Comparison of clinical and physical measures of image quality in chest and pelvis computed radiography at different tube voltages.

Michael Sandborg1, Anders Tingberg, Gustaf Ullman, David R Dance, Gudrun Alm Carlsson.   

Abstract

The aim of this work was to study the dependence of image quality in digital chest and pelvis radiography on tube voltage, and to explore correlations between clinical and physical measures of image quality. The effect on image quality of tube voltage in these two examinations was assessed using two methods. The first method relies on radiologists' observations of images of an anthropomorphic phantom, and the second method was based on computer modeling of the imaging system using an anthropomorphic voxel phantom. The tube voltage was varied within a broad range (50-150 kV), including those values typically used with screen-film radiography. The tube charge was altered so that the same effective dose was achieved for each projection. Two x-ray units were employed using a computed radiography (CR) image detector with standard tube filtration and antiscatter device. Clinical image quality was assessed by a group of radiologists using a visual grading analysis (VGA) technique based on the revised CEC image criteria. Physical image quality was derived from a Monte Carlo computer model in terms of the signal-to-noise ratio, SNR, of anatomical structures corresponding to the image criteria. Both the VGAS (visual grading analysis score) and SNR decrease with increasing tube voltage in both chest PA and pelvis AP examinations, indicating superior performance if lower tube voltages are employed. Hence, a positive correlation between clinical and physical measures of image quality was found. The pros and cons of using lower tube voltages with CR digital radiography than typically used in analog screen-film radiography are discussed, as well as the relevance of using VGAS and quantum-noise SNR as measures of image quality in pelvis and chest radiography.

Mesh:

Year:  2006        PMID: 17153395     DOI: 10.1118/1.2362871

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  14 in total

1.  A method to produce and validate a digitally reconstructed radiograph-based computer simulation for optimisation of chest radiographs acquired with a computed radiography imaging system.

Authors:  C S Moore; G P Liney; A W Beavis; J R Saunderson
Journal:  Br J Radiol       Date:  2011-10       Impact factor: 3.039

2.  Visual grading regression: analysing data from visual grading experiments with regression models.

Authors:  O Smedby; M Fredrikson
Journal:  Br J Radiol       Date:  2010-03-11       Impact factor: 3.039

3.  Optimization of image quality and patient dose in radiographs of paediatric extremities using direct digital radiography.

Authors:  A Jones; C Ansell; C Jerrom; I D Honey
Journal:  Br J Radiol       Date:  2015-03-27       Impact factor: 3.039

4.  Comparison of visual grading and free-response ROC analyses for assessment of image-processing algorithms in digital mammography.

Authors:  F Zanca; C Van Ongeval; F Claus; J Jacobs; R Oyen; H Bosmans
Journal:  Br J Radiol       Date:  2012-07-27       Impact factor: 3.039

5.  Characterization and validation of the thorax phantom Lungman for dose assessment in chest radiography optimization studies.

Authors:  Sunay Rodríguez Pérez; Nicholas William Marshall; Lara Struelens; Hilde Bosmans
Journal:  J Med Imaging (Bellingham)       Date:  2018-02-06

6.  Visual grading analysis of digital neonatal chest phantom X-ray images: Impact of detector type, dose and image processing on image quality.

Authors:  M H Smet; L Breysem; E Mussen; H Bosmans; N W Marshall; L Cockmartin
Journal:  Eur Radiol       Date:  2018-02-19       Impact factor: 5.315

7.  Use of a digitally reconstructed radiograph-based computer simulation for the optimisation of chest radiographic techniques for computed radiography imaging systems.

Authors:  C S Moore; G Avery; S Balcam; L Needler; A Swift; A W Beavis; J R Saunderson
Journal:  Br J Radiol       Date:  2012-01-17       Impact factor: 3.039

8.  Optimisation of tube voltage range (kVp) for AP abdomen, pelvis and spine imaging of average patients with a digital radiography (DR) imaging system using a computer simulator.

Authors:  Craig Steven Moore; Tim Wood; Stephen Balcam; Liam Needler; Tim Guest; Wee Ping Ngu; Lee Wun Chong; John Saunderson; Andrew Beavis
Journal:  Br J Radiol       Date:  2020-08-12       Impact factor: 3.039

9.  Implementation of a patient dose monitoring system in conventional digital X-ray imaging: initial experiences.

Authors:  Christina Heilmaier; Niklaus Zuber; Dominik Weishaupt
Journal:  Eur Radiol       Date:  2016-06-23       Impact factor: 5.315

10.  Use of a computer simulator to investigate optimized tube voltage for chest imaging of average patients with a digital radiography (DR) imaging system.

Authors:  Craig Steven Moore; Tim Wood; Ged Avery; Steve Balcam; Liam Needler; Hiten Joshi; Najeeb Ahmed; John Saunderson; Andrew Beavis
Journal:  Br J Radiol       Date:  2019-10-07       Impact factor: 3.039

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.