Scatter rejection in multislit digital mammography.

Magnus Aslund1, Björn Cederström, Mats Lundqvist, Mats Danielsson.   

Abstract

The scatter to primary ratio (SPR) was measured on a scanning multislit full-field digital mammography system for different thickness of breast equivalent material and different tube voltages. Scatter within the detector was measured separately and was found to be the major source of scatter in the assembly. Measured total SPRs below 6% are reported for breast range 3-7 cm. The performance of the multislit assembly is compared to other imaging geometries with different scatter rejection schemes by using the scatter detective quantum efficiency.

Mesh:

Year:  2006        PMID: 16696469     DOI: 10.1118/1.2179122

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  13 in total

1.  Quantification of breast density with spectral mammography based on a scanned multi-slit photon-counting detector: a feasibility study.

Authors:  Huanjun Ding; Sabee Molloi
Journal:  Phys Med Biol       Date:  2012-07-06       Impact factor: 3.609

2.  The effect of scatter and glare on image quality in contrast-enhanced breast imaging using an a-Si/CsI(TI) full-field flat panel detector.

Authors:  Ann-Katherine Carton; Raymond Acciavatti; Johnny Kuo; Andrew D A Maidment
Journal:  Med Phys       Date:  2009-03       Impact factor: 4.071

3.  On image quality metrics and the usefulness of grids in digital mammography.

Authors:  Han Chen; Mats Danielsson; Cheng Xu; Björn Cederström
Journal:  J Med Imaging (Bellingham)       Date:  2015-02-17

4.  A comparative analysis of OTF, NPS, and DQE in energy integrating and photon counting digital x-ray detectors.

Authors:  Raymond J Acciavatti; Andrew D A Maidment
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

5.  Energy weighting improves dose efficiency in clinical practice: implementation on a spectral photon-counting mammography system.

Authors:  Johan Berglund; Henrik Johansson; Mats Lundqvist; Björn Cederström; Erik Fredenberg
Journal:  J Med Imaging (Bellingham)       Date:  2014-08-28

6.  Ideal-observer detectability in photon-counting differential phase-contrast imaging using a linear-systems approach.

Authors:  Erik Fredenberg; Mats Danielsson; J Webster Stayman; Jeffrey H Siewerdsen; Magnus Aslund
Journal:  Med Phys       Date:  2012-09       Impact factor: 4.071

7.  Impact of flat panel-imager veiling glare on scatter-estimation accuracy and image quality of a commercial on-board cone-beam CT imaging system.

Authors:  Dimitrios Lazos; Jeffrey F Williamson
Journal:  Med Phys       Date:  2012-09       Impact factor: 4.071

Review 8.  A review of breast tomosynthesis. Part I. The image acquisition process.

Authors:  Ioannis Sechopoulos
Journal:  Med Phys       Date:  2013-01       Impact factor: 4.071

9.  Microcalcification detectability using a bench-top prototype photon-counting breast CT based on a Si strip detector.

Authors:  Hyo-Min Cho; Huanjun Ding; William C Barber; Jan S Iwanczyk; Sabee Molloi
Journal:  Med Phys       Date:  2015-07       Impact factor: 4.071

10.  Quantitative contrast-enhanced spectral mammography based on photon-counting detectors: A feasibility study.

Authors:  Huanjun Ding; Sabee Molloi
Journal:  Med Phys       Date:  2017-06-28       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.