| Literature DB >> 25563495 |
M J Suttorp1, P R Stella, J Dens, J M McKenzie, K S Park, P Frambach.
Abstract
AIM: To report clinical follow-up at 6 months after implantation of the ultra-thin strut cobalt chromium SolarFlex stent in a real-world setting. METHODS ANDEntities:
Year: 2015 PMID: 25563495 PMCID: PMC4315800 DOI: 10.1007/s12471-014-0629-6
Source DB: PubMed Journal: Neth Heart J ISSN: 1568-5888 Impact factor: 2.380
Fig. 1Schematic illustration showing the uniform distance between stent struts over all five stent sizes
Baseline clinical and angiographic characteristics
| Number of patients | 240 |
| Age (years) | 66 ± 12.6 |
| Male (%) | 178 (74 %) |
| Baseline angina status | |
| Asymptomatic | 7 (2.9 %) |
| Silent ischaemia | 12 (5.0 %) |
| Stable | 108 (45.0 %) |
| Unstable | 40 (16.7 %) |
| AMI | 73 (30.4 %) |
| History of | |
| MI | 67 (27.9 %) |
| CABG | 16 (6.7 %) |
| PCI | 71 (29.6 %) |
| Risk factors | |
| Diabetes mellitus type 1 | 38 (16 %) |
| Diabetes mellitus type 2 | 31 (13 %) |
| Hypertension | 138 (58 %) |
| Cholesterol | 142 (59 %) [ |
| Family history | 114 (48 %) [ |
| Smoking | 186 (22 %) [ |
| Number of lesions | 292 |
| Target lesion classification | |
| A | 50 (17.1 %) |
| B1 | 92 (31.5 %) |
| B2 | 110 (37.7 %) |
| C | 37 (12.7 %) |
| Unknown | 3 (1.0 %) |
| Vessels treated | |
| Heft anterior descending | 106 (36.3 %) |
| Right coronary artery | 108 (37.0 %) |
| Circumflex | 74 (25.3 %) |
| Saphenous valve graft | 3 (1.0 %) |
| Ramus | 1 (0.3 %) |
| Angulation | |
| < 45° | 236 (80.8 %) |
| ≥ 45° | 47 (16.1 %) |
| ≥ 90° | 7 (2.4 %) |
| Unknown | 2 (0.7 %) |
| Tortuosity | |
| Low | 186 (63.7 %) |
| Moderate | 86 (29.5 %) |
| High | 18 (6.2 % |
| Unknown | 2 (0.7 %) |
| Complexities | |
| Bifurcation | 35 (12.0 %) |
| Ostial | 15 (5.1 %) |
| Thrombus | 32 (11.0 %) |
| Calcification | 77 (26.4 %) |
| Diffuse disease | 32 (11.0 %) |
| Involves side branch | 12 (4.1 %) |
aWhere data were not available for all patients the number of patients with data is indicated (n)
Clinical outcomes
| Number of stents | 293 |
| Average stent dimensions | |
| Diameter (mm) | 3.15 ± 0.38 |
| Length (mm) | 17.47 ± 4.82 |
| Predilation | 176 (60.1 %) |
| Direct stenting | 117 (39.9 %) |
| Post-dilation | 60 (20.5 %) |
| Reason for stenting | |
| Elective | 265 (90.4 %) |
| Dissection | 23 (7.8 %) |
| Other | 5 (1.7 %) |
| Delivery success (stents reaching lesion) | 293 (100 %) |
| Device success (per stent) | 292 (99.7 %) |
| Procedural success (per patient) | 234 (97.5 %) |
| Follow-up | 240 (100 %) |
| In-hospital events | |
| Total MACE | 5 (2.1 %) |
| Cardiac death | 0 |
| Myocardial infarction | 1 (0.4 %) |
| Target lesion revascularisation | 4 (1.7 %) |
| Stent thrombosis (definite) | 3 (1.3 %) |
| 6-month outcome (cumulative) | |
| Total MACE | 14 (5.8 %) |
| Cardiac death | 1 (0.4 %) |
| Myocardial infarction | 3 (1.2 %) |
| Target lesion revascularisation | 12 (5.0 %) |
| Stent thrombosis (probable) | 1 (0.4 %) |
| Stent thrombosis (definite) | 4 (1.7 %) |
Data from a review of all six- to nine-month multi-centre registries or clinical trials of single Co-Cr stent types with more 100 patients (MEDLINE search results extend to 31 May 2013; table excludes studies restricted to patients presenting only with STEMI)
| Registry | Stent name | Material | Strut thickness (μm) | Number of patients | Average stent length (mm) | Average stent diameter (mm) | Presenting conditions | Follow-up (months) | Delivery success | MACE | TLR/TVRa |
|---|---|---|---|---|---|---|---|---|---|---|---|
| MULTIBENE [ | PRO-Kinetik | SiC-coated L605 | 60–120 | 198 | 15.7 ± 3 .5 | 3.21 ± 0.45 | No AMI | 6 | – | 11.4 % | 7.6 % |
| Arthos Pico Austrian Multicentre [ | Arthos Pico | Phynox Co-Cr | 65 | 203 | 16.2 ± 6.5 | 2.58 ± 0.22 | Real-world | 6 | 98.0 % | 13.0 % | 6.0 % |
| SOLSTICE | SolarFlex | L605 Co-Cr | 65 | 240 | 17.5 ± 4.82 | 3.15 ± 0.38 | Real-world | 6 | 100 % | 5.8 % | 5.0 % |
| CoroFlex Blue registry [ | CoroFlex Blue | L605 Co-Cr | 65 | 2315 | 15.6 ± 4.4 | 3.01 ± 0.43 | Real-world | 6 | 98.5 % | 9.2 % | 5.5 % |
| MILES [ | Skylor | L605 Co-Cr | 70–95 | 1020 | 19.5 ± 10.3 | 3.03 ± 0.43 | Real-world | 6+ | 98 % | 9.9 % | 5.0 % |
| VISION registry [ | Vision | L605 Co-Cr | 81 | 267 | 17.2 ± 6.3 | 3.34 ± 0.37 | No AMI | 6 | 100 % | 6.2 % | 4.3 % |
| VIVE [ | Vision | L605 Co-Cr | 81 | 429 | – | – | Real-world | 6 | 99.3 % | 6.8 % | 1.4 % |
| RISICO [ | Vision | L605 Co-Cr | 81 | 143 | 17.0 ± 3.9 | 2.41 ± 0.14 | No AMI | 6 | 96.7 % | 11.6 % | 5.8 % |
| Driver registry [ | Driver | MP35N Co-Cr | 91 | 298 | – | – | Real-world | 9 | 100 % | 8.4 % | 7.0 % |
| CLASS [ | Driver | MP35N Co-Cr | 91 | 202 | – | – | No AMI | 6 | 100 % | 12.4 % | 9.4 % |
aWhere available as an endpoint, cTLR data were been used preferentially for comparison purposes