Reliability of shoulder range of motion comparing a goniometer to a digital level.

Michael J Mullaney1, Malachy P McHugh, Christopher P Johnson, Timothy F Tyler.   

Abstract

The clinical use of digital levels, for joint measurement, may be a viable alternative to standard goniometry. The purpose of this study was to determine the intra- and intertester reliability of a construction grade digital level compared to the standard universal goniometer for measurements for active assisted shoulder range of motion (ROM). Two experienced physical therapists measured shoulder flexion, external rotation (ER), and internal rotation (IR) ROM bilaterally, on two different occasions, in 20 patients (9 males, 11 females, 18-79 years old) with unilateral shoulder pathology, using a goniometer and a digital level. Relative reliability was assessed by using intraclass correlation coefficients (ICC), and absolute reliability was assessed by using 95% limits of agreement (LOA). Intratester ICCs ranged from 0.91 to 0.99, and LOA ranged from 3 degrees to 9 degrees for measurements made with the goniometer and digital level. Intertester ICCs ranged from 0.31 to 0.95, and LOA ranged from 6 degrees to 25 degrees . For the comparison of goniometric vs. digital level ROM, ICCs ranged from 0.71 to 0.98. ER and IR ROM were 3-5 degrees greater for the digital level than the goniometer (p < 0.01). Goniometric vs. digital level LOA ranged from 6 degrees to 11 degrees for shoulder flexion. Both measurement techniques had excellent intratester reliability, but for intertester reliability ICCs were 20% lower and LOA were 2.3 times higher than intratester values. Reliability estimates were similar between the digital level and the goniometer. However, because glenohumeral rotation was 3-5 degrees greater for the digital level than the goniometer (systematic error), the two methods cannot be used interchangeably. On the basis of the average intratester LOA for the goniometer and the digital level, a change of 6-11 degrees is needed to be certain that true change has occurred. For comparison of measures made by two different therapists, a change is of 15 degrees is required to be certain a true change has occurred. A digital level can be used to reliably measure shoulder ROM but should not be used interchangeably with a standard goniometer.

Entities:  

Mesh:

Year:  2010        PMID: 20557263     DOI: 10.3109/09593980903094230

Source DB:  PubMed          Journal:  Physiother Theory Pract        ISSN: 0959-3985            Impact factor:   2.279


  41 in total

1.  The reliability and concurrent validity of shoulder mobility measurements using a digital inclinometer and goniometer: a technical report.

Authors:  Morey J Kolber; William J Hanney
Journal:  Int J Sports Phys Ther       Date:  2012-06

2.  Concurrent validity of digital inclinometer and universal goniometer in assessing passive hip mobility in healthy subjects.

Authors:  Sean Roach; Jun G San Juan; David N Suprak; Marc Lyda
Journal:  Int J Sports Phys Ther       Date:  2013-10

3.  The relationship between glenohumeral joint total rotational range of motion and the functional movement screen™ shoulder mobility test.

Authors:  Peter A Sprague; G Monique Mokha; Dustin R Gatens; Rudy Rodriguez
Journal:  Int J Sports Phys Ther       Date:  2014-10

4.  Use of thoracic spine manipulation in the treatment of adhesive capsulitis: a case report.

Authors:  Joshua R McCormack
Journal:  J Man Manip Ther       Date:  2012-02

5.  Flexibility responses to different stretching methods in young elite basketball players.

Authors:  Angela Notarnicola; Fabrizio Perroni; Alessio Campese; Giuseppe Maccagnano; Antonio Monno; Biagio Moretti; Silvio Tafuri
Journal:  Muscles Ligaments Tendons J       Date:  2018-04-16

6.  THE EFFECT OF HUMERAL RETROTORSION ON PITCH VELOCITY IN YOUTH BASEBALL PLAYERS.

Authors:  Elliot M Greenberg; Alicia Fernandez-Fernandez; J Todd Lawrence; Philip McClure
Journal:  Int J Sports Phys Ther       Date:  2020-05

7.  The effectiveness of high-intensity laser therapy in the treatment of post-stroke patients with hemiplegic shoulder pain: a prospective randomized controlled study.

Authors:  Nurdan Korkmaz; Eda Gurcay; Yasin Demir; Özge Tezen; İzzet Korkmaz; Merve Örücü Atar; Evren Yaşar
Journal:  Lasers Med Sci       Date:  2021-04-08       Impact factor: 3.161

8.  ASSESSMENT OF HUMERAL TORSION BY PALPATION IN BASEBALL PITCHERS: A VALIDATION STUDY.

Authors:  Lee S Yaari; Michael J Mullaney; Takumi Fukunaga; Ran Thein; Malachy P McHugh; Stephen J Nicholas
Journal:  Int J Sports Phys Ther       Date:  2020-12

9.  DRY NEEDLING AND PHYSICAL THERAPY VERSUS PHYSICAL THERAPY ALONE FOLLOWING SHOULDER STABILIZATION REPAIR: A RANDOMIZED CLINICAL TRIAL.

Authors:  Rob Halle; Michael Crowell; Donald Goss
Journal:  Int J Sports Phys Ther       Date:  2020-02

10.  Intramuscular Electrical Stimulation Combined with Therapeutic Exercises in Patients with Shoulder Adhesive Capsulitis: A Randomised Controlled Trial.

Authors:  Sukumar Shanmugam; Lawrence Mathias; Nagarajan Manickaraj; K U Dhanesh Kumar; Praveen Kumar Kandakurti; Sathees Kumar Dorairaj; Ramprasad Muthukrishnan
Journal:  Int J Surg Protoc       Date:  2021-05-18
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.