Literature DB >> 22474091

Modularity of the femoral component in total hip arthroplasty.

Anand Srinivasan1, Edward Jung, Brett Russell Levine.   

Abstract

Modular femoral components have been developed to aid in recreating native femoral version, limb length, and offset in total hip arthroplasty. Use of modular implants results in cost savings, as well. Inventory can be reduced while allowing intraoperative flexibility and options. With modular implants, the femoral prosthesis can be built in situ, which is helpful in minimizing incision length and surgical dissection. However, additional modular junctions are associated with increased concern for component failure through taper fretting, fatigue fracture, and local corrosion, which may contribute to elevated serum metal ion levels. The recent trend toward using larger diameter femoral heads may impart higher loads and stress than were seen previously. Although modular components offer a plethora of intraoperative options in primary and revision total hip arthroplasty, the long-term effects of these additional junctions remains unknown.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22474091     DOI: 10.5435/JAAOS-20-04-214

Source DB:  PubMed          Journal:  J Am Acad Orthop Surg        ISSN: 1067-151X            Impact factor:   3.020


  18 in total

1.  Stem-Sleeve Junction Failure of a Modular Femoral Hip System: a Retrieval Analysis.

Authors:  Feras Waly; Fahad H Abduljabbar; Trevor Gascoyne; Thomas R Turgeon; Olga Huk
Journal:  HSS J       Date:  2015-07-09

2.  Dissociation of Modular Total Hip Arthroplasty at the Neck-stem Interface: A Unique but Possible Complication.

Authors:  Angelos Trellopoulos; Stavros Angelis; George Komnos; Grigorios Avramidis; Euaggelos Gikas
Journal:  Cureus       Date:  2019-09-02

3.  Are titanium-on-titanium TiAl6V4 modular necks safe in total hip arthroplasty for non-overweight patients? Results of a prospective series at a minimum follow-up of 7 years.

Authors:  Matthieu Ollivier; Sébastien Parratte; Alexandre Galland; Alexandre Lunebourg; Jean-Noel Argenson
Journal:  Eur J Orthop Surg Traumatol       Date:  2015-07-10

4.  The Irish National Joint Registry: where are we now?

Authors:  B J O'Neill; M Nugent; J P Cashman; S J O'Flanagan; P Keogh; P J Kenny
Journal:  Ir J Med Sci       Date:  2013-06-19       Impact factor: 1.568

5.  Fretting-corrosion in Hip Implant Modular Junctions: New Experimental Set-up and Initial Outcome.

Authors:  D Royhman; M Patel; M J Runa; J J Jacobs; N J Hallab; M A Wimmer; M T Mathew
Journal:  Tribol Int       Date:  2015-11-01       Impact factor: 4.872

Review 6.  Head-neck taper corrosion in hip arthroplasty.

Authors:  S Hussenbocus; D Kosuge; L B Solomon; D W Howie; R H Oskouei
Journal:  Biomed Res Int       Date:  2015-04-14       Impact factor: 3.411

7.  Histopathological characterization of corrosion product associated adverse local tissue reaction in hip implants: a study of 285 cases.

Authors:  Benjamin F Ricciardi; Allina A Nocon; Seth A Jerabek; Gabrielle Wilner; Elianna Kaplowitz; Steven R Goldring; P Edward Purdue; Giorgio Perino
Journal:  BMC Clin Pathol       Date:  2016-02-27

8.  Fracture of the Modular Neck in Total Hip Arthroplasty.

Authors:  A Hernandez; A Gargallo-Margarit; V Barro; I Gallardo-Calero; A Sallent
Journal:  Case Rep Orthop       Date:  2015-07-22

9.  Implant based differences in adverse local tissue reaction in failed total hip arthroplasties: a morphological and immunohistochemical study.

Authors:  Giorgio Perino; Benjamin F Ricciardi; Seth A Jerabek; Guido Martignoni; Gabrielle Wilner; Dan Maass; Steven R Goldring; P Edward Purdue
Journal:  BMC Clin Pathol       Date:  2014-09-05

10.  Spontaneous Dissociation of Anatomic Medullary Locking A Plus (AML A Plus) Femoral Component at the Head-Neck Interface.

Authors:  Ketan Pande; Juzaily Fekry Leong; Ngai Nung Lo
Journal:  J Orthop Case Rep       Date:  2015 Jul-Sep
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.