Literature DB >> 26069448

Feasibility of a novel atrioventricular delay optimization method using transmitral and pulmonary venous flow in patients with sequential ventricular pacing or cardiac resynchronization therapy.

Kenzo Fukuhara1, Hiroyuki Okura1, Terumasa Koyama1, Teruyoshi Kume1, Yoji Neishi1, Akihiro Hayashida1, Kiyoshi Yoshida1.   

Abstract

BACKGROUND: Although several echo-Doppler methods were proposed to optimize atrioventricular (AV) delay in patients with sequential ventricular pacing, "echo-guided" AV optimization has not been widely adopted clinically. A combination of trasmitral flow (TMF) and pulmonary venous flow (PVF) measurements may be beneficial to further optimize AV delay to achieve better cardiac function. The aim of this study was to assess the feasibility and usefulness of AV delay optimization by combined use of TMF and PVF.
METHODS: A total of 32 patients after sequential ventricular pacemaker implantation were enrolled and studied. The optimal AV delay was defined as the timing to minimize the duration between PVF reversal (a) wave and the duration of the "A" wave of TMF. Stroke volume was measured at the "optimized" AV delay (AVD(OPT)) and was compared with that obtained at shorter (AVD(OPT) - 50 ms) and longer (AVD(OPT) + 50 ms) AV delays.
RESULTS: AV optimization was feasible in 27 of 32 patients (87%). Stroke volume at AVD(OPT) was significantly higher than that at shorter or longer AV delay (63 ± 18 ml vs. 57 ± 15 ml vs. 56 ± 16 ml, P = 0.001).
CONCLUSIONS: AV delay optimization using TMF and PV flow was feasible. Usefulness of this method requires further investigation with a larger study population.

Entities:  

Keywords:  AV delay; Cardiac resynchronization therapy; Doppler echocardiography; Left ventricular function; Pacemaker

Mesh:

Year:  2014        PMID: 26069448      PMCID: PMC4454827          DOI: 10.1007/s12574-014-0237-x

Source DB:  PubMed          Journal:  J Echocardiogr        ISSN: 1349-0222


Introduction

Left ventricular (LV) dysfunction may develop as a result of LV dyssynchrony and/or inappropriate atrioventricular (AV) delay in some patients after single chamber, ventricular pacing. Even after dual-chamber sequential pacing, maintenance of AV synchrony is necessary to preserve cardiac function and to achieve a better prognosis [1, 2]. AV delay optimization is, therefore, important to maintain better cardiac function and a favorable long-term outcome after sequential pacing [3, 4] or cardiac resynchronization therapy [5, 6]. Although several echo-Doppler- [7-13] as well as electrocardiogram- [14-17] based methods to optimize AV interval have been proposed, routine or systematic use of AV optimization remains controversial [5, 6, 18–20]. Transmitral flow (TMF) by transthoracic Doppler echocardiography is commonly used to optimize AV delay. However, the advantage of echo-Doppler-based AV optimization over fixed AV delay or a commercially available AV optimization algorithm based on electrocardiogram has not been proven yet. Theoretically (based on the Frank–Starling law), AV delay should be optimized to achieve maximal LV filling without deterioration of LV function [2]. Because TMF alone does not reflect both systolic function and LV filling pressure, TMF-based AV optimization may not provide enough advantage over the other methods. A previous echo-Doppler study demonstrated that the difference between the duration of pulmonary venous flow reversal (PVa) and mitral forward flow during atrial systole (A) reliably estimates LV filling pressure [21]. We hypothesized that a combination of TMF and PV flow measurements may be beneficial to further optimize AV delay to achieve better cardiac function with adequate LV filling pressure. Therefore, the aim of this study was to assess feasibility of the AV delay optimization by combined use of TMF and PV flow.

Materials and methods

Study population

This study included 32 patients after dual-chamber pacing for complete AV block (n = 26, mean age = 79 ± 8 years; 12 males) or cardiac resynchronization therapy (n = 6, mean age = 65 ± 16 years; 4 males). The exclusion criteria were current atrial arrhythmia and frequent premature ventricular beats. Informed consent was provided by each participant before enrollment in this study.

Study protocol

Echocardiography was performed with a Sonos 5500 and S3 transducer (Philips Medical Systems, Andover, MA, USA). TMF was obtained from apical 3-chamber or 4-chamber views with the sample volume positioned at the tip of the mitral leaflets. TMF consists of 2 distinct flow signals, early or E wave and late or A wave during atrial contraction. PV flow was obtained from an apical 4-chamber view with the sample volume placed in the left superior pulmonary vein. An effort was made to maintain the same position of the pulsed Doppler sample throughout the echo-Doppler examination. AV delay optimization was performed using TMF and PV flow at rest. Optimal AV delay (AVDOPT) was defined as the AV delay where the duration of PVa minus A was the minimum (=0). Because the onset of the A wave cannot be always detected, the difference between the duration of PVa and A was alternatively measured as (time interval between the onset of the Q wave and the end of the A wave) − (time interval between the onset of the Q wave and the end of the PVa wave). To simplify this method, TMF and PV flow were recorded at the pre-set AV delay. Then, AVDOPT was determined as (pre-set AV delay) + (duration of PVa − duration of A) (Fig. 1). Stroke volume (SV) was measured by a pulsed Doppler method obtained at the LV outflow tract and was used as an index to assess cardiac function during AV optimization. SV obtained at the AVDOPT was compared with SV obtained at shorter (AVDOPT − 50 ms) or longer (AVDOPT + 50 ms) AV delays.
Fig. 1

AV delay optimization using TMF and PV flow. a (Step 1) At a pre-set AV delay (=180 ms in this case), both TMF and PV flow signal were recorded. (Step 2) The difference in duration between PVa and A wave was measured (=−30 ms). b (Step 3) Optimal AV delay was calculated as (pre-set AV delay) + (duration of PVa − duration of A). In this case, the optimal AV delay was calculated as 180 ms + (−30 ms) = 150 ms

AV delay optimization using TMF and PV flow. a (Step 1) At a pre-set AV delay (=180 ms in this case), both TMF and PV flow signal were recorded. (Step 2) The difference in duration between PVa and A wave was measured (=−30 ms). b (Step 3) Optimal AV delay was calculated as (pre-set AV delay) + (duration of PVa − duration of A). In this case, the optimal AV delay was calculated as 180 ms + (−30 ms) = 150 ms

Statistical analysis

The measurements are expressed as mean ± standard deviation. Statistical analyses were performed with one-way analysis of variance (ANOVA) using the Bonferroni post hoc test. Values of P < 0.05 were considered statistically significant.

Results

Clinical characteristics of the 32 patients are shown in Table 1. Dual chamber (DDD) pacing was used in 12 patients and ventricular (VDD) pacing in 20 patients. All patients after dual-chamber sequential pacing were in New York Heart Association (NYHA) class I or П. On the other hand, all patients after cardiac resynchronization therapy (CRT) were in NYHA class III. Left ventricular ejection fraction (LVEF) was 55 ± 15 %.
Table 1

Clinical characteristics

(n = 32)
Age (years)76 ± 11
Male gender, n (%)16 (50)
DDD/VDD12/20
Pacemaker/CRT6/26
Ischemic heart disease, n (%)9 (28)
Diabetes, n (%)12 (38)
Hypertension, n (%)20 (62)
Dyslipidemia, n (%)14 (44)
NYHA class (I/II/III/IV)24/2/6/0
Medication, n (%)
 β-blockers10 (31)
 ACE-I/ARB18 (56)
 Loop diuretics11 (34)
 Spironolactone6 (19)
 Digitalis1 (3)
 Statins12 (38)

ACE-I angiotensin-converting enzyme inhibitors, ARB angiotensin receptor blockers, CRT cardiac resynchronization therapy

Clinical characteristics ACE-I angiotensin-converting enzyme inhibitors, ARB angiotensin receptor blockers, CRT cardiac resynchronization therapy AV optimization using our current method could be performed in 27 of 32 patients (84 %). In the remaining 5 patients, adequate PV flow signal could not be recorded. The measurements made in all patients are summarized in Tables 2 and 3. Mean AVDOPT was 143 ± 35 ms. As expected, the mean AVDOPT was significantly lower in the VDD than in the DDD mode (133 ± 32 ms vs. 170 ± 37 ms, P = 0.014). SV at AVDOPT was significantly higher than shorter or longer AV delay (63 ± 18 ml vs. 57 ± 15 ml vs. 56 ± 16 ml, P = 0.001) (Fig. 2).
Table 2

Hemodynamic and Doppler echocardiography parameters

Patient no.GenderAgeDiseaseLVEF (%)LVDd (mm)LVDs (mm)Ao TVI (cm)SV (ml)
1M75Complete AV block6048302253
2F46Complete AV block6048302261
3F81Complete AV block4846371243
4M73Complete AV block5842262386
5F84Complete AV block5441322149
6M75Complete AV block5951312679
7F88Complete AV block6243233061
8F77Complete AV block5843292044
9M73Complete AV block54443234108
10M78Complete AV block5842281754
11M80Complete AV block6542242183
12M81Complete AV block5636211531
13F82Complete AV block6342262655
14M78Complete AV block5554361952
15M83Complete AV block7145205574
16M81Complete AV block7143302681
17F82Complete AV block6750332056
18F90Complete AV block7532171835
19F79Complete AV block6737191226
20F72Complete AV block6747292989
21F82Complete AV block6148282668
22M74Complete AV block5454331965
23M38DCM3670581561
24M75Ischemic heart disease3261502178
25F84Ischemic heart disease7634163344
26M64DCM4557392060
27M71DCM3163602056

LVEF left ventricular ejection fraction; LVDd left ventricular end-diastolic diameter; LVDs left ventricular end-systolic diameter; Ao TVI aorta time velocity integral; SV stroke volume; DCM dilated cardiomyopathy

Table 3

Pacing mode, pacing rate at initial enrollment and TMF A, PVa duration pre and post AV delay optimization

Patient no.Pacing modeHRPre AV delay optimizationPost AV delay optimization
TMF A durationPVa durationTMF A durationPVa duration
1DDD (A sense V pace)70115120130130
2VDD (A sense V pace)72145140145140
3DDD (A sense V pace)60165155165155
4DDD (A sense V pace)60160160160160
5VDD (A sense V pace)70150125150125
6VDD (A sense V pace)60120130120130
7VDD (A sense V pace)75170100122122
8VDD (A sense V pace)70115130126122
9VDD (A sense V pace)55145145145145
10VDD (A sense V pace)60140150155145
11DDD (A sense V pace)60165165165165
12VDD (A sense V pace)62135160135160
13VDD (A sense V pace)70135110150135
14DDD (A sense V pace)69150145115125
15VDD (A sense V pace)60135115135115
16VDD (A sense V pace)60175180185190
17VDD (A sense V pace)60145105135130
18VDD (A sense V pace)75180140130130
19VDD (A sense V pace)70130125115115
20VDD (A sense V pace)60140130140130
21VDD (A sense V pace)76110125110125
22DDD (A pace V pace)80150115150115
23DDD (A sense V pace)758775
24VDD (A sense V pace)65
25DDD (A pace V pace)60127
26DDD (A pace V pace)60
27VDD (A sense V pace)65150145150145
28DDD (A sense V pace)70140170140170
29VDD (A sense V pace)96130125130125
30VDD (A sense V pace)80150170130145
31DDD (A pace V pace)60115155115155
32VDD (A sense V pace)60

TMF transmitral flow

Fig. 2

Comparison of SV in all patients. SV obtained at the AVDOPT was compared with SV obtained at shorter (AVDOPT − 50 ms) or longer (AVDOPT + 50 ms) AV delays

Hemodynamic and Doppler echocardiography parameters LVEF left ventricular ejection fraction; LVDd left ventricular end-diastolic diameter; LVDs left ventricular end-systolic diameter; Ao TVI aorta time velocity integral; SV stroke volume; DCM dilated cardiomyopathy Pacing mode, pacing rate at initial enrollment and TMF A, PVa duration pre and post AV delay optimization TMF transmitral flow Comparison of SV in all patients. SV obtained at the AVDOPT was compared with SV obtained at shorter (AVDOPT − 50 ms) or longer (AVDOPT + 50 ms) AV delays In a subset of patients after sequential dual-chamber pacing for complete AV block, AV optimization could be performed in 22 of 26 patients (85 %). The AVDOPT in VDD mode was 128 ± 38 ms and the AVDOPT in the DDD pacing mode was 177 ± 39 ms. SV with AVDOPT was significantly higher than shorter or longer AV delay (64 ± 19 ml vs. 57 ± 16 ml vs. 56 ± 17 ml, P = 0.0001) (Fig. 3).
Fig. 3

Comparison of SV in patients after sequential, dual chamber pacing for complete AV block. SV obtained at the AVDOPT was compared with SV obtained at shorter (AVDOPT − 50 ms) or longer (AVDOPT + 50 ms) AV delays

Comparison of SV in patients after sequential, dual chamber pacing for complete AV block. SV obtained at the AVDOPT was compared with SV obtained at shorter (AVDOPT − 50 ms) or longer (AVDOPT + 50 ms) AV delays Similarly, in a subset of patients after CRT, AV optimization could be performed in 5 of 6 patients (83 %). The AVDOPT in VDD mode was 128 ± 38 ms and the AVDOPT in the DDD pacing mode was 177 ± 39 ms. SV with AVDOPT was significantly higher than shorter or longer AV delay (61 ± 13 ml vs. 53 ± 11 ml vs. 57 ± 10 ml, P = 0.026) (Fig. 4).
Fig. 4

Comparison of SV in CRT recipients. SV obtained at the AVDOPT was compared with SV obtained at shorter (AVDOPT − 50 ms) or longer (AVDOPT + 50 ms) AV delays

Comparison of SV in CRT recipients. SV obtained at the AVDOPT was compared with SV obtained at shorter (AVDOPT − 50 ms) or longer (AVDOPT + 50 ms) AV delays Reproducibility of PV flow measurements was analyzed. Correlation coefficients were high for repeated measurements by the same observer (r = 0.98 for duration of PVa minus A) and measurements by 2 different observers (r = 0.88 for duration of PVa minus A).

Discussion

This study shows that AV delay optimization based on a new echo-Doppler method using TMF and PV flow is feasible. In addition, increased SV during AV delay optimized by this method may suggest a potential favorable impact on cardiac function and possibly prognosis. A previous randomized, prospective study comparing echo-guided AV delay optimization and an empiric, fixed AV delay of 120 ms demonstrated improved clinical outcome at 3 months in patients with echo-guided AV optimization [19]. In their study, optimal AV delay was defined as the largest aortic velocity–time integral at one of eight tested AV intervals (between 60 and 200 ms). On the other hand, a more recent large-scale randomized prospective multicenter trial (SMART-AV trial) to compare between a fixed empirical AV delay (120 ms), echocardiographically optimized AV delay, and AV delay optimized with SmartDelay electrocardiogram-based algorithm did not show superiority of echocardiography or SmartDelay over a fixed AV delay of 120 ms [18]. In their study, Ritter’s method [10, 22] and/or an iterative method [23] using TMF were used to optimize AV delay as endorsed by the American Society of Echocardiography [23, 24]. Based on their negative results, the authors stated that routine echocardiographic AV optimization using the American Society of Echocardiography recommended method for patients with CRT should be abandoned [18]. However, it is not certain whether all echo-Doppler methods should be abolished. Ritter et al. [22] first reported an echo-Doppler method to optimize AV delay in patients with complete AV block and a normal LV systolic function. Ritter et al. defined the AV delay with the echo method that provided the longest diastolic filling time without interruption of the A wave. Ritter’s formula, which can be regarded as the current “gold standard” in AV delay optimization [24] requires 2 measurements: (1) QA short = the time interval between the onset of the Q wave and the end of the truncated “A” wave of the TMF at a short (30–60 ms) AV delay; and (2) QA long = the time interval between the onset of the Q wave and the end of the “A” wave of the TMF at a long (200 ms) AV interval. According to the formula, optimal AV delay was calculated as AV long − (QA short − QA long). This method has been used in several clinical trials because it is a simple, non-invasive and reproducible method [20]. On the other hand, Ishikawa et al. used diastolic mitral regurgitation to optimize AV delay. As compared with Ritter’s method in which AV delay was optimized to achieve the highest cardiac output, Ishikawa’s method is to achieve the lowest possible left atrial or LV filling pressure [9, 25]. In our present study, we used both TMF and PV flow to achieve the lowest LV filling pressure and the highest SV. The concept of Doppler assessment of LV filling pressure using both TMF and PV flow was first reported in 1993 by Rossvoll and Hatle [21]. The difference in duration between PVa of the PV flow and antegrade A wave by the TMF was positively and strongly correlated with LV end-diastolic pressure (r = 0.68, P < 0.001). A longer duration of PVa versus A wave predicted increased (>15 mmHg) LV end-diastolic pressure [21]. The mechanisms for a longer duration of PVa than the A wave was explained by the increased LV end-diastolic pressure as a result of reduced LV compliance. Therefore, an AV delay that does not prolong PVa more than the A wave could be considered as a hemodynamically optimal AV delay. Although our preliminary data suggest that AV optimization based on the TMF and PV flow is feasible, it was not possible for AV optimization to be performed in some patients in whom PV flow could not be detected. This is a possible limitation of this study. Detection of the PV flow signal using the transthoracic approach depends upon the image quality of the echo-Doppler machine. Although the sensitivity of the Doppler measurements for some specific conditions was not sufficient when using old echo-Doppler machines and therefore required contrast enhancement [26, 27], modern echo-Doppler machines have sufficiently sensitive Doppler equipment [28, 29]. Another apparent limitation is that 2 different Doppler measurements are required for our method which appears to be time consuming. However, as compared with Ritter’s method, which requires 2 TMF recordings at 2 different AV delay settings, our method is less time consuming. Because this is a small pilot study, further investigations will be necessary. First, the hemodynamically favorable acute results should be confirmed by invasive hemodynamic monitoring. Second, the long-term clinical impact of the acute results should be investigated by a serial observation of the study population. Finally, the advantages of the current method should be investigated by comparing it with other echo-Doppler methods or empirical fixed AV delay prospectively.

Conclusions

A novel AV delay optimization method using TMF and PV flow has been shown to be feasible. The usefulness of this method requires further investigation with a larger study population.
  28 in total

1.  Optimal atrioventricular delay setting determined by evoked QT interval in patients with implanted stimulus-T-driven DDDR pacemakers.

Authors:  T Ishikawa; T Sugano; S Sumita; T Nakagawa; K Hanada; M Kosuge; I Kobayashi; K Kimura; O Tochikubo; T Usui; S Umemura
Journal:  Europace       Date:  2001-01       Impact factor: 5.214

Review 2.  Optimization of the atrioventricular delay in sequential and biventricular pacing: physiological bases, critical review, and new purposes.

Authors:  Lanfranco Antonini; Antonio Auriti; Vincenzo Pasceri; Antonella Meo; Christian Pristipino; Antonio Varveri; Salvatore Greco; Massimo Santini
Journal:  Europace       Date:  2012-02-06       Impact factor: 5.214

3.  Primary results from the SmartDelay determined AV optimization: a comparison to other AV delay methods used in cardiac resynchronization therapy (SMART-AV) trial: a randomized trial comparing empirical, echocardiography-guided, and algorithmic atrioventricular delay programming in cardiac resynchronization therapy.

Authors:  Kenneth A Ellenbogen; Michael R Gold; Timothy E Meyer; Ignacio Fernndez Lozano; Suneet Mittal; Alan D Waggoner; Bernd Lemke; Jagmeet P Singh; Francis G Spinale; Jennifer E Van Eyk; Jeffrey Whitehill; Stanislav Weiner; Maninder Bedi; Joshua Rapkin; Kenneth M Stein
Journal:  Circulation       Date:  2010-11-15       Impact factor: 29.690

4.  Prevalence and correlates of physiological valvular regurgitation in healthy subjects.

Authors:  Hiroyuki Okura; Yuko Takada; Azusa Yamabe; Takeshi Ozaki; Hiroyuki Yamagishi; Iku Toda; Minoru Yoshiyama; Junichi Yoshikawa; Kiyoshi Yoshida
Journal:  Circ J       Date:  2011-09-01       Impact factor: 2.993

5.  Pulmonary venous flow velocities recorded by transthoracic Doppler ultrasound: relation to left ventricular diastolic pressures.

Authors:  O Rossvoll; L K Hatle
Journal:  J Am Coll Cardiol       Date:  1993-06       Impact factor: 24.094

6.  Determination of the optimal atrioventricular delay in DDD pacing. Comparison between echo and peak endocardial acceleration measurements.

Authors:  P Ritter; L Padeletti; L Gillio-Meina; G Gaggini
Journal:  Europace       Date:  1999-04       Impact factor: 5.214

7.  Doppler index and plasma level of atrial natriuretic hormone are improved by optimizing atrioventricular delay in atrioventricular block patients with implanted DDD pacemakers.

Authors:  N Toda; T Ishikawa; N Nozawa; I Kobayashi; H Ochiai; K Miyamoto; S Sumita; K Kimura; S Umemura
Journal:  Pacing Clin Electrophysiol       Date:  2001-11       Impact factor: 1.976

8.  Adjusting the timing of left-ventricular pacing using electrocardiogram and device electrograms.

Authors:  Yaariv Khaykin; Derek Exner; David Birnie; John Sapp; Sandeep Aggarwal; Aleksandre Sambelashvili
Journal:  Europace       Date:  2011-05-19       Impact factor: 5.214

9.  A fast and simple echocardiographic method of determination of the optimal atrioventricular delay in patients after biventricular stimulation.

Authors:  JaroslaV Meluzín; Miroslav Novák; Jolana Müllerová; Jan Krejcí; Petr Hude; Martin Eisenberger; Ladislav Dusek; Ivo Dvorák; Lenka Spinarová
Journal:  Pacing Clin Electrophysiol       Date:  2004-01       Impact factor: 1.976

Review 10.  Doppler echocardiographic assessment of the effect of varying atrioventricular delay and pacemaker mode on left ventricular filling.

Authors:  A C Pearson; D L Janosik; R R Redd; T A Buckingham; R I Blum; A J Labovitz
Journal:  Am Heart J       Date:  1988-03       Impact factor: 4.749

View more
  2 in total

1.  Echocardiographic evaluation of cardiac dyssynchrony in patients with congestive heart failure.

Authors:  Chuan Qin; Li Zhang; Zi-Ming Zhang; Bin Wang; Zhou Ye; Yong Wang; Navin C Nanda; Ming-Xing Xie
Journal:  J Huazhong Univ Sci Technolog Med Sci       Date:  2016-07-05

2.  AV timing in pacemaker patients with first-degree AV block: which is preferable, intrinsic AV conduction or pacing?

Authors:  Yoshihiro Aizawa; Toshiko Nakai; Yukitoshi Ikeya; Rikitake Kogawa; Yuki Saito; Kazuto Toyama; Tetsuro Yumikura; Naoto Otsuka; Koichi Nagashima; Yasuo Okumura
Journal:  Heart Vessels       Date:  2022-02-08       Impact factor: 1.814

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.