Literature DB >> 25796205

Evaluation of micro-emboli in a patient with ventricular assist device support with hemolysis.

Osamu Seguchi1, Kozue Saito, Kazuki Fukuma, Keiko Shimamoto, Takuma Sato, Seiko Nakajima, Haruki Sunami, Kensuke Kuroda, Takamasa Sato, Takuya Watanabe, Hiroki Hata, Masanobu Yanase, Tomoyuki Fujita, Junjiro Kobayashi, Kazuyuki Nagatsuka, Takeshi Nakatani.   

Abstract

The clinical relevance of transcranial Doppler (TCD) detection of micro-embolic signals (MES) in patients with ventricular assist devices (VADs) has been described. However, all of the previous studies concerning TCD in patients with VADs were conducted in patients with old devices; the clinical relevance of TCD in patients with newer devices has not been fully elucidated. We recently encountered a patient with a continuous-flow VAD with hemolysis. TCD monitoring was useful for the direct evaluation of micro-emboli in this patient. A 50-year-old male who underwent HeartMate II(®) VAD (Thoratec Corporation; Pleasanton, CA) implantation with a diagnosis of ischemic cardiomyopathy 15 months prior was admitted to our institution because of findings suggestive of hemolysis, such as elevated lactate dehydrogenase (LDH) and total bilirubin. Unfractionated heparin was started after admission and hemolysis gradually improved. On admission, TCD detected 146 MES during 30 min of monitoring. During the hospital course, the MES count decreased to 20 signals on hospital day 4 and further decreased to 2 signals on hospital day 15 along with decreases in LDH and total bilirubin. Since hemolysis in VAD patients is thought to be associated with thromboembolic outcomes, MES detected by TCD indicate subclinical micro-emboli. TCD monitoring may be useful for assessing the risk of thromboembolic events in newer continuous-flow VAD patients through direct visualization of micro-emboli.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25796205     DOI: 10.1007/s10047-015-0832-2

Source DB:  PubMed          Journal:  J Artif Organs        ISSN: 1434-7229            Impact factor:   1.731


  19 in total

1.  ULTRASOUND AS A METHOD TO DETECT BUBBLES OR PARTICULATE MATTER IN THE ARTERIAL LINE DURING CARDIOPULMONARY BYPASS.

Authors:  W G AUSTEN; D H HOWRY
Journal:  J Surg Res       Date:  1965-06       Impact factor: 2.192

2.  Clinically silent microemboli in patients with artificial prosthetic aortic valves are predominantly gaseous and not solid.

Authors:  M Kaps; J Hansen; M Weiher; K Tiffert; I Kayser; D W Droste
Journal:  Stroke       Date:  1997-02       Impact factor: 7.914

3.  Transcranial detection of microembolic signals in patients with a novel nonpulsatile implantable LVAD.

Authors:  E V Potapov; B A Nasseri; M Loebe; M Kukucka; A Koster; H Kuppe; G P Noon; M E DeBakey; R Hetzer
Journal:  ASAIO J       Date:  2001 May-Jun       Impact factor: 2.872

4.  Influence of inflammation and pump dynamic on cerebral microembolization in patients with Continuous-Flow DeBakey LVAD.

Authors:  Nils H Thoennissen; Achim Allroggen; Martin Ritter; Ralf Dittrich; Christof Schmid; Hans H Schmid; E Bernd Ringelstein; Darius G Nabavi
Journal:  ASAIO J       Date:  2006 May-Jun       Impact factor: 2.872

5.  Low thromboembolism and pump thrombosis with the HeartMate II left ventricular assist device: analysis of outpatient anti-coagulation.

Authors:  Andrew J Boyle; Stuart D Russell; Jeffrey J Teuteberg; Mark S Slaughter; Nader Moazami; Francis D Pagani; O Howard Frazier; Gerald Heatley; David J Farrar; Ranjit John
Journal:  J Heart Lung Transplant       Date:  2009-09       Impact factor: 10.247

6.  Advanced heart failure treated with continuous-flow left ventricular assist device.

Authors:  Mark S Slaughter; Joseph G Rogers; Carmelo A Milano; Stuart D Russell; John V Conte; David Feldman; Benjamin Sun; Antone J Tatooles; Reynolds M Delgado; James W Long; Thomas C Wozniak; Waqas Ghumman; David J Farrar; O Howard Frazier
Journal:  N Engl J Med       Date:  2009-11-17       Impact factor: 91.245

7.  Doppler microembolic load predicts risk of thromboembolic complications in Novacor patients.

Authors:  Darius G Nabavi; Joerg Stockmann; Christof Schmid; Michael Schneider; Dieter Hammel; Hans H Scheld; E Bernd Ringelstein
Journal:  J Thorac Cardiovasc Surg       Date:  2003-07       Impact factor: 5.209

8.  Cerebral and systemic embolization during left ventricular support with the Novacor N100 device.

Authors:  C Schmid; M Weyand; D G Nabavi; D Hammel; M C Deng; E B Ringelstein; H H Scheld
Journal:  Ann Thorac Surg       Date:  1998-06       Impact factor: 4.330

9.  Transcranial Doppler-detected microemboli in patients with acute stroke.

Authors:  D C Tong; G W Albers
Journal:  Stroke       Date:  1995-09       Impact factor: 7.914

10.  Low thromboembolic risk for patients with the Heartmate II left ventricular assist device.

Authors:  Ranjit John; Forum Kamdar; Kenneth Liao; Monica Colvin-Adams; Leslie Miller; Lyle Joyce; Andrew Boyle
Journal:  J Thorac Cardiovasc Surg       Date:  2008-09-14       Impact factor: 5.209

View more
  2 in total

1.  Haemolysis induced by mechanical circulatory support devices: unsolved problems.

Authors:  Inge Köhne
Journal:  Perfusion       Date:  2020-06-23       Impact factor: 1.972

2.  Simulation study on the effects of cancellous bone structure in the skull on ultrasonic wave propagation.

Authors:  Itsuki Michimoto; Kazuki Miyashita; Hidehisa Suzuyama; Keita Yano; Yasuyo Kobayashi; Kozue Saito; Mami Matsukawa
Journal:  Sci Rep       Date:  2021-09-02       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.