardiac catheterization remains a major source of radiation exposure for patients with congenital heart disease. This study reports the magnitude of radiation with a 3.75 frame per second (fps) pulse fluoroscopy rate and compares the reduction with the previous 15 fps protocol during cardiac catheterization for pediatric and adult congenital heart disease.
MATERIAL AND METHODS: All diagnostic and interventional cardiac catheterization procedures from a single tertiary center were analyzed from January 1, 2014 to December 31, 2015, one year before and after implementing lower starting pulse fluoroscopy rates. The radiation dose was quantified as air kerma dose (mGy) and dose-area product (DAP; µGy/m2). Radiation exposure was analyzed for diagnostic and interventional procedures; the diagnostic group was subdivided into cyanotic and acyanotic patients, whereas the interventional group was subdivided according to the most common indications.
RESULTS: A total of 786 procedures were analyzed. The median fluoroscopy times and contrast amounts did not show a statistically significant difference between both periods (487 vs. 456 seconds and 42.5 vs. 45.3 cm3). The median air kerma for all procedures showed an 88% reduction after implementing lower pulse fluoroscopy rates (340-41 mGy). The doses were reduced significantly for diagnostic and interventional angiograms from 470 mGy and 162 mGy to 40 mGy and154 mGy. Among all patient groups, the most striking decrease was observed in the diagnostic procedures we use, of which fluoroscopy is more prominent than cine angiography.
CONCLUSION: We claim that novel radiation dose reduction protocols could be easily applied without increasing fluoroscopy time or losing image quality. © Copyright 2021 by The Turkish Archives of Pediatrics.

Entities:  

Keywords:  cardiac catheterization; congenital heart disease; radiation dose

Year:  2021        PMID: 35005727      PMCID: PMC8655960          DOI: 10.5152/TurkArchPediatr.2021.20068

Source DB:  PubMed          Journal:  Turk Arch Pediatr        ISSN: 2757-6256


Introduction

Cardiac catheterization is still one of the most important sources of radiation exposure in patients with congenital heart disease (CHD). Catheter interventions expose these groups of patients to high radiation doses due to the technically challenging procedures (patent ductus arteriosus stenting, percutaneous pulmonary valve implantation, etc.), complex anatomic features, manipulation difficulties, prolonged procedure time, heart rate, and frequent use of magnification result. In addition, the radiation dose may be further increased depending on the operative-dependent (distance between the patient and device) and the operative-independent (angiography device, image intensifier factors, the nature of procedure) factors. CHD patients are more prone to deterministic (direct dose–response relationship) and stochastic radiation effects due to the growing organism's biological properties. A stochastic effect is one in which the probability of the effect, rather than its severity, increases with dose. Radiation-induced cancer and genetic injuries are stochastic. Considering the long life expectancy of children and the cumulative radiation doses that patients and laboratory workers are exposed to during catheter angiography, both groups are at risk of radiation's deleterious effects such as immune dysfunction, cataract, and congenital anomalies and malignancy. In order to minimize this risk, safe and effective new methods are required that reduce fluoroscopy time and radiation dose. These new methods aim to provide the most accurate diagnostic and therapeutic benefit with the lowest possible radiation dose by applying as low as reasonably achievable (ALARA) concept to cardiac catheterization, as radiation has no known safe-dose range for the patient and health personnel. The second aim is to reduce the radiation dose without compromising the image quality. The use of low pulse rate fluoroscopy to achieve these 2 objectives was reported in a limited number of studies. In this study, we determined radiation doses using a 3.75 frame rate per second (fps) pulse rate (lowest possible pulse rate) fluoroscopy during cardiac catheterization of patients with CHD, both children and adults. The results were compared with the 15 fps standard pulse rate fluoroscopy findings that we previously used. What is already known on this topic? Cardiac catheterization is still one of the most important sources of radiation exposure in patients with congenital heart disease. In order to minimize this risk, safe and effective new radiation exposure reducing methods are required . What this study adds on this topic? Implementing lower fluoroscopy rates could minimize radiation’s possible side effects, concomitantly increasing the safety of the patient and the health care provider without increasing the total fluoroscopy time and the amount of contrast.

Materials and Methods

We evaluated all the data of CHD patients who underwent cardiac catheterization between January 1, 2014 and December 31, 2015. All data were retrospectively obtained from the pediatric cardiology department's angiographic database. Fluoroscopic procedures for pericardiocentesis, central catheter insertion, electrophysiology studies, and hybrid cases were excluded from the study. A total of 4 primary interventional pediatric cardiology staff were involved in the study. A descriptive table including age, weight, sex, echocardiographic diagnosis, procedure and intervention type (if performed), fluoroscopy time, procedure time, and radiation doses was obtained. The catheter angiographies performed between January 1, 2015 and December 31, 2015 at 3.75 fps were defined as group I. Catheter angiographies. The standard 15 fps method was used between January 1, 2014 and December 31, 2014 formed group II. Measurements of radiation dose are reported for each case by the catheterization system of Philips Allura Xper FD20/10® (Philips Medical Systems, Eindhoven, Netherlands), and categorized as those that are obtained through fluoroscopy alone versus those obtained through digital acquisition. To monitor radiation exposure, the patient dose was indirectly recorded using standard techniques, including total fluoroscopy time (minutes), air kerma (mGy), and dose-area product (DAP; µGy m ). Air kerma dose is the dose measured in air at a fixed distance from the X-ray tube and is the best surrogate of the radiation absorbed at the skin surface at the site of beam entrance. It is correlated with the risk of skin injury; doses > 2000 mGy at a single skin site are known to increase the risk of acute skin injury. DAP; µGy m ) is the instantaneous air kerma dose times the X-ray field area, reflecting the total dose given to the patient. Radiation exposure was analyzed for diagnostic and interventional procedures. The diagnostic group was subdivided into cyanotic and acyanotic patients (>30 procedures). The interventional group was subdivided according to the most common indications (>15 procedures). To minimize the effect of body weight and age, patient groups were subdivided according to weight and age. There were 6 groups, according to the body weight of the patients, as 0-5, 5-15, 15-40, 40-55, 55-70, and >70 kg; the patients were also divided into 6 age groups, as newborn (0-30 days), infants (1-12 months), 1-5 years, 5-10 years, 10-15 years, and >15 years.

Statistical Analysis

The Statistical Package for the Social Sciences for Windows (SPSS) Version 15 (SPSS, Chicago, IL, USA) was used for the statistical analyses. The distribution of each continuous variable was tested for normality using the Kolmogorov–Smirnov test. Non-parametric tests were used in cases where normality was not provided. Continuous variables are expressed as the median and interquartile range (IQR, first, and third quartiles); categorical variables are expressed as percentages. The Mann–Whitney U-test was used to compare the 2 groups' median values, while the chi-square and Fisher's exact tests were used to compare the findings between groups. P values < .05 were considered statistically significant.

Ethical Standards

The authors assert that all procedures contributing to this work comply with the ethical standards of the relevant national guidelines on human experimentation and the Helsinki Declaration of 1975, as revised in 2008, and approved by the institutional committees (Reference Number 2020-46). Written informed consent was obtained from all children and/or their parents.

Results

A total of 786 cases were included in the study. Among them, 415 patients were studied with 15 fps and 371 patients with 3.75 fps. In the first year of study, the median air kerma for all procedures was 340 mGy; it was decreased to 41 mGy (≈88%). The most striking decrease was observed in diagnostic procedures (470 mGy vs. 40 mGy), of which we use fluoroscopy more prominently than cine angiography. Although the median procedure (30 min vs. 30 min) and fluoroscopy times did not show a statistically significant difference for both periods (487 seconds vs. 456 seconds), DAP had changed dramatically (4731 µGy m2 vs. 3149 µGy m ). Again, the most important decrease was seen in the diagnostic group (9512 µGy m2 vs. 3490 µGy m ). Moreover, the mean contrast amount was almost identical without increasing the contrast for better visualization of the anatomic structure (42.5 cm vs. 45.3 cm ). Basic demographic and procedural data for the study cohort are summarized in Table 1.
Table 1.

Comparison of Patient Groups According to Pulse Fluoroscopy Rates

Characteristics15 fps(n = 371)3.75 fps(n = 415) P
Age at procedure (month) (median, IQR) 16.6 (7.2-69.8)22.3 (7.6-73)NS
Weight (kg) (median, IQR)9.8 (6.7-19.4)10 (6.2-20)NS
Body surface area (m2) (median, IQR)0.46 (0.36-0.76)0.48 (0.34-0.78)NS
Sex (Male/Female)193/178224/191NS
Procedure type, n (%)* NS
 Diagnostic 250(60)
 Interventional 165(40)
Procedure time minutes (median, IQR)30 (20-50)30 (20-45)NS
Fluoroscopy time seconds (median, IQR)487 (283-766)456 (283-679)NS
Air kerma (mGy) total (median, IQR)340 (190-460)41 (25.6-79.1) .001
DAP (µGy m2) total (median, IQR)4731 (2627-9388)3149 (1936-6102) .04
Air kerma (mGy) diagnostic (median, IQR)470 (310-550)40 (27-71) .02
DAP (µGy m2) diagnostic (median, IQR)9512 (3509-14 125)3490 (2250-6340) .001
Air kerma (mGy) interventional (median, IQR)162 (81-235)154 (92-220)NS
DAP (µGy.m2) interventional (median, IQR)11 325 (3420-18 952)8623 (2786-6874) .01

DAP, dose-area product; IQR, interquartile range.

Table 2 summarizes the radiation doses of the patients according to age distribution. According to age groups, there was a statistically significant decrease in radiation dose for newborns and infants.
Table 2.

Radiation Exposure According to Procedure Type and Age Groups

NewbornInfant1-5 years5-10 years10-15 years>15 years
15 fps3.75 fps15 fps3.75 fps15 fps3.75 fps15 fps3.75 fps15 fps3.75 fps15 fps3.75 fps
All procedures (N)2726118130124139404831383034
Diagnostic (N)6780849010417209162119
Interventional (N)21193846343524282222915
Air kerma (mGy), All procedures87591681091034482713891202624354
Air kerma (mGy), Diagnostic505417345684383668641313645273
Air kerma (mGy), Interventional10063163227205618174195110245455
DAP (µGy m2) All procedures46042745*4540*3320*818742608928899623 89011 98535 923*32 169
DAP (µGy m2) Diagnostic220028624771313662563938996511 52520 87712 96538 15826 754
DAP (µGy m2) Interventional542827443731343012 96147048051695824 94911 06427 78240 758
Fluoroscopy time972882571588503480574559855584652837

*Values are given as median. Interquartile ranges are not given, for simplicity. P values are given only for statistically significant results (P < .05). Newborns DAP P = .045, infants DAP P = .004, 10-15 age group DAP P = .027.

Table 3 summarizes the radiation doses of the patients according to weight distribution. Air kerma and DAP values were ­significantly decreased, especially in 15-40 kg group (P = .023 and P = .04, respectively).
Table 3.

Radiation Exposure According to Body Weight

0-5 kg5-15 kg15-40 kg40-55 kg55-70 kg>70 kg
15 fps3.75 fps15 fps3.75 fps15 fps3.75 fps15 fps3.75 fps15 fps3.75 fps15 fps3.75 fps
All procedures (N)526620221476871926131795
Diagnostic (N)2223145162334211138743
Interventional (N)30435752434581351052
Air kerma (mGy) All procedures140566685118 71 * 41262092413381906551
Air kerma (mGy) Diagnostic47476545118657321782933822708351
Air kerma (mGy) Interventional2066169210118761242401583071261851
DAP (µGy m2) All procedures387228506510444213 750*7195*21 27220 19527 60631 96953 50549 916
DAP (µGy m2) Diagnosti 286127165550435513 808677126 81917 57632 17137 92759 17034 425
DAP (µGy m2) Interventional459129109011465613 221722713 58022 79520 27327 79148 93773 138
Fluoroscopy time755725534527691517572675701928405585

* P values are given only for statistically significant results (P < .05). 15-40 kg air kerma P = 0.023. DAP P = .04.

Table 4 summarizes the radiation doses according to the subgroup of diagnostic procedures, which includes at least 30 cases. There was a striking reduction in dosage in radiation performed for evaluation before Glenn and Fontan operation (P < .05), especially in angiographies.
Table 4.

Radiation Exposure According to Subgroups (Diagnostic Catheterizations)

DiagnosticProceduresTetralogy of FallotPre-FontanPre-Glenn
15 fps3.75 fps P 15 fps3.75 fps P 15 fps3.75 fps P
Patient (n)4337.971430.634738.83
Age (months), (median, IQR)12(4-17)11(5-17).9239(30-45)38(29-42).936(4-8)5(4-7).32
Weight (median, IQR)7(2.9-0.8)8(3-9.6).4325(18-32)21(10-29).477.6(5.6-9.5)8.1(5.9-10.6).48
BMI (median, IQR)0.4(0.2-0.5)0.4(0.2-0.5).300.9(0.4-1.2)0.8(0.4-1.1).450.4(0.3-0.6)0.4(0.3-0.5).53
Procedure time (minutes) (median, IQR)26(15-36)21(12-30).0845(22-58)32(25-39).00635(21-47)34(22-45).59
Fluoroscopy time (seconds) (median, IQR)359(80-430)348(110-400).81750(240-960)570(310-830).08640(215-852)611(201-811).75
Air kerma (mGy) (median, IQR)59(10-92)35(14-84).07154(30-220)81(20-180).0565(28-96)38(16-51).001
DAP (µGy m2) (median, IQR)3755(1211-5320)3018(950-4839).1217 159(2300-21 236)8650(1954-14 589).025400(2700-7600)3200(2100-6100).001
Air kerma (mGy/kg) (median, IQR)7.2(1.5-18.3)5.2(2.1-12.6).105.9(2.1-8)3.7(2-4.8).0089.3(3.5-12.7)5.4(2.3-8.1).004
DAP (µGy m2/kg) (median, IQR)51(20-80)39(16-60).18120(25-150)70(30-142).00877(25-96)42(15-80).001
Air kerma (mGy/BMI) (median, IQR)155(18-210)102(25-195).10150(60-205)90(50-160).01185(59-269)98(40-203).001
DAP (µGy m2/BMI) (median, IQR)9326(3952-13 598)7522(3125-12 369).2019 102(6325-23 587)10 256(7589-17 589).00213 500(4300-18 900)8100(5200-12 450).001

Values are given as median and interquartile ranges.

Table 5 summarizes the radiation doses according to the subgroup of interventional procedures, which includes at least 15 cases. It was determined that all radiation dose parameters decreased significantly, especially in atrial septal defect (ASD) and patent ductus arteriosus (PDA) closure procedures (P < .05).
Table 5.

Radiation Exposure According to the Subgroups (Interventional Catheterizations)

Interventional ProceduresAS BalloonPS BalloonCoA StentCoA BalloonPDA StentingASD ClosurePDA Closure
15 fps 3.75 fps 15 fps 3.75 fps 15 fps 3.75 fps 15 fps 3.75 fps 15 fps 3.75 fps 15 fps3.75 fps15 fps3.75 fps
Patient 157153096916131119293136
Age (months) 18.030.13043.81561353543.721.21331154863
Weight 8.911.21313.8504012.414.443.440331819
BMI 0.420.450.50.531.41.30.570.60.230.21.210.70.7
Procedure time 3633.52936.141.842.533.838.4586626253226.2
Fluoroscopy time (seconds)474490473562650615347504120011203723005.65.9
Air kerma (mGy) 50451088224012961723366956*18*187*30*
DAP (µGy m2)3706353312 240686425 92013 248480052325472345662402014*3552*2400*
Air kerma (mGy/kg) 42 28742 170842 22942 22742 08042 11242 318131212*1*5*2*
DAP (µGy m2/kg) 41033252949346130744246114401056216*72*240*142*
Air kerma (mGy/BMI) 13210916814815496129152192033653*22*154*45*
DAP (µGy m2/BMI) 8448641313 20010 56016 12896969024936024 96016 3206000*2400*5520*3264*

Values are given as median. Interquartile ranges are not given for simplicity. * P < .05.

Discussion

In this study, radiation dose during fluoroscopy was significantly decreased with a simple and easily applicable protocol—changing 15 fps to 3.75 fps—during diagnostic or interventional catheterization in patients with CHD. It was proved that radiation exposure could be significantly reduced without compromising image quality and increasing the amount of radiopaque used. Besides, both patients and healthcare personnel could be significantly protected from radiation exposure. Many patients with CHD undergo multiple cardiac catheterization procedures; they are exposed to ionizing radiation, which can have both immediate and long-term effects. The negative consequences of ionizing radiation can also be categorized as either deterministic or stochastic effects. While deterministic effects, like cataract formation or skin injury have a somewhat predictable dose–response relationship with the degree of injury directly correlating with absorbed radiation dose, stochastic effects, like cancer and genetic mutations, are unpredictable without a threshold effect. Strides have been made, particularly in the past decade, in improving radiation safety profiles as well as public awareness. Initiatives such as the ALARA conference, conducted by the Society for Pediatric Radiology in 2006, concluded that fluoroscopy dose optimization and reduction were critical areas of concern. Campaigns such as "Image Gently, Step Lightly," first launched in August of 2009, incorporated a standard safety checklist to encourage proper preparation, technique, and lower radiation exposure. In particular, this checklist encouraged utilizing pulse fluoroscopy rather than continuous fluoroscopy when possible, as well as using the lowest pulse rate possible. In their study, Covi et al. classified patients into 3 different groups, according to pulse fluoroscopy rates of 15, 7.5, and 5 frames. All 3 groups were equivalent in terms of difficulty, duration, and complication rate of procedures. They showed that reducing the frame rate from 15 to 7.5 fps significantly reduced radiation dose without compromising on image quality. There were no significant differences in physician-perceived ability to complete the procedure or impact of frame rate on the procedural length. Lamers et al. compared radiation doses during the standard imaging method and the new generation pediatric imaging method (10 fps) in their study, including 21 patients of less than 20 kg body weight undergoing PDA closure. Patient demographics, procedural technique, PDA dimensions, closure devices, and fluoroscopy time were similar for the 2 groups. Air kerma and DAP decreased by 65-70% by the new method (P values < .001). Recently, Amdani et al. showed that it is possible to reduce radiation exposure by lowering the frame rate in children undergoing cardiac catheterization without compromising the efficacy and safety of catheterization. They reported that fluoroscopy time, contrast volume, and complication rates did not increase, while diagnostic image quality was maintained. Boudjemline applied a similar principle by reducing the frame rate from 7.5 to 4 fps during transcatheter atrial septal defect closure, while maintaining excellent clinical results. He argued that there was no increase in the median procedure and fluoroscopic times or complications. Similarly, in our study, a significant decrease in air kerma and DAP levels was shown for diagnostic and interventional procedures. Although the link between high levels of radiation exposure and cancer risk is unequivocal, translating the relatively low level of radiation exposure from pediatric cardiac catheterization into a demonstrably increased cancer risk is more challenging. There was no demonstrable increase in cancer risk or cancer-related mortality either in an initial study of 4891 children exposed to pediatric cardiac catheterization or a later study of the same cohort with up to 35 years of follow-up. A separate study of 674 patients who underwent cardiac catheterization as children between 1950 and 1970 did find a significantly increased risk of lymphoma. Despite the difficulties in definitively proving an increased cancer risk after childhood radiation exposure, the theoretical possibility remains clear, as evidence for chromosomal damage has been seen immediately following cardiac catheterization.

Limitations

This single-institution, a retrospective study has several limitations. First, the air kerma and DAP doses reported here are directly reported from the X-ray system. It is important to understand that these measures reflect what is generated at the energy source and not necessarily what is actually absorbed by the patient. Effective dose and equivalent dose were not reported, but as simple logic, the reduction in generated radiation should be reflected as a reduction in both parameters.

Conclusion

In this study, we have demonstrated that by using the lowest possible fluoroscopy rate of 3.75 fps, the radiation dose can be significantly reduced during cardiac catheterization. This method effectively reduces the radiation dose, especially in diagnostic procedures where fluoroscopy is used extensively instead of cine angiography. Implementing this simple and effective radiation dose reduction protocol could minimize radiation's possible side effects, concomitantly increasing the safety of the patient and the health care provider without increasing the total fluoroscopy time and the amount of contrast.
  17 in total

1.  Estimated risks of radiation-induced fatal cancer from pediatric CT.

Authors:  D Brenner; C Elliston; E Hall; W Berdon
Journal:  AJR Am J Roentgenol       Date:  2001-02       Impact factor: 3.959

2.  Patient radiation exposure in a modern, large-volume, pediatric cardiac catheterization laboratory.

Authors:  Andrew C Glatz; Akash Patel; Xiaowei Zhu; Yoav Dori; Brian D Hanna; Matthew J Gillespie; Jonathan J Rome
Journal:  Pediatr Cardiol       Date:  2014-01-18       Impact factor: 1.655

3.  Effects of reducing frame rate from 7.5 to 4 frames per second on radiation exposure in transcatheter atrial septal defect closure.

Authors:  Younes Boudjemline
Journal:  Cardiol Young       Date:  2018-07-30       Impact factor: 1.093

4.  Cancer following cardiac catheterization in childhood.

Authors:  B Modan; L Keinan; T Blumstein; S Sadetzki
Journal:  Int J Epidemiol       Date:  2000-06       Impact factor: 7.196

5.  Characterization of radiation exposure and effect of a radiation monitoring policy in a large volume pediatric cardiac catheterization lab.

Authors:  George R Verghese; Doff B McElhinney; Keith J Strauss; Lisa Bergersen
Journal:  Catheter Cardiovasc Interv       Date:  2011-12-08       Impact factor: 2.692

Review 6.  Minimizing radiation injury and neoplastic effects during pediatric fluoroscopy: what should we know?

Authors:  Louis Kenneth Wagner
Journal:  Pediatr Radiol       Date:  2006-09

7.  Pulse fluoroscopy radiation reduction in a pediatric cardiac catheterization laboratory.

Authors:  Stuart H Covi; Wendy Whiteside; Sunkyung Yu; Jeffrey D Zampi
Journal:  Congenit Heart Dis       Date:  2014-06-26       Impact factor: 2.007

8.  Radiation dose reduction in the invasive cardiovascular laboratory: implementing a culture and philosophy of radiation safety.

Authors:  Kenneth A Fetterly; Verghese Mathew; Ryan Lennon; Malcolm R Bell; David R Holmes; Charanjit S Rihal
Journal:  JACC Cardiovasc Interv       Date:  2012-08       Impact factor: 11.195

9.  gamma-H2AX foci as a biomarker for patient X-ray exposure in pediatric cardiac catheterization: are we underestimating radiation risks?

Authors:  Laurence Beels; Klaus Bacher; Daniël De Wolf; Joke Werbrouck; Hubert Thierens
Journal:  Circulation       Date:  2009-10-26       Impact factor: 29.690

Review 10.  Cancer risks following diagnostic and therapeutic radiation exposure in children.

Authors:  Ruth A Kleinerman
Journal:  Pediatr Radiol       Date:  2006-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.