Satya Pal Sharma1, Anders Bærheim2, Alice Kvåle3.
Abstract
BACKGROUND: Measuring range of motion (ROM) in the shoulder joint is important for the diagnosis and monitoring of change over time. To what degree passive ROM can be trusted as a reliable outcome measure was examined as part of an on-going randomized controlled trial for patients with shoulder capsulitis. The aim of this study was to examine intertester reliability of passive ROM in the shoulder joint over a period of eight weeks in patients with adhesive capsulitis stage II.Entities:
Mesh:
Year: 2015 PMID: 25888419 PMCID: PMC4340115 DOI: 10.1186/s12891-015-0495-4
Source DB: PubMed Journal: BMC Musculoskelet Disord ISSN: 1471-2474 Impact factor: 2.362
Summary of intra- and intertester reliability studies for range of motion on both the affected and the non-affected shoulders using different measuring modalities
|
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|---|
| Pandya et al. [ | Muscular dystrophy N = 150 | PROM: ABD | Affected | Goniometer | Point estimates | Inter-tester & intra-tester | 0.67 | No |
| Riddle et al.[ | Shoulder pain N = 100, two groups of 50 each | PROM: ABD | Affected | Goniometer | Point estimates | Inter-tester & intra-tester | 0.87 | No |
| PROM: ER | 0.88 | |||||||
| PROM: IR | 0.55 | |||||||
| Croft et al. [ | Shoulder complaints N = 6 | PROM: EL | Affected | Protractor | Point estimates | Inter-tester | 0.95 | No |
| PROM: ER | Affected | 0.43 | ||||||
| Green et al.[ | Shoulder pain & stiffness N = 6 | AROM: FLEX, ABD, ER, IR, HBB | Affected | Plurimeter | Point estimates | Inter-tester & intra-tester | 0.62–0.95 | No |
| Sabari et al. [ | Rehab.patients N = 11 | AROM: FLEX, ABD | Mixed | Goniometer | Point estimates | Intra-tester | 0.73–0.78 | No |
| Normal N = 19 | PROM: FLEX, ABD | 0.78–0.81 | ||||||
| MacDermid et al. [ | Shoulder pathology | PROM: ER | Affected | Goniometer | Point estimates | Inter-tester & intra-tester | 0.85 & 0.80 | Yes |
| N = 34 | ||||||||
| Hayes et al. [ | Shoulder pathology | AROM: ABD | Affected | Goniometer, visual est., photography | Point estimates | Inter-tester & intra-tester | ICC(Rho) 0.69 | Yes |
| N = 8 | ||||||||
| AROM:ER | Affected | Goniometer | 0.64 | |||||
| AROM: HBB | Measuring tape | 0.39 | ||||||
| Hoving et al. [ | Shoulder pain & stiffness N = 6 | AROM: ABD | Affected | Plurimeter | Point estimates | Inter-tester & intra-tester | 0.51 | No |
| ER in neutral | 0.29 | |||||||
| ER in abd | 0.11 | |||||||
| IR in abd | 0.06 | |||||||
| HBB | 0.80 | |||||||
| Awan et al. [ | Normal (athletes) N = 56 | ER without scapular stabilization | Unaffected, right & left side | Digital inclinometer | Point estimates | Inter-tester/ intra-tester | 0.41–0.51 | No |
| IR standard technique | 0.50–0.66 | |||||||
| IR with scap.stab. | ||||||||
| IR without scap. stab | ||||||||
| de Winter et al. [ | Shoulder complaints N = 155 | PROM: ABD | Affected/ non-affected | Digital inclinometer | Point estimates | Inter-tester | 0.83/0.28 | No |
| PROM: ER | 0.90/0.56 | |||||||
| Terwee et al. [ | Shoulder complaints N = 201 | PROM: ABD | Affected/ non-affected | Visual estimation | Point estimates | Inter-tester | 0.67/0.15 | Yes |
| AROM: EL | 0.88/0.76 | |||||||
| PROM: ER | 0.73/0.34 | |||||||
| Nadeau et al. [ | Normal N = 30 | AROM: ELEV, PROTR, RETRAC | Non-affected | Goniometer & tape measure | Point estimates | Inter-tester & intra-tester | 0.78–0.46 | No |
| 0.75–0.32 | ||||||||
| Tveitå et al. [ | Shoulder adhesive capsulitis N = 32 | PROM: ABD | Affected/non-affected | Digital inclinometer | Point estimates | Intra-tester | 0.72/0.89 | Yes |
| FLEX | 0.76/0.61 | |||||||
| IR | 0.81/0.88 | |||||||
| ER | 0.91/0.86 | |||||||
| Mullaney et al. [ | Shoulder pain N = 20 | AROM: FLEX | Affected/non-affected | Goniometer/ digital level | Point estimates | Inter-tester & intra-tester | 0.93/0.74, 0.92/0.79, 0.82/0.62 | Yes |
| ER | ||||||||
| IR | ||||||||
| Kolber et al. [ | Normal N = 30 | AROM: ABD | Non-affected | Inclinometer | Point estimates | Inter-tester & intra-tester | 0.95 | Yes |
| AROM: ER | Non-affected | 0.88 | ||||||
| AROM: IR | 0.93 | |||||||
| De Jong et al. [ | Hemiplegic shoulder N = 43 | PROM: ABD | Affected | Hydro-goniometer | Over time 4, 8 & 20 weeks | Inter-tester | 0.97 | Yes |
| PROM: ER | Affected | 0.94 |
ABD = abduction, ER = external rotation, IR = internal rotation, HBB = hand behind back, EL = elevation, PROM = Passive range of motion, AROM = Active range of motion, Sw = standard error of measurement, scap. stab = scapular stabilization.
With few exceptions only the inter-tester ICC values are written in the table.
Figure 1Measurement of passive range of motion in shoulder. a) Abduction. b) External rotation. c) Internal rotation. d) Hand behind back.
Mean range and standard deviation (SD) for two testers for abduction (ABD) in 50 patients with shoulder capsulitis
|
|
|
|
|
|
|
|---|---|---|---|---|---|
|
|
|
|
| ||
| 1. | ABD non-affected | 87.3 (7.7) | 88.4 (6.6) | 0.91 | 2.1 |
| ABD affected | 53.2 (17.0) | 54.1 (15.1) | 0.83 | 6.7 | |
| 2. | ABD non-affected | 88.3 (6.8) | 89.0 (4.7) | 0.88 | 2.0 |
| ABD affected | 60.6 (17.9) | 61.7 (18.7) | 0.90 | 5.9 | |
| 3. | ABD non-affected | 89.7 (2.9) | 89.7 (1.6) | 0.64 | 1.4 |
| ABD affected | 65.4 (19.5) | 68.3 (19.5) | 0.86 | 7.2 |
Relative agreement is reported with intraclass correlation coefficient, ICC (2.1), and absolute measurement error is reported with within-subject standard deviation, sw, between the two testers. ROM = Range of motion.
Mean range and standard deviation (SD) for two testers for external rotation (ER) in 50 patients with shoulder capsulitis
|
|
|
|
|
|
|
|---|---|---|---|---|---|
|
|
|
|
| ||
| 1. | ER non-affected | 70.8 (16.0) | 75.0 (17.9) | 0.83 | 6.5 |
| ER affected | 18.6 (16.2) | 22.1 (16.1) | 0.90 | 4.5 | |
| 2. | ER non-affected | 71.8 (14.1) | 76.4 (15.7) | 0.80 | 6.0 |
| ER affected | 25.0 (19.2) | 28.6 (17.6) | 0.89 | 5.8 | |
| 3. | ER non-affected | 72.0 (13.2) | 78.5 (11.3) | 0.69 | 5.7 |
| ER affected | 34.0 (22.6) | 35.3 (19.2) | 0.91 | 6.2 |
Mean range and standard deviation (SD) for two testers for internal rotation (IR) and hand behind back (HBB) in 50 patients with shoulder capsulitis
|
|
|
|
|
|
|
|---|---|---|---|---|---|
|
|
|
|
| ||
| 1. | IR non-affected | 73.0 (14.3) | 77.9 (13.9) | 0.76 | 6.2 |
| IR affected | 40.7 (14.9) | 42.4 (14.1) | 0.85 | 7 | |
| 2. | IR non-affected | 72.9 (13.0) | 78.0 (13.4) | 0.78 | 5.4 |
| IR affected | 44.3 (16.9) | 47.3 (17.8) | 0.85 | 6.4 | |
| 3. | IR non-affected | 73.8 (10.8) | 80.3 (9.9) | 0.63 | 5.1 |
| IR affected | 51.0 (18.3) | 54.1 (18.0) | 0.89 | 5.6 |
Mean range and standard deviation (SD) for two testers for hand behind back (HBB) in 50 patients with shoulder capsulitis
|
|
|
|
|
|
|
|---|---|---|---|---|---|
|
|
|
|
| ||
| 1. | HBB non-affected (cm) | 18.3 (5.7) | 18.4 (5.8) | 0.97 | 2.1 |
| HBB affected (cm) | 0.2 (6.9) | 1.5 (7.0) | 0.91 | 1.9 | |
| 2. | HBB non-affected (cm) | 19.0 (5.8) | 19.0 (5.5) | 0.98 | 0.9 |
| HBB affected (cm) | 4.9 (7.3) | 4.3 (7.9) | 0.94 | 1.8 | |
| 3. | HBB non-affected (cm) | 19.1 (5.0) | 19.4 (5.3) | 0.96 | 1.1 |
| HBB affected (cm) | 7.9 (8.6) | 8.7 (7.9) | 0.96 | 1.6 |
Figure 2Bland-Altman plot with mean abduction values for a) affected and b) non-affected arm in 50 patients with adhesive capsulitis at visit 1 for both testers plotted against the difference between testers. For the non-affected arm equal values are dispersed by a factor of 0.1 degrees at 90 degrees.