Literature DB >> 20821172

Clinical evaluation of a newly developed method for avoiding artifacts caused by dental fillings on X-ray CT.

Yasuo Nakae1, Kiyoshi Sakamoto, Takahiro Minamoto, Toshiko Kamakura, Yuji Ogata, Mitsuhiro Matsumoto, Takeshi Johkou.   

Abstract

To evaluate the clinical usefulness of gantry tilt scanning as an image reconstruction technique for avoiding artifacts caused by metallic dental fillings. Gantry tilt scanning was used with multidetector-row computed tomography for imaging in patients with dental fillings. Using a novel PC-based program, the oblique images obtained were reconstructed to transverse images using nearest neighbor and bilinear interpolation methodologies in order to avoid metallic streak artifacts. Coronal images were reformatted with the reconstructed transverse images, and the continuity of the reconstructed images was evaluated. Gantry tilt scanning was performed in 12 patients with metal artifacts, and the original and reconstructed images were classified into four grades and assessed by two radiologists. Results of the clinical evaluation indicated that the original images with artifacts, only 4% had good image quality in the region around the medial pterygoid muscle, only 8% depicted areas around the internal carotid artery and internal jugular vein, and only 12% could depict the areas around the parotid gland in the clinical evaluation. These values were improved to 60, 96, and 100%, respectively, in the reconstructed transverse images. Gantry tilt scanning as an image reconstruction technique improves image quality and removes most, if not all, artifacts caused by metallic dental fillings. The resulting images can be used in the evaluation of oropharyngeal lesions in patients with dental fillings.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 20821172     DOI: 10.1007/s12194-007-0016-8

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  15 in total

1.  Fast iterative algorithm for metal artifact reduction in X-ray CT.

Authors:  G Wang; T Frei; M W Vannier
Journal:  Acad Radiol       Date:  2000-08       Impact factor: 3.173

2.  [Evaluation of gantry tilt correction in multi detector-row CT: effect on multi-planar reconstruction images].

Authors:  Toshiya Nasada; Shinichi Hirayama; Masao Tanooka; Tomoko Fujita; Toshiyuki Sakai
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2003-01

Review 3.  An extensible software-based platform for reconstruction and evaluation of CT images.

Authors:  Dirk-Alexander Sennst; Marc Kachelriess; Christianne Leidecker; Bernhard Schmidt; Oliver Watzke; Willi A Kalender
Journal:  Radiographics       Date:  2004 Mar-Apr       Impact factor: 5.333

4.  The use of CT-MR image registration to define target volumes in pelvic radiotherapy in the presence of bilateral hip replacements.

Authors:  N Charnley; A Morgan; E Thomas; S Wilson; S Bacon; D Wilson; D Bottomley
Journal:  Br J Radiol       Date:  2005-07       Impact factor: 3.039

5.  [Improvement of metal artifacts in dental structures by X-ray CT: reconstruction of transverse images using oblique images by gantry tilt scanning].

Authors:  Yasuo Nakae; Kiyoshi Sakamoto; Takahiro Minamoto; Noriko Kotoura; Takao Ozaki; Takeshi Johkoh
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2006-06-20

6.  New edge-directed interpolation.

Authors:  X Li; M T Orchard
Journal:  IEEE Trans Image Process       Date:  2001       Impact factor: 10.856

7.  Reduction of CT artifacts caused by metallic implants.

Authors:  W A Kalender; R Hebel; J Ebersberger
Journal:  Radiology       Date:  1987-08       Impact factor: 11.105

8.  An algorithm for the reduction of metal clip artifacts in CT reconstructions.

Authors:  G H Glover; N J Pelc
Journal:  Med Phys       Date:  1981 Nov-Dec       Impact factor: 4.071

9.  [influence and improvement of metal artifacts in dental structures by CT for radiation treatment planning: reconstruction of transverse images using oblique images by gantry tilt scanning].

Authors:  Yasuo Nakae; Hiroyuki Inoue; Takahiro Minamoto; Toshijiro Yamamoto; Takeshi Johkou
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2007-03-20

10.  Artifacts in CT: recognition and avoidance.

Authors:  Julia F Barrett; Nicholas Keat
Journal:  Radiographics       Date:  2004 Nov-Dec       Impact factor: 5.333

View more
  5 in total

1.  Metal artefact reduction in CT imaging of hip prostheses—an evaluation of commercial techniques provided by four vendors.

Authors:  K M Andersson; P Nowik; J Persliden; P Thunberg; E Norrman
Journal:  Br J Radiol       Date:  2015-05-27       Impact factor: 3.039

2.  Reduction of metal artifacts due to dental hardware in computed tomography angiography: assessment of the utility of model-based iterative reconstruction.

Authors:  Keita Kuya; Yuki Shinohara; Ayumi Kato; Makoto Sakamoto; Masamichi Kurosaki; Toshihide Ogawa
Journal:  Neuroradiology       Date:  2017-03-02       Impact factor: 2.804

3.  Clinical evaluation of the normalized metal artefact reduction algorithm caused by dental fillings in CT.

Authors:  X-Y Gong; E Meyer; X-J Yu; J-H Sun; L-P Sheng; K-H Huang; R-Z Wu
Journal:  Dentomaxillofac Radiol       Date:  2013-02-18       Impact factor: 2.419

4.  Metal Artifact Suppression in Dental Cone Beam Computed Tomography Images Using Image Processing Techniques.

Authors:  Masoumeh Johari; Milad Abdollahzadeh; Farzad Esmaeili; Vahideh Sakhamanesh
Journal:  J Med Signals Sens       Date:  2018 Jan-Mar

5.  Evaluation of Orthopedic Metal Artifact Reduction Application in Three-Dimensional Computed Tomography Reconstruction of Spinal Instrumentation: A Single Saudi Center Experience.

Authors:  Amir Monir Ali
Journal:  J Clin Imaging Sci       Date:  2018-03-12