Chest circumference as a predictive parameter of computed tomography coronary angiography radiation doses from dual-source computed tomography.

M Erdem Sagsoz1, Ummugulsum Bayraktutan2, Hayri Ogul2, Mecit Kantarci2.   

Abstract

OBJECTIVE: Our aim in this study was to determine the relationship between chest circumference and the radiation doses received by breast tissues during Dual-Source Computed Tomography (DSCT) cardiac scans.
MATERIALS AND METHODS: Routine cardiac DSCT examinations with similar exposure lengths were applied to 30 female patients. The scanogram image, multi-slice helical scan x-ray tube voltage and anode-cathode current were adjusted automatically according to attenuation for each patient during the scanogram acquisition. The standard protocol was otherwise applied to all patients. The 30 patients had an average weight of 67.8±15.3 kg, and the average length of the scanned region was 278.0±11.6 mm. Radiation doses were calculated from the dose-length product (DLP) and the computed tomography dose index (CTDI) scanner data values. The correlations between radiation dose and chest circumference were investigated. The level of significance was set at p<0.05.
RESULTS: For routine cardiac DSCT scans, the average values were as follows: total DLP: 715.54±317.01 mGycm, CTDIvol: 40.79±19.41 mGy, and effective dose (ED): 17.89±7.93 mSv. The chest circumference of patients correlates well with their radiation exposure (p<0.01).
CONCLUSION: The distribution of different tissues throughout the human body may vary among races and genders. Because of this, many researchers use body mass index (BMI) to set image quality and predict the radiation dose distribution from general computed tomography (CT) examinations. Additional anthropomorphic phantom studies should be conducted to determine more accurate conversion factors and, hence, better ED predictions.

Entities:  

Keywords:  Breast; Cardiac; DSCT; Radiation dose

Year:  2013        PMID: 25610247      PMCID: PMC4261497          DOI: 10.5152/eajm.2013.07

Source DB:  PubMed          Journal:  Eurasian J Med        ISSN: 1308-8734


  19 in total

1.  X-ray induced DNA double-strand breaks in coronary CT angiography: comparison of sequential, low-pitch helical and high-pitch helical data acquisition.

Authors:  Michael Brand; Matthias Sommer; Stephan Achenbach; Katharina Anders; Michael Lell; Markus Löbrich; Michael Uder; Michael A Kuefner
Journal:  Eur J Radiol       Date:  2011-12-16       Impact factor: 3.528

2.  Image quality and radiation dose of low dose coronary CT angiography in obese patients: sinogram affirmed iterative reconstruction versus filtered back projection.

Authors:  Rui Wang; U Joseph Schoepf; Runze Wu; Ryan P Reddy; Chuanchen Zhang; Wei Yu; Yi Liu; Zhaoqi Zhang
Journal:  Eur J Radiol       Date:  2012-05-10       Impact factor: 3.528

3.  Computing effective doses from dose-length product in CT.

Authors:  Walter Huda
Journal:  Radiology       Date:  2008-07       Impact factor: 11.105

4.  High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography.

Authors:  Nico R Mollet; Filippo Cademartiri; Carlos A G van Mieghem; Giuseppe Runza; Eugène P McFadden; Timo Baks; Patrick W Serruys; Gabriel P Krestin; Pim J de Feyter
Journal:  Circulation       Date:  2005-10-03       Impact factor: 29.690

5.  Estimation and comparison of effective dose (E) in standard chest CT by organ dose measurements and dose-length-product methods and assessment of the influence of CT tube potential (energy dependency) on effective dose in a dual-source CT.

Authors:  Jijo Paul; Rosemarie Banckwitz; Bernhard Krauss; Thomas J Vogl; Werner Maentele; Ralf W Bauer
Journal:  Eur J Radiol       Date:  2011-06-23       Impact factor: 3.528

6.  Dual-source CT coronary angiography: effectiveness of radiation dose reduction with lower tube voltage.

Authors:  Akmal Sabarudin; Ahmad Khairuddin Md Yusof; May Fang Tay; Kwan-Hoong Ng; Zhonghua Sun
Journal:  Radiat Prot Dosimetry       Date:  2012-07-17       Impact factor: 0.972

7.  Gender influence on dose saving allowed by prospective-triggered 64-slice multidetector computed tomography coronary angiography as compared with retrospective-gated mode.

Authors:  Antonio Esposito; Francesco De Cobelli; Caterina Colantoni; Gianluca Perseghin; Antonella Del Vecchio; Tamara Canu; Riccardo Calandrino; Alessandro Del Maschio
Journal:  Int J Cardiol       Date:  2011-02-16       Impact factor: 4.164

8.  Dual-source computed tomography: estimation of radiation exposure of ECG-gated and ECG-triggered coronary angiography.

Authors:  Dominik Ketelsen; Christoph Thomas; Matthias Werner; Marie H Luetkhoff; Markus Buchgeister; Ilias Tsiflikas; Anja Reimann; Christof Burgstahler; Harald Brodoefel; Andreas F Kopp; Claus D Claussen; Martin Heuschmid
Journal:  Eur J Radiol       Date:  2008-12-18       Impact factor: 3.528

9.  Initial evaluation of coronary images from 320-detector row computed tomography.

Authors:  Frank J Rybicki; Hansel J Otero; Michael L Steigner; Gabriel Vorobiof; Leelakrishna Nallamshetty; Dimitrios Mitsouras; Hale Ersoy; Richard T Mather; Philip F Judy; Tianxi Cai; Karl Coyner; Kurt Schultz; Amanda G Whitmore; Marcelo F Di Carli
Journal:  Int J Cardiovasc Imaging       Date:  2008-03-27       Impact factor: 2.357

10.  Radiation dose of cardiac dual-source CT: the effect of tailoring the protocol to patient-specific parameters.

Authors:  Hatem Alkadhi; Paul Stolzmann; Hans Scheffel; Lotus Desbiolles; Stephan Baumüller; André Plass; Michele Genoni; Borut Marincek; Sebastian Leschka
Journal:  Eur J Radiol       Date:  2008-10-30       Impact factor: 3.528

View more
  2 in total

1.  Multidetector computed tomographic imaging of Erdheim-Chester disease.

Authors:  Zeyneb Yuceler; Mecit Kantarci; Nevzat Karabulut; Hayri Ogul; Ummugulsum Bayraktutan; Canan Akman
Journal:  Tex Heart Inst J       Date:  2014-06-01

2.  Effectiveness of Using Dual-source CT and the Upshot it creates on Both Heart Rate and Image Quality.

Authors:  Tuba Selçuk; Hafize Otçu; Zeyneb Yüceler; Çiğdem Bilgili; Mesut Bulakçı; Yıldıray Savaş; Ömer Çelik
Journal:  Balkan Med J       Date:  2016-05-01       Impact factor: 2.021

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.