Metal artefact reduction in CT imaging of hip prostheses—an evaluation of commercial techniques provided by four vendors.

K M Andersson1, P Nowik2, J Persliden1, P Thunberg1,3, E Norrman1.   

Abstract

OBJECTIVE: The aim of this study was to evaluate commercial metal artefact reduction (MAR) techniques in X-ray CT imaging of hip prostheses.
METHODS: Monoenergetic reconstructions of dual-energy CT (DECT) data and several different MAR algorithms, combined with single-energy CT or DECT, were evaluated by imaging a bilateral hip prosthesis phantom. The MAR images were compared with uncorrected images based on CT number accuracy and noise in different regions of interest.
RESULTS: The three MAR algorithms studied implied a general noise reduction (up to 67%, 74% and 77%) and an improvement in CT number accuracy, both in regions close to the prostheses and between the two prostheses. The application of monoenergetic reconstruction, without any MAR algorithm, did not decrease the noise in the regions close to the prostheses to the same extent as did the MAR algorithms and even increased the noise in the region between the prostheses.
CONCLUSION: The MAR algorithms evaluated generally improved CT number accuracy and substantially reduced the noise in the hip prostheses phantom images, both close to the prostheses and between the two prostheses. The study showed that the monoenergetic reconstructions evaluated did not sufficiently reduce the severe metal artefact caused by large orthopaedic implants. ADVANCES IN KNOWLEDGE: This study evaluates several commercially available MAR techniques in CT imaging of large orthopaedic implants.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26110201      PMCID: PMC4651384          DOI: 10.1259/bjr.20140473

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  15 in total

1.  Development, implementation and evaluation of a dedicated metal artefact reduction method for interventional flat-detector CT.

Authors:  D Prell; W A Kalender; Y Kyriakou
Journal:  Br J Radiol       Date:  2010-09-21       Impact factor: 3.039

2.  Clinical evaluation of a newly developed method for avoiding artifacts caused by dental fillings on X-ray CT.

Authors:  Yasuo Nakae; Kiyoshi Sakamoto; Takahiro Minamoto; Toshiko Kamakura; Yuji Ogata; Mitsuhiro Matsumoto; Takeshi Johkou
Journal:  Radiol Phys Technol       Date:  2007-12-14

3.  An adaptive approach to metal artifact reduction in helical computed tomography for radiation therapy treatment planning: experimental and clinical studies.

Authors:  Mehran Yazdi; Mehran Yazdia; Luc Gingras; Luc Beaulieu
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-07-15       Impact factor: 7.038

Review 4.  Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multi-detector CT.

Authors:  Mi-Jung Lee; Sungjun Kim; Sung-Ah Lee; Ho-Taek Song; Yong-Min Huh; Dae-Hong Kim; Seung Hwan Han; Jin-Suck Suh
Journal:  Radiographics       Date:  2007 May-Jun       Impact factor: 5.333

5.  Reduction of dental metallic artefacts in CT: value of a newly developed algorithm for metal artefact reduction (O-MAR).

Authors:  M Kidoh; T Nakaura; S Nakamura; S Tokuyasu; H Osakabe; K Harada; Y Yamashita
Journal:  Clin Radiol       Date:  2013-10-22       Impact factor: 2.350

6.  Metal artifact reduction software used with abdominopelvic dual-energy CT of patients with metal hip prostheses: assessment of image quality and clinical feasibility.

Authors:  Seung Chol Han; Yong Eun Chung; Young Han Lee; Kwan Kyu Park; Myeong Jin Kim; Ki Whang Kim
Journal:  AJR Am J Roentgenol       Date:  2014-10       Impact factor: 3.959

7.  Artifacts in CT: recognition and avoidance.

Authors:  Julia F Barrett; Nicholas Keat
Journal:  Radiographics       Date:  2004 Nov-Dec       Impact factor: 5.333

8.  Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation.

Authors:  Fabian Bamberg; Alexander Dierks; Konstantin Nikolaou; Maximilian F Reiser; Christoph R Becker; Thorsten R C Johnson
Journal:  Eur Radiol       Date:  2011-01-20       Impact factor: 5.315

9.  Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning.

Authors:  Karin M Andersson; Anders Ahnesjö; Christina Vallhagen Dahlgren
Journal:  J Appl Clin Med Phys       Date:  2014-09-08       Impact factor: 2.102

10.  The CT number accuracy of a novel commercial metal artifact reduction algorithm for large orthopedic implants.

Authors:  Guido Hilgers; Tonnis Nuver; André Minken
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

View more
  22 in total

1.  Metal Artifact Reduction in Head CT Performed for Patients with Deep Brain Stimulation Devices: Effectiveness of a Single-Energy Metal Artifact Reduction Algorithm.

Authors:  Y Nagayama; S Tanoue; S Oda; D Sakabe; T Emoto; M Kidoh; H Uetani; A Sasao; T Nakaura; O Ikeda; K Yamada; Y Yamashita
Journal:  AJNR Am J Neuroradiol       Date:  2019-12-26       Impact factor: 3.825

2.  Reduction of artifacts caused by orthopedic hardware in the spine in spectral detector CT examinations using virtual monoenergetic image reconstructions and metal-artifact-reduction algorithms.

Authors:  Nils Große Hokamp; V Neuhaus; N Abdullayev; K Laukamp; S Lennartz; A Mpotsaris; J Borggrefe
Journal:  Skeletal Radiol       Date:  2017-09-21       Impact factor: 2.199

3.  CT metal artifacts in patients with total hip replacements: for artifact reduction monoenergetic reconstructions and post-processing algorithms are both efficient but not similar.

Authors:  Kai Roman Laukamp; Simon Lennartz; Victor-Frederic Neuhaus; Nils Große Hokamp; Robert Rau; Markus Le Blanc; Nuran Abdullayev; Anastasios Mpotsaris; David Maintz; Jan Borggrefe
Journal:  Eur Radiol       Date:  2018-05-03       Impact factor: 5.315

4.  Clinical evaluation of the iterative metal artefact reduction algorithm for post-operative CT examination after maxillofacial surgery.

Authors:  Arsany Hakim; Johannes Slotboom; Olivier Lieger; Fabian Schlittler; Roland Giger; Chantal Michel; Roland Wiest; Franca Wagner
Journal:  Dentomaxillofac Radiol       Date:  2017-03-07       Impact factor: 2.419

5.  Metal artifact reduction using mono-energy images combined with metal artifact reduction software in spectral computed tomography: a study on phantoms.

Authors:  Xiaoli Sun; Qingqing Zhao; Pengtao Sun; Zhipeng Yao; Rengui Wang
Journal:  Quant Imaging Med Surg       Date:  2020-07

6.  Metal implants on abdominal CT: does split-filter dual-energy CT provide additional value over iterative metal artifact reduction?

Authors:  Hildegard M Wichtmann; Kai R Laukamp; Sebastian Manneck; Konrad Appelt; Bram Stieltjes; Daniel T Boll; Matthias R Benz; Markus M Obmann
Journal:  Abdom Radiol (NY)       Date:  2022-09-30

7.  Evaluation of raw-data-based and calculated electron density for contrast media with a dual-energy CT technique.

Authors:  Daisuke Kawahara; Shuichi Ozawa; Kazushi Yokomachi; Toru Higaki; Takehiro Shiinoki; Yoshimi Ohno; Yuji Murakami; Kazuo Awai; Yasushi Nagata
Journal:  Rep Pract Oncol Radiother       Date:  2019-08-20

Review 8.  [Iterative algorithms for artifact reduction in computed tomography].

Authors:  S Skornitzke
Journal:  Radiologe       Date:  2018-03       Impact factor: 0.635

9.  Comparison of virtual non-contrast dual-energy CT and a true non-contrast CT for contouring in radiotherapy of 3D printed lung tumour models in motion: a phantom study.

Authors:  Dominik Alexander Hering; Kai Kröger; Ralf W Bauer; Hans Theodor Eich; Uwe Haverkamp
Journal:  Br J Radiol       Date:  2020-10-01       Impact factor: 3.039

10.  Evaluation of Orthopedic Metal Artifact Reduction Application in Three-Dimensional Computed Tomography Reconstruction of Spinal Instrumentation: A Single Saudi Center Experience.

Authors:  Amir Monir Ali
Journal:  J Clin Imaging Sci       Date:  2018-03-12
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.