| Literature DB >> 27625684 |
Krzysztof Milewski1, Paweł Gąsior1, Stefan Samborski1, Piotr P Buszman1, Aleksandra Błachut1, Adam Wojtaszczyk1, Adam Młodziankowski1, Artur Mendyk1, Marek Król1, Wojciech Wojakowski2, Paweł E Buszman1.
Abstract
INTRODUCTION: Despite the dominance of drug-eluting stents in modern interventional cardiology, there is still a niche for bare metal stents. AIM: The aim of the Polish NexGen registry was to evaluate the safety and efficacy of a new generation cobalt-chromium NexGen stent in a real life patient population.Entities:
Keywords: all-comers patient population; cobalt-chromium; hybrid cell design; stent; ultrathin struts
Year: 2016 PMID: 27625684 PMCID: PMC5011537 DOI: 10.5114/aic.2016.61642
Source DB: PubMed Journal: Postepy Kardiol Interwencyjnej ISSN: 1734-9338 Impact factor: 1.426
Figure 1Hybrid cell design and morphology mediated expansion of NexGen stent
Baseline clinical characteristics
| Parameter | Number | Percentage or standard deviation |
|---|---|---|
| Age | 66.6 | SD = 10.22 |
| Male | 257 | 67.1% |
| Female | 126 | 32.9% |
| Diabetes mellitus | 116 | 30.29% |
| Hypertension | 311 | 81.41% |
| Dyslipidemia | 192 | 50.13% |
| Current smoker | 87 | 22.72% |
| History of MI | 109 | 28.46% |
| History of PCI | 93 | 24.28% |
| History of CABG | 12 | 3.13% |
| SA | 39 | 10.18% |
| UA | 127 | 33.16% |
| NSTEMI | 83 | 21.67% |
| STEMI | 134 | 34.99% |
| LVEF (%) | 50.66 | SD = 10.2 |
MI – myocardial infarction, PCI – percutaneous coronary intervention, CABG – coronary artery bypass graft, SA – stable angina, UA – unstable angina, NSTEMI – non-ST segment elevation myocardial infarction, STEMI – ST segment elevation myocardial infarction, LVEF – left ventricular ejection fraction.
Baseline angiographic characteristics
| Variable | Number | Percentage or interquartile range |
|---|---|---|
| 1-vessel CAD | 105 | 27.42% |
| 2-vessel CAD | 135 | 35.25% |
| 3-vessel CAD | 143 | 37.44% |
| Lesions distribution: | ||
| LM | 24 | 6.27% |
| LAD | 247 | 64.49% |
| Cx | 230 | 60.05% |
| RCA | 303 | 79.11% |
| SVG | 6 | 1.57% |
| LIMA | 1 | 0.26% |
| Ostial lesion | 17 | 4.51% |
| Bifurcations | 29 | 7.71% |
| Thrombus | 123 | 32.62% |
| ACC/AHA lesion type based on core lab (42 patients): | ||
| Type A | 7 | 16.7% |
| Type B1 | 13 | 31.0% |
| Type B2 | 16 | 38.1% |
| Type C | 6 | 14.3% |
| Number of stents per lesion: | ||
| 1 | 276 | 72.44% |
| 2 | 94 | 24.67% |
| 3 | 8 | 2.1% |
| 4 | 2 | 0.52% |
| 5 | 1 | 0.26% |
| Average stent length per lesion [mm] | 29.0 | IQR: 19.0–40.0 |
| Mean stent diameter [mm] | 3.0 | IQR: 2.75–3.50 |
| Mean inflation pressure [atm] | 16.0 | IQR: 14.0–18.0 |
| Direct stenting | 175 | 46.54% |
CAD – coronary artery disease, LM – left main, LAD – left anterior descending artery, Cx – circumflex artery, RCA – right coronary artery, SVG – saphenous vein graft, LIMA – left internal mammary artery.
Rate of stent thrombosis during hospitalization and follow-up
| Stent thrombosis | Number | Percentage |
|---|---|---|
| Acute ST (hospitalization) | 1 | 0.26 |
| ST at 6-month follow-up | 1 | 0.27 (0.26) |
| ST at 12-month follow-up | 2 | 0.54 (0.52) |
ST – stent thrombosis.
Rate of composite MACE and individual components of MACE and stroke at 30-day, 6- and 12-month follow-up
| Variable | Number | Percentage |
|---|---|---|
| Thirty-day follow-up ( | ||
| MACE | 23 | 6.01 |
| TVR | 10 | 2.61 |
| TLR | 10 | 2.60 |
| Subacute stent thrombosis | 1 | 0.26 |
| Myocardial infarction | 7 | 1.83 |
| Stroke | 0 | 0.00 |
| All cause death | 9 | 2.35 |
| 6-month follow-up ( | ||
| MACE | 69 | 18.50 |
| TVR | 47 | 12.60 |
| TLR | 39 | 10.46 |
| Stent thrombosis | 1 | 0.27 |
| Myocardial infarction | 13 | 3.49 |
| Stroke | 4 | 1.07 |
| All cause death | 14 | 3.75 |
| 12-month follow-up ( | ||
| MACE | 92 | 25.21 |
| TVR | 65 | 17.81 |
| TLR | 50 | 13.70 |
| Stent thrombosis | 2 | 0.54 |
| Myocardial infarction | 16 | 4.38 |
| Stroke | 5 | 1.37 |
| All cause death | 19 | 5.21 |
MACE – major adverse cardiac events, TVR – target vessel revascularization, TLR – target lesion revascularization.
Quantitative coronary artery analysis
| Parameter | Value |
|---|---|
| Before implantation: | |
| MLD [mm] | 0.57 ±0.63 |
| RD [mm] | 2.96 ±0.77 |
| %DS | 80.93 ±19.28 |
| %AS | 92.74 ±8.41 |
| After implantation: | |
| MLD [mm] | 2.60 ±0.57 |
| RD [mm] | 3.03 ±0.56 |
| %DS | 14.42 ±7.21 |
| %AS | 26.30 ±11.86 |
| Acute gain [mm] | 2.06 ±0.60 |
| 6-month follow-up: | |
| MLD [mm] | 1.98 ±0.94 |
| RD [mm] | 2.90 ±0.67 |
| %DS | 33.09 ±24.77 |
| %AS | 49.24 ±24.84 |
| Late loss [mm] | 0.66 ±0.71 |
| Binary restenosis | 16.67% ( |
MLD – minimal lumen diameter, RD – reference diameter, %DS – % diameter stenosis, %AS – % area stenosis.