Optimization of image quality and patient dose in radiographs of paediatric extremities using direct digital radiography.

A Jones1, C Ansell, C Jerrom, I D Honey.   

Abstract

OBJECTIVE: The purpose of this study was to evaluate the effect of beam quality on the image quality (IQ) of ankle radiographs of paediatric patients in the age range of 0-1 year whilst maintaining constant effective dose (ED).
METHODS: Lateral ankle radiographs of an infant foot phantom were taken at a range of tube potentials (40.0-64.5 kVp) with and without 0.1-mm copper (Cu) filtration using a Trixell Pixium 4600 detector (Trixell, Morains, France). ED to the patient was computed for the default exposure parameters using PCXMC v. 2.0 and was fixed for other beam qualities by modulating the tube current-time product. The contrast-to-noise ratio (CNR) was measured between the tibia and adjacent soft tissue. The IQ of the phantom images was assessed by three radiologists and a reporting radiographer. Four IQ criteria were defined each with a scale of 1-3, giving a maximum score of 12. Finally, a service audit of clinical images at the default and optimum beam qualities was undertaken.
RESULTS: The measured CNR for the 40 kVp/no Cu image was 12.0 compared with 7.6 for the default mode (55  0.1 mm Cu). An improvement in the clinical IQ scores was also apparent at this lower beam quality.
CONCLUSION: Lowering tube potential and removing filtration improved the clinical IQ of paediatric ankle radiographs in this age range. ADVANCES IN KNOWLEDGE: There are currently no UK guidelines on exposure protocols for paediatric imaging using direct digital radiography. A lower beam quality will produce better IQ with no additional dose penalty for infant extremity imaging.

Entities:  

Mesh:

Year:  2015        PMID: 25816115      PMCID: PMC4628445          DOI: 10.1259/bjr.20140660

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  24 in total

1.  The influence of different technique factors on image quality of lumbar spine radiographs as evaluated by established CEC image criteria.

Authors:  A Almén; A Tingberg; S Mattsson; J Besjakov; S Kheddache; B Lanhede; L G Månsson; M Zankl
Journal:  Br J Radiol       Date:  2000-11       Impact factor: 3.039

2.  On site evaluation of three flat panel detectors for digital radiography.

Authors:  Giovanni Borasi; Andrea Nitrosi; Paolo Ferrari; Davide Tassoni
Journal:  Med Phys       Date:  2003-07       Impact factor: 4.071

3.  Routine skeleton radiography using a flat-panel detector: image quality and clinical acceptance at 50% dose reduction.

Authors:  Markus Völk; Christian Paetzel; Peter Angele; Johannes S Seitz; Bernd Füchtmeier; Reiner Hente; Stefan Feuerbach; Michael Strotzer
Journal:  Invest Radiol       Date:  2003-04       Impact factor: 6.016

4.  Comparison of clinical and physical measures of image quality in chest and pelvis computed radiography at different tube voltages.

Authors:  Michael Sandborg; Anders Tingberg; Gustaf Ullman; David R Dance; Gudrun Alm Carlsson
Journal:  Med Phys       Date:  2006-11       Impact factor: 4.071

5.  Optimizing image quality and dose for digital radiography of distal pediatric extremities using the contrast-to-noise ratio.

Authors:  R Hess; U Neitzel
Journal:  Rofo       Date:  2012-05-22

6.  The formulation of tissue substitute materials using basic interaction data.

Authors:  D R White
Journal:  Phys Med Biol       Date:  1977-09       Impact factor: 3.609

7.  Dose optimization in pediatric cardiac x-ray imaging.

Authors:  Amber J Gislason; Andrew G Davies; Arnold R Cowen
Journal:  Med Phys       Date:  2010-10       Impact factor: 4.071

8.  Amorphous silicon, flat-panel, x-ray detector versus screen-film radiography: effect of dose reduction on the detectability of cortical bone defects and fractures.

Authors:  M Strotzer; J Gmeinwieser; M Spahn; M Völk; R Fründ; J Seitz; V Spies; J Alexander; S Feuerbach
Journal:  Invest Radiol       Date:  1998-01       Impact factor: 6.016

9.  Photostimulable phosphor digital radiography of the extremities: diagnostic accuracy compared with conventional radiography.

Authors:  A J Wilson; F A Mann; W A Murphy; B S Monsees; M R Linn
Journal:  AJR Am J Roentgenol       Date:  1991-09       Impact factor: 3.959

10.  Detection of simulated inflicted metaphyseal fractures in a fetal pig model: image optimization and dose reduction with computed radiography.

Authors:  Patricia L Kleinman; David Zurakowski; Keith J Strauss; Robert H Cleveland; Jeannette M Perez-Rosello; David P Nichols; Kelly H Zou; Paul K Kleinman
Journal:  Radiology       Date:  2008-05       Impact factor: 11.105

View more
  4 in total

1.  Neonatal digital chest radiography- should we be using additional copper filtration?

Authors:  Jenna Ruth Tugwell-Allsup; Rhys Wyn Morris; Kate Thomas; Richard Hibbs; Andrew England
Journal:  Br J Radiol       Date:  2021-12-14       Impact factor: 3.039

2.  Comparison and optimization of imaging techniques in suspected physical abuse paediatric radiography.

Authors:  Jodie Rebecca Hampel; Ana Pascoal
Journal:  Br J Radiol       Date:  2018-01-18       Impact factor: 3.039

3.  Paediatric image-guided radiation therapy: determining and evaluating appropriate kilovoltage planar exposure factors for the Varian on-board imager.

Authors:  John Ryan; David Willis
Journal:  J Med Radiat Sci       Date:  2019-09-03

4.  An evaluation of the effect of tube potential on clinical image quality using direct digital detectors for pelvis and lumbar spine radiographs.

Authors:  Nicole E Peacock; Adam L Steward; Peter J Riley
Journal:  J Med Radiat Sci       Date:  2020-06-03
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.