Kaweh Mansouri1, Robert N Weinreb2, John H K Liu2.
Abstract
PURPOSE: To study performance of a contact lens sensor (CLS) for 24-hour monitoring of IOP-related short-term patterns and compare with IOP obtained by pneumatonometry.Entities:
Mesh:
Year: 2015 PMID: 25942434 PMCID: PMC4420265 DOI: 10.1371/journal.pone.0125530
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1CONSORT flow chart.
Fig 2Twenty-four-hour IOP pattern measured in CLS (1B) and pneumatonometer eye (1A) (N = 30).
Coefficient of determination between the two eyes was R2 = 0.956. Pneumatonometer data were collected sitting during the diurnal/wake period and supine during the nocturnal/sleep period.
Accuracy of ocular pulse frequency by contact lens sensor in relation to heart rate interval.
| OPF interval | Proportion of accurate cases | Lower 95.25% CI | Upper 95.25% CI |
|---|---|---|---|
| T0(per protocol) | 86.5%(45 out of 52 plots) | 75.0% | 93.2% |
| T1 | 85.5%(49 out of 56 plots) | 64.1% | 95.1% |
| T2 | 86.0%(58 out of 68 plots) | 74.5% | 92.9% |
Abbreviations: CI = confidence interval; OPF = ocular pulse frequency
Fig 4Example of 24-h CLS pattern recorded in a study subject.
The red line represents the wake/sleep (W/S) slope as determined from 1 hour pre- to 1 hour post-sleep.
Fig 5Calculation of wake/sleep (W/S) slope as determined from 1 hour pre- to 1 hour post-sleep with CLS (blue).
For the pneumatonometer means (red), actual relative time is not used, but rather one point before, one point closely following and one point after “lights off time”, with a two hour difference between each.
Baseline Demographic and Ocular Characteristics of Study Participants (n = 33).
| Age, yrs (mean ± SD) | 35.0 ± 14.4 |
|---|---|
| Gender | |
| Male | 17 (51.5%) |
| Female | 16 (48.5%) |
| Ancestry | |
| Caucasian | 26 (78.8%) |
| Asian | 6 (18.2%) |
| African-American | 1 (3.0%) |
| SER, D (mean ± SD) | |
| CLS eye | -0.7 ± 1.3 |
| Pneumatonometer eye | -0.8 ± 1.2 |
| CCT, μm (mean ± SD) | |
| CLS eye | 560.3 ± 42.4 |
| Pneumatonometer eye | 560.6 ± 41.1 |
| IOP, mmHg (mean ± SD) | |
| CLS eye | 17.8 ± 2.8 |
| Pneumatonometer eye | 17.8 ± 2.9 |
| Body Mass Index (kg/m2), (mean ± SD) | 25.0 ± 4.1 |
Abbreviations: CCT = Central corneal thickness; CLS = Contact lens sensor; SER = Spherical equivalent refraction